Use of augmented reality surgical navigation reduces intraoperative fluoroscopic imaging in anterior total hip arthroplasty with equivalent accuracy for component positioning
Article History
-
Received: ,
Accepted: ,
Abstract
Abstract
Computer-assisted surgical navigation was designed to reduce the rates of acetabular component malpositioning, a common cause for revision following primary total hip arthroplasty (THA). We compared intraoperative measures such as radiation exposure, radiographic acetabular cup placement, and outcomes of patients who underwent a THA using computer-assisted surgical navigation with augmented reality (AR) compared to those who underwent traditional manual THA with fluoroscopic assistance (FA).
We retrospectively reviewed 115 hips who underwent primary direct anterior THAs using the intraoperative AR surgical navigation system between 3/30/2021 and 12/30/2022. These were compared to a matched cohort of 115 primary THAs performed with FA through a direct anterior approach during the same time period. Patients were matched based on age, gender, American Society of Anesthesiologists score (ASA), and BMI. Goal inclination and anteversion were set to 42° (°) and 22°, respectively.
THA with AR patients had decreased intraoperative radiation exposure (mean 2.0 versus 3.2 mGy, p < 0.001) and radiation time (mean 0.2 versus 0.4 min, p < 0.001). Deviation from the preoperative plan was similar between AR and FA for both radiographic inclination and anteversion. There was no difference in the operative time, estimated blood loss, or reoperations, all of which were for infection (1.7 versus 0.0 %, p = 0.5).
AR surgical navigation in primary THA has similar radiographic, operative and clinical results as THA with intraoperative FA. However, AR surgical navigation is associated with decreased intraoperative fluoroscopic radiation and time. Further research is needed to evaluate accuracy of acetabular component positioning and complications.
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
1 Introduction
1.1 Background/rationale
Augmented reality (AR) describes the simulation of virtual objects within real-world surroundings; this approach often utilizes headsets that project images on top of translucent lenses, allowing users to interact with these simulated elements. The hardware used in these extended reality devices includes headsets, smartphones, and external monitor-based device categories.1
While the consumer AR market has centered around entertainment, commercial applications in medicine have steadily increased. Surgical fields initially used these technologies for training,2 but have increasingly sought to apply them in the operating room to improve clinical outcomes. For example, AR has been used in spine procedures to address challenges with accurate hardware placement.3 Multiple specialties, including adult reconstruction, may benefit from this technology and its ability to improve visualization of structures while performing hardware placement.4–6
THA has a steep learning curve that generally plateaus around 100 cases post-fellowship.7 One crucial and challenging aspect of THAs is the precise placement of the acetabular cup, defined by ideal anteversion and inclination angles. Multiple strategies have been developed to improve cup positioning, including intraoperative radiographs, fluoroscopy, mechanical navigation devices, CT-based assistance, and robots.8,9
Multiple studies, including one meta-analysis, have examined the accuracy of THA alignment using AR relative to conventional approaches.10–16 Four of the studies found decreased absolute errors of anteversion and inclination.12–14 while two studies found decreased absolute errors of either anteversion or inclination.15,16 One study found no differences in absolute errors.11 The meta-analysis found improved accuracy of both anteversion and inclination overall.10 Of note, these results are difficult to interpret because comparative conventional approaches were not uniform across studies. Some compare AR to accelerometer-based devices,11,13,16 while others compare to manual approaches utilizing mechanical guides12,14,15 or both.10
AR is a relatively broad term, and the technology behind these studies is also somewhat heterogeneous. Most studies utilized a smartphone-based AR system called AR-Hip.17 This technology does not utilize any preoperative or intraoperative imaging. It instead identifies the functional pelvic plane based on anatomic landmarks. A less commonly utilized AR approach involves the display of preoperative imaging on top of the patient's anatomy, viewed on an external monitor outside the surgical field.12 The ideal AR approach would be a combination of these two technologies: a method of integrating preoperative imaging while directly viewing the operative field. The technology demonstrated here allowed us to achieve that goal. This study utilized the Microsoft HoloLens headset to overlay 3D reconstructions of preoperative CT scans onto the patient intraoperatively. Thus, this represents a new approach to the integration of AR into THA.
