CASE REPORT

Early Femoral Stem Fracture in an Obese Patient Following Total Hip Arthoplasty.

Address for Correspondence

Stephen R Manning

Clinical Fellow

Department of Medical Education 

Sandwell General Hospital

Lyndon ,West Bromwich

B71 4NG ,(0044)7866 859959


E-mail: stephen.manning@doctors.org.uk

Abstract:

Herein is presented a case of an unusual injury following a total hip replacement. Many of us would consider the metal implant to be the strongest, most resilient part of a hip joint after replacement. This may not always be the case, as the following case illustrates, metal may break when force is applied leaving the bone intact.

J.Orthopaedics 2012;9(1)e8

Keywords:

Metalwork failure; femoral stem fracture; prosthetic fracture


Case Report:

A 66 year old lady had undergone a right total hip replacement some 4 weeks earlier. She had been recovering well and was at home mobilising comfortably with the aid of crutches. Her medical history was of diabetes, controlled with metformin. She is a non-smoker who lives in a house with her husband. She had been managing the stairs safely on her crutches. Of note, is that she is a lady of considerable size with a BMI of 40. She presented to casualty unable to flex or extend her right hip after falling off a small step in her garden, landing on her right hip. It was not particularly painful. The hip was held in a varus position with flexion of 10 degrees. Her surgical wound was healed. She had not sustained any other injury. An AP radiograph of the pelvis was taken to look for dislocation or peri-prosthetic fracture. (figure 1).

Figure 1. AP Pelvis radiograph

The radiograph however, did not show either of these conditions. Instead, a break in the metal of the femoral prosthesis was seen. The impact from the fall had snapped and bent the femoral neck. The bone was intact and the femoral head was sitting in the acetabulum. This lady was admitted to hospital under the care of the orthopaedic surgeons and a revision of her femoral component was carried out, from which she recovered well.

Discussion:
Dislocation was considered as a differential diagnosis. Andrew et al., (2008)1 looked at hip replacement surgery in the obese patient and found that there was no significant difference in the rates of dislocation between the non-obese, obese and morbidly obese patient (1.3, 2.7 and 5.6% respectively) Dislocation due to trauma is described by von Knoch et al., (2002)2 as a cause of late dislocation (median time 11.3 years post operation) however it is associated with substantial trauma. Khan et al., (1981)3 also looked at dislocations and identified 142 patients with dislocation after hip replacement. Just 10 of these clearly resulted from trauma and again, were late dislocations. From the literature available dislocation would be uncommon after a fall 4 weeks post operatively, and the evidence suggests the risk of dislocation was not increased by her body mass.

Periprosthetic fracture is an uncommon but important complication of cemented total hip replacement. Cook et al.,(2008)4 have shown that the incidence of fracture at 5 years post surgery is 0.8% and at 10 years post surgery is 3.5%. The only risk factor for peri-prosthetic fracture identified was age over 70 years. This correlated with the work of Lindahl et al., (2006)5 who identified an average age of 73.9 years for patients suffering peri-prosthetic fractures. They found that minor trauma, such as a fall from standing or spontaneous fracture were the most common causes but also that 66% of those suffering fracture after primary hip replacement had some evidence of implant loosening. Neither of these studies identified any metal work fractures. The patient described here certainly had a fall that could be classed as minor trauma, however, allowing for her body mass the impact into the hip would be expected to have been higher than that experienced by a person of lesser body mass. She is below the average age seen in the described studies and at 4 weeks post op there is no evidence of implant loosening. It would seem that peri-prosthetic fracture would be an unlikely finding in this patient.

Busch et al., (2005)6 identified 5 metal work fractures out of a series of 219 revision hip procedures using a cementless system. Risk factors for metalwork fracture in this series were identified as BMI >30, poor proximal bone support, extended trochanteric oseteotomy, and small stem diameter. Other studies have consistently identified increased BMI as a risk factor for metal work failure of both primary and revision hip prostheses. 7, 8, 9, 10, 11 Although this patient had not undergone a revision hip replacement it would be reasonable to consider her BMI to be a risk factor for fracture of the metal femoral prosthesis she was given.