Notable drawbacks to this AR technology include high upfront costs, increased setup time for equipment, and decreased intraoperative mobility due to line-of-sight requirements between sensors and devices. Additionally, in the case of CT-based navigation, this method may introduce unnecessary radiation to patients and is susceptible to changes in patient pelvis positioning intraoperatively.10,12 However, the potential benefits of improved stability and better joint alignment through AR-guided THA may be sufficient to outweigh these factors. We hypothesized that this AR system would more accurately place the acetabular cup for THA relative to manual approaches utilizing intraoperative fluoroscopic assistance.
1.2 Objective
Therefore, the objective of this study was to evaluate whether there are perioperative differences in case duration, anesthesia, operative time between cases which utilized the AR headset-based system and those that did not. Additionally, we sought to answer whether there is radiographic evidence of superior acetabular cup placement accuracy using this AR headset-based system compared to conventional approaches.
2 Methods
2.1 Study design
After obtaining approval from our institutional review board, we retrospectively reviewed 125 THAs who underwent a direct anterior THA using AR (AR THA) and 118 patients who underwent a direct anterior THA with the use of fluoroscopic assistance (FA THA) alone between March 30, 2021, and December 30, 2022. All THAs were performed using a direct anterior approach by a single fellowship-trained arthroplasty surgeon. Medical records were reviewed for demographics, intraoperative details, immediate postoperative details, radiographic acetabular cup placement, and radiographic leg length discrepancy. All patients were followed for a minimum of 1 year postoperatively to assess for any complications and reoperations, including the need for revision surgery. Patients were excluded for BMI less than 20 kg/m2 (n = 12). Patients were then matched on age, gender, ASA score, and BMI, leaving 115 THAs in the AR cohort and 115 THAs in the FA cohort in the final study population.
2.2 Participants
Patients were similar across age, gender, height, laterality, and indication for surgery (Table 1). Between the THA with AR group and manual THA group, BMI (AR THA mean 27.2 (range, 20.0 to 39.9) versus FA THA mean 28.4 (range, 20.7 to 37.6), p = 0.02) and weight (AR THA mean 76.5 kg (range, 47.0 to 108.0) versus FA THA (mean 80.8 (range, 50.3 to 122.3), p = 0.03) were statistically different, though this difference is likely not clinically significant.
| Augmented Reality n = 115 | Fluoroscopic Assistance n = 115 | Total n = 230 | p-value | |
| Age (years), mean (range) | 69.4 (43–88) | 68.3 (45–88) | 68.9 (43–88) | p = 0.4 |
| Gender, n (%) | N/A | N/A | N/A | p = 0.2 |
| Women | 66 (57.4) | 76 (66.1) | 142 (61.7) | N/A |
| Men | 49 (42.6) | 39 (33.9) | 88 (38.3) | N/A |
| Height (cm), mean (range) | 166.9 (143.5–188.0) | 167.5 (150.5–193.0) | 167.2 (143.5–193.0) | p = 0.7 |
| Weight (kg), mean (range) | 76.5 (47.0–108.0) | 80.8 (50.3–122.3) | 78.6 (47.0–122.3) | p = 0.03 |
| BMIa (kg/m2), mean (range) | 27.2 (20.0–39.9) | 28.4 (20.7–37.6) | 27.8 (20.0–39.9) | p = 0.02 |
| ASAbscore, mean (range) | 2.2 (1–3) | 2.2 (2–3) | 2.2 (1–3) | p = 0.5 |
| Laterality, n (%) | N/A | N/A | N/A | p = 0.6 |
| Left | 46 (40.0) | 51 (44.3) | 97 (42.2) | N/A |
| Right | 69 (60.0) | 64 (55.7) | 133 (57.8) | N/A |
| Indication for THA, n (%) | N/A | N/A | N/A | p = 0.3 |
| Avascular Necrosis | 1 (0.9) | 0 (0.0) | 1 (0.4) | N/A |
| Femoral Neck Fracture | 1 (0.9) | 0 (0.0) | 1 (0.4) | N/A |
| Osteoarthritis | 112 (97.4) | 111 (96.5) | 223 (97.0) | N/A |
| Posttraumatic Arthritis | 1 (0.9) | 4 (3.5) | 5 (2.2) | N/A |
2.3 Variables
2.3.1 Radiographic evaluation
Radiographic analysis was performed using the most recent preoperative and postoperative standing anteroposterior (AP) radiographs for each patient. Anteversion and inclination were calculated using the technique described by Widmer.18 In accordance with author's surgical experience as well as both historical and recent evidence, goal anteversion and inclination were defined as 22° and 42°, respectively, for both the AR and FA cohorts.19–22 Leg length discrepancy was measured as the distance from the inferior aspects of the obturator foramina to the superior aspect of the lesser trochanter as described by Tsang et al.23
2.4 Statistical method
R (R Core Team, Vienna, Austria) was used for matching. THAs were matched on age, gender, ASA score, and BMI using 1:1 propensity score optimal matching. Python (Python Software Foundation, Beaverton, OR) was used for data analysis. Chi-square and independent-samples t-tests were used to compare cohorts. Significance was set at p < 0.05.