Could metal fatigue have lead to a fatigue fracture? This also seems unlikely as fatigue fractures are the result of repetitive loading and unloading cycles. Fatigue failure may occur in implanted materials but within a short period after implantation it is unlikely that cyclical loading would have occurred to a level to cause fatigue failure. We expect and design hip prostheses to last 10 or more years. It is worth considering the offset used in placing this prosthesis. As offset increases so too does stress through the femoral prosthesis neck. The level of stress placed upon this prosthesis therefore would be greater with an increased offset and greater still in this patient due to her increased BMI.

From the literature available, it seems that fractures of the femoral component of a cemented total hip replacement are a very rare event, but are associated with having a higher BMI unlike dislocations or peri-prosthetic fractures

    

Conclusion

Peri-prosthetic fractures after hip replacement surgery are not common and tend to occur after minor trauma in older patients with loosening of the prosthesis. Dislocations are more common but are not associated with trauma until later and are not associated with higher body mass Metalwork fractures after total hip replacement are rare but do happen and are associated with increased BMI. This case is unusual in that the metal work fracture occurred so soon after implantation. Consider this diagnosis in the larger patient suffering trauma to a prosthetic hip joint. Remember to inspect the metalwork seen on radiographs as well as the bones.

References:

    1.Andrew JG, Palan J, Murray DW, Beard DJ, Kurup HV., and Gibson P. Obesity in total hip replacement. J Bone Joint Surg Br 2008 90 424-429

    2. von Knoch M, Berry DJ, Harmsen WS, Morrey BF.Late dislocation after total hip arthroplasty J Bone Joint Surg Am. 2002 Nov;84-A(11):1949-53.

    3. Khan MAA, Brackenbury PH, and Reynolds ISR. Dislocation following total hip replacement. J Bone Joint Surg Br 1981 63-B(2) 214-218

    4. Cook RE, Jenkins PJ, Walmsley PJ, Patton JT, and Robinson CM. Risk factors for periprosthetic fractures of the hip: a survivorship analysis. Clin Orthop Relat Res. 2008 Jul;466(7):1652-6.

    5. Lindahl H, Garellick G, Regnér H, Herberts P, and Malchau H. Three hundred and twenty-one periprosthetic femoral fractures. J Bone Joint Surg Am. 2006 Jun;88(6):1215-22.

    6 Busch CA, Charles MN, Haydon CM, Bourne RB, Rorabeck CH, Macdonald SJ, McCalden RW. Fractures of distally-fixed femoral stems after revision arthroplasty. J Bone Joint Surg Br. 2005 Oct;87(10):1333-6.

    7. Andriacchi TP, Gelante JO, Belytschko TB and Hampton S. A stress analysis of the femoral stem in total hip prostheses. J Bone Joint Surg Am 1976; 58:618-24

    8. Miller EH, Shastri R, and Shih CI. Fracture failure of a forged vitallium prosthesis. A case report. J Bone Joint Surg Am 1982; 64:1359-63

    9. Mollan RA, Watters PH, Steele R and McClelland CJ. Failure of the femoral component in the Howse total hip arthroplasty. Clin Orthop Relat Res. 1984; 190:142-7

    10. Buttaro MA, Mayor MB, Van Citters D, and Piccaluga F. Fatigue fracture of a proximally modular, distally tapered fluted implant with diaphyseal fixation. J Arthoplasty 2007; 22:780-3

    11. Lakstein D, Eliaz N, Levi O, Backstein D, Kosashvili Y, Safir O and Gross AE. Fracture of cementless femoral stems at the mid-stem junction in modular revision hip arthroplasty systems. J Bone Joint Surg Am 2011; 93:57-65

 

 

This is a peer reviewed paper 

Please cite as :Early Femoral Stem Fracture in an Obese Patient Following Total Hip Arthoplasty.

J.Orthopaedics 2012;9(1)e8

URL: http://www.jortho.org/2012/9/1/e8

ANNOUNCEMENTS


 

Arthrocon 2011


Refresher Course in Hip Arthroplasty

13th March,  2011

At Malabar Palace,
Calicut, Kerala, India

Download Registration Form

For Details
Dr Anwar Marthya,
Ph:+91 9961303044

E-Mail:
anwarmh@gmail.com

 

Powered by
VirtualMedOnline

 

   
© Copyright of articles belongs to the respective authors unless otherwise specified.Verbatim copying, redistribution and storage of this article permitted provided no restrictions are imposed on the access and a hyperlink to the original article in Journal of Orthopaedics maintained. All opinion stated are exclusively that of the author(s).
Journal of Orthopaedics upholds the policy of Open Access to Scientific literature.