3 Results
3.1 Intraoperative results
Table 2 represents analysis on intraoperative results. Patients who underwent AR THA had decreased intraoperative radiation exposure 2.0 milligray (mGy) (range, 0.3–8.0 mGy) versus 3.2 mGy (range, 0.4–10.8 mGy), p < 0.001) and radiation time (mean 0.2 min (range, 0.0–0.5 min) versus 0.4 min (range, 0.1–0.8 min), p < 0.001) compared to those who underwent FA THA. There was no difference in anesthesia type, operative time, (mean 2.7 h (range, 2.2 to 3.4) versus 2.7 h (range, 2.0 to 3.5), p = 0.5), use of 1 or more cables (7.0 versus 7.0 %, p = 1.0), or estimated blood loss (mean 343.9 mL (range, 150–1200 mL) versus 342.6 mL (range, 50–1500 mL), p = 1.0).
| Augmented Reality n = 115 | Fluoroscopic Assistance n = 115 | Total n = 230 | p-value | |
| Anesthesia, n (%) | N/A | N/A | N/A | p = 0.7 |
| General | 6 (5.2) | 4 (3.5) | 10 (4.3) | N/A |
| Spinal | 109 (94.8) | 111 (96.5) | 220 (95.7) | N/A |
| Cumulative Radiation Dose (mGya), mean (range) | 2.0 (0.3–8.0) | 3.2 (0.4–10.8) | 2.6 (0.3–10.8) | p < 0.001 |
| Radiation Time (min), mean (range) | 0.2 (0.0–0.5) | 0.4 (0.1–0.8) | 0.3 (0.0–0.8) | p < 0.001 |
| Number of Cables, n (%) | N/A | N/A | N/A | p = 0.6 |
| 0 | 107 (93.0) | 107 (93.0) | 214 (93.0) | N/A |
| 1 | 6 (5.2) | 4 (3.5) | 10 (4.3) | N/A |
| 2 | 2 (1.7) | 4 (3.5) | 6 (2.6) | N/A |
| Operative Time (hours), mean (range) | 2.7 (2.2–3.4) | 2.7 (2.0–3.6) | 2.7 (2.0–3.6) | p = 0.5 |
| Estimated Blood Loss (mL), mean (range) | 343.0 (150–1200) | 342.6 (50–1500) | 342.8 (50–1500) | p = 1.0 |
3.2 Postoperative outcomes
Inpatient versus same day status was equivalent for AR THA (20.9 % inpatient, 79.1 % same day) and FA THA (24.3 % inpatient, 75.7 % same day). There was no significant difference between AR THA and FA THA in postoperative length of stay (mean 35.6 h (range, 5.0–146.3 h) versus 30.5 h (range, 3.1 to 121.8), p = 0.1), highest VAS pain score (mean 5.9 (range, 2 to 10) versus FA mean = 5.4 (range, 0 to 10), p = 0.1), or reoperation (AR THA 2 infection reoperations, FA THA 0 reoperations, p = 0.5) (Table 3.
| Augmented Reality n = 115 | Fluoroscopic Assistance n = 115 | Total n = 230 | p-value | |
| Inpatient versus Same Day Surgery, n (%) | N/A | N/A | N/A | p = 0.6 |
| Inpatient | 24 (20.9) | 28 (24.3) | 52 (22.6) | N/A |
| Same Day Surgery | 91 (79.1) | 87 (75.7) | 178 (77.4) | N/A |
| Postoperative Length of Stay (hours), mean (range) | 35.6 (5.0–146.3) | 30.5 (3.1–121.8) | 33.1 (3.1–146.3) | p = 0.1 |
| Highest VASaPain Score, mean (range) | 5.9 (2–10) | 5.4 (0–10) | 5.7 (0–10) | p = 0.1 |
| Reoperations, n (%) | N/A | N/A | N/A | p = 0.5 |
| Infection | 2 (1.7) | 0 (0.0) | 2 (0.9) | N/A |
| None | 113 (98.3) | 115 (100.0) | 228 (99.1) | N/A |
3.3 Radiographic outcomes
There was no significant difference between AR THA and FA THA in inclination deviation (mean 2.3° (range, −10.2 to 13.9°) versus 2.2° (range, −7.9 to 13.6°), p = 0.8) or anteversion deviation (mean 1.9° (range, −10.0 to 15.2°) versus mean 2.3° (range, −14.1 to 15.8°), p = 0.5) (Table 4). The same held true for both inclination deviation (Fig. 1) and anteversion deviation (Fig. 2) after groups were subdivided into 6 bins according to the following deviation ranges: <-10°, −10° to −5°, −5°–0°, 0°–5°, 5°–10°, >10°.
| Augmented Reality n = 115 | Fluoroscopic Assistance n = 115 | Total n = 230 | p-value | |
| Inclination, mean (range) | 44.3 (31.9–55.9) | 44.2 (34.1–55.6) | 44.2 (31.9–55.9) | p = 0.8 |
| Anteversion, mean (range) | 23.9 (12.0–37.2) | 24.3 (7.9–37.8) | 24.1 (7.9–37.8) | p = 0.5 |
| Leg Length Discrepancy compared to Preoperative, mean (range) | 4.6 (−5.6–22.0) | 3.4 (−9.1–14.1) | 4.0 (−9.1–22.0) | p = 0.1 |
| Leg Length Discrepancy compared to Contralateral Side, mean (range) | 3.1 (−9.6–14.5) | 2.2 (−9.9–19.0) | 2.6 (−9.9–19.0) | p = 0.2 |
| Deviation From Goal Inclination, mean (range) | 2.3 (−10.2–13.9) | 2.2 (−7.9–13.6) | 2.2 (−10.2–13.9) | p = 0.8 |
| Deviation From Goal Anteversion, mean (range) | 1.9 (−10.0–15.2) | 2.3 (−14.1–15.8) | 2.1 (−14.1–15.8) | p = 0.5 |


There was no significant difference between inclination for AR THA versus FA THA (mean 44.3° (range, 31.9 to 55.9 versus 44.2° (range, 34.1–55.6°), p = 0.8), anteversion (mean 23.9° (range, 12.0–37.2°) versus 24.3° (range, 7.9–37.8°), p = 0.5), mean preoperative-postoperative leg length discrepancy difference (mean 4.6 (range, −5.6 to 22.0) versus 3.4 (range, −9.1 to 14.1), p = 0.1), or leg length discrepancy compared to contralateral side (mean 3.1 (range, −9.6 to 14.5) versus 2.2 (range, −9.9 to 19.0), p = 0.2) (Table 4).
4 Discussion
4.1 Key results
In primary direct anterior THA, AR surgical navigation delivers similar results to manual THA with intraoperative fluoroscopic assistance. Radiographic anteversion and inclination, operative time, estimated blood loss, and reoperations were comparable between the two groups. However, AR surgical navigation is associated with decreased intraoperative fluoroscopic radiation and time.
Previous studies within this space are challenging to directly compare to this AR approach because of the heterogeneity of both conventional control approaches and the new technologies being investigated.10–16 Generally, AR-guided surgeries have had favorable outcomes with decreased errors in angles of anteversion or inclination.
While the AR approach outlined here has demonstrated similar results to FA THA, the potential of AR technologies extends far beyond final anteversion and inclination angles. For hip arthroplasty performed on an outpatient basis at an ambulatory surgical center (ASC), patients receive preoperative x-rays during their work up. While this has been the gold standard for diagnosis and surgical planning, the high turnover of staff (including x-ray technologists) and the expansion of cost-cutting measures have decreased the consistency of preoperative imaging quality. This AR methodology still requires preoperative imaging, but the use of a preoperative CT scan removes some variability in imaging quality and improves preoperative planning.
Although preoperative CT imaging is required for AR approaches, delivering more radiation to the patient, this was slightly offset by a decrease in intraoperative radiation, which also decreased radiation exposure to the operative staff.
We believe AR strategies can transform current practice standards due to their potential for cost savings relative to alternative technologies entering the OR, such as robotics. Robotic approaches offer opportunities to increase the accuracy and consistency of hip arthroplasty while reducing complications,24,25 but come with the downside of increased invasiveness and more drastic changes to the operative workflow. Additionally, as with any new device integrated into surgical procedures, operative staff must undergo additional training; there is also a learning curve before devices become second nature to these processes. However, AR approaches like the one described here can supplement traditional workflows with minimal additional training.
Furthermore, these robotic systems are accompanied by significant upfront costs paired with high disposable costs that can result in higher medical bills for patients.25 While a new robotic system, such as Mako, can carry with it an initial investment of upwards of $1,000,000 USD,26 an AR approach that can supplement existing approaches would likely be less than $10,000 USD when integrated into a finalized commercial product.21 While it is not uncommon to find devices in healthcare utilized successfully for decades, newer trends in both the commercial and private sector utilize a planned obsolescence approach. As such, a case could be made against integrating robotic approaches that might be supplanted by the next version of surgical robots, necessitating additional capital investments. Comparatively, AR approaches are more likely to have significant software improvements that might not necessitate full system replacements. Even still, fifty full system AR replacements would be less costly than a single robotic system. These new, less expensive technologies would also be considerably easier for smaller, financially constrained hospitals that are often based in rural or smaller community settings.
Lastly, the ability to superimpose a patient's skeletal anatomy on the surgical field has made surgeons re-think the traditional goal angles of anteversion. When planned visually using AR approaches, the surgeons carrying out these procedures consistently found that an anteversion angle of 22° was most appropriate for patients.
4.2 Limitations & strength
There are several limitations to this study that must be considered. This was a retrospective study, which inherently weakens the analysis. Additionally, the AR approach was compared to FA THA, whereas a case could be made that it should be compared to new operative technologies such as robotics. Another limitation was the relatively short follow-up; a longer follow-up period with a larger sample size is needed to elucidate any potential long-term discrepancies that may exist between FA and AR surgical approaches.
5 Conclusion
We demonstrate that primary direct THA performed with an AR navigation had similar deviation rates from ideal radiographic anteversion and inclination angles, operative times, and reoperation rates compared to manual approaches. There was decreased intraoperative radiation exposure while using the AR navigation. Our findings demonstrate that AR approaches can assist surgeons in performing THA with equal accuracy to manual approaches. The future of surgery is rapidly changing as new, sometimes costly, technologies enter the OR. These devices must not only be evaluated on clinical results, but also considered in the context of the value they may add to surgery centers.
CRediT authorship contribution statement
Gloria Coden: Data curation, Formal analysis, Investigation, Project administration, Resources, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing. Mikhail Kuznetsov: Investigation, Project administration, Supervision, Validation, Writing – original draft, Writing – review & editing. Lauren Schoeller: Investigation, Resources, Validation, Visualization, Writing – review & editing. James Patti: Investigation, Resources, Validation, Visualization, Writing – review & editing, Investigation, Resources, Validation, Visualization, Writing – review & editing. Andrew Grant: Formal analysis, Investigation, Project administration, Validation, Writing – original draft, Writing – review & editing. James Penn Miller: Investigation, Resources, Validation, Visualization, Writing – review & editing. Carl Talmo: Methodology, Conceptualization, Formal analysis, Investigation, Project administration, Resources, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing.
Ethical approval and patient consent
The work described has been carried out in accordance with The Code of Ethics of the World Medical Association (Declaration of Helsinki) for experiments involving humans. All procedures were performed in compliance 160 with relevant laws and institutional guidelines. This work was approved by our institutional review board.
Funding
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
References
- MakoTM robotic-arm-assisted total hip and total knee arthroplasty outcomes in an orthopedic oncology setting: a case series. J Orthop. 2023;46:70-77.
- [Google Scholar]
- The effectiveness of virtual reality, augmented reality, and mixed reality training in total hip arthroplasty: a systematic review and meta-analysis. J Orthop Surg. 2023;18(1):121.
- [Google Scholar]
- A novel 3D guidance system using augmented reality for percutaneous vertebroplasty: technical note. J Neurosurg Spine. 2013;19(4):492-501.
- [Google Scholar]
- Application of virtual and augmented reality technology in hip surgery: systematic review. J Med Internet Res. 2023;25
- [Google Scholar]
- Augmented reality in the operating room: a clinical feasibility study. BMC Muscoskel Disord. 2021;22(1):451.
- [Google Scholar]
- Accuracy of augmented reality-assisted pedicle screw placement: a systematic review. Eur Spine J. 2024;33(3):974-984.
- [Google Scholar]
- The learning curve for the direct anterior total hip arthroplasty: a systematic review. Int Orthop. 2021;45(8):1971-1982.
- [Google Scholar]
- What safe zone? The vast majority of dislocated THAs are within the lewinnek safe zone for acetabular component position. Clin Orthop. 2016;474(2):386-391.
- [Google Scholar]
- Improving cup positioning using a mechanical navigation instrument. Clin Orthop. 2011;469(2):423-428.
- [Google Scholar]
- Augmented reality-assisted versus conventional total hip arthroplasty: a systematic review and meta-analysis. J Orthop Surg. 2023;18(1):920.
- [Google Scholar]
- Accuracy comparison of cup positioning during total hip arthroplasty using Hip Align and AR-Hip in the supine position. Arch Orthop Trauma Surg. 2023;143(12):7229-7235.
- [Google Scholar]
- Accuracy of augmented reality with computed tomography-based navigation in total hip arthroplasty. J Orthop Surg. 2023;18(1):662.
- [Google Scholar]
- Augmented reality- vs accelerometer-based portable navigation system to improve the accuracy of acetabular cup placement during total hip arthroplasty in the lateral decubitus position. J Arthroplasty. 2022;37(3):488-494.
- [Google Scholar]
- Does a commercially available augmented reality–based portable hip navigation system improve cup positioning during THA compared with the conventional technique? A randomized controlled study. Clin Orthop. 2024;482(3):458-467.
- [Google Scholar]
- Does an augmented reality-based portable navigation system improve the accuracy of acetabular component orientation during THA? A randomized controlled trial. Clin Orthop. 2020;478(5):935-943.
- [Google Scholar]
- Does augmented reality-based portable navigation improve the accuracy of cup placement in THA compared with accelerometer-based portable navigation? A randomized controlled trial. Clin Orthop. 2023;481(8):1515-1523.
- [Google Scholar]
- A pilot study of augmented reality technology applied to the acetabular cup placement during total hip arthroplasty. J Arthroplasty. 2018;33(6):1833-1837.
- [Google Scholar]
- A simplified method to determine acetabular cup anteversion from plain radiographs. J Arthroplasty. 2004;19(3):387-390.
- [Google Scholar]
- Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60(2):217-220.
- [Google Scholar]
- The John Charnley Award: risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital. Clin Orthop. 2011;469(2):319-329.
- [Google Scholar]
- Redefining the 3D topography of the acetabular safe zone: a multivariable study evaluating prosthetic hip stability. J Bone Joint Surg Am. 2022;104(3):239-245.
- [Google Scholar]
- Acetabular cup positioning in primary routine total hip arthroplasty—a review of current concepts and technologies. Arthroplasty. 2023;5:59.
- [Google Scholar]
- The intraoperative pelvic radiograph during total hip arthroplasty: is it reliable to estimate leg length? Arthroplasty Today. 2022;16:9-14.
- [Google Scholar]
- The current role of robotics in total hip arthroplasty. EFORT Open Rev. 2019;4(11):618-625.
- [Google Scholar]
- Lower intraoperative and immediate postoperative complications in robotic versus conventional primary total hip arthroplasty: a retrospective cohort analysis of over 360,000 patients. Cureus 2024
- [Google Scholar]
- Comparative assessment of current robotic-assisted systems in primary total knee arthroplasty. Bone Jt Open. 2023;4(1):13-18.
- [Google Scholar]