The effect of c-arm tilt on accuracy of intraoperative fluoroscopy in assessing acetabular component position during direct anterior approach for hip arthroplasty
Article History
-
Received: ,
Accepted: ,
Abstract
Abstract
Intraoperative fluoroscopy can improve the precision of acetabular component positioning during direct anterior hip arthroplasty. However, changes in pelvic tilt with c-arm positioning may compromise accuracy. A geometric model of an acetabular component’s projection onto fluoroscopy images was created and manipulated to simulate c-arm tilt. An apparently ideally-placed cup will have its effective anteversion changed by 7.5° with just 10° of caudal/cephalad tilt of the c-arm. This effect is greater in cups that are more horizontally placed or less anteverted. Accurate c-arm positioning is crucial, as small errors in tilt can have considerable effects on final cup position.
Keywords
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
1 Introduction
The direct anterior approach to the hip is an increasingly common technique in total hip arthroplasty. Early clinical results suggest that the direct anterior approach results in shorter length of stay, improved early postoperative pain and function, and more rapid weaning of walking aids. There is controversy over the incidence of postoperative instability and other complications in anterior versus posterior approach, but a large-scale meta-analysis did not demonstrate clear superiority of either approach.1
One of the advantages of using a direct anterior approach is that it may be performed in a supine position, facilitating real-time fluoroscopic assessment of the position of the acetabular component during reaming and implant placement. The use of fluoroscopy can offer a degree of precision in terms of cup positioning that is unavailable to the surgeon relying on physical landmarks.2
One underappreciated pitfall with intraoperative fluoroscopy is how much the tilt of the c-arm or pelvis in the sagittal plane can affect the perceived positioning of the acetabular component. A fluoroscope that is vertically oriented with respect to the operating room floor may not correspond with a true standing anteroposterior (AP) view of the pelvis, and in reality be more of an inlet or outlet view. Subtle intraoperative errors in positioning of the patient, the operating room table, or the c-arm may have large effects on the eventual functional component position. A better understanding of how c-arm positioning affects cup position would help surgeons recognize and prevent these issues.
The purpose of this study was to derive a mathematical model of how the degree of c-arm tilt in a sagittal plane can create a mismatch between the true position of a cup and its radiographic appearance on fluoroscopy, and translate this derivation to a clinically relevant correlative model. A better understanding of this effect should help surgeons to recognize when they are in danger of malpositioning the acetabular component.
2 Methods
Models were constructed of an acetabular component in a 3-dimensional space in relation to a patient supine on an operating room table (Fig. 1). The true amount of anteversion and inclination of the cup were defined as AT and IT, respectively. A normal vector n orthogonal to the acetabular component (i.e. the direction of the reamer) may then be defined in terms of true inclination and true anteversion:(1)Let n= xyz=cosATsinITcosATcosITsinATWhen a c-arm is malpositioned by a certain degree Ɵ, a new normal vector n’ is created:(2)n′=x'y'z'=Axyz=1000cosθsinθ0−sinθcosθxyz= xycosθ+zsinθzcosθ−ysinθ=x'y'z'

The values of the distances x’, y’, and z’ define the degree of perceived anteversion and inclination (AP and IP) that is apparent through the (now-malpositioned) c-arm.(3)x'y'z'=cosAPsinIPcosAPcosIPsinAP(4)x'=x= cosATsin(IT)y'=ycosθ+zsinθ=cosATcosITcosθ+sinATsinθz′=zcosθ−ysinθ=sinATcosθ−cosATcosITsinθ
Combining Formulas (3) and (4) above yields:(5)x′= cosAPsinIPy′= cosATcosITcosθ+sinATsinθ= cosAPcosIPz'=sinATcosθ−cosATcosITsinθ =sinAP
Using standard trigonometric identity formulas, the equations were derived for perceived anteversion and inclination caused by c-arm error of θ degrees for any amount of true anteversion and inclination:(6)AP=arcsinsinATcosθ−cosATcosITsinθ(7)IP=arccotcotITcosθ+tanATsinITsinθ
This equation is consistent with a previously published derivation examining a similar problem (resting cup position as a function of pelvic tilt).3 Representative values produced by this formula on a cup positioned in 40° of inclination and 15° of anteversion are presented in Table 1.
| Perceived Cup Anteversion (AP) | Perceived Cup Inclination (IP) | Relative C-arm tilt (Ɵ) | True Cup Anteversion (AT) | True Cup Inclination (IT) |
| 15 | 40 | 20 | −0.6 | 38.4 |
| 15 | 40 | 15 | 3.4 | 38.5 |
| 15 | 40 | 10 | 7.3 | 38.7 |
| 15 | 40 | 5 | 11.1 | 39.3 |
| 15 | 40 | 0 | 15.0 | 40.0 |
| 15 | 40 | −5 | 18.8 | 41.0 |
| 15 | 40 | −10 | 22.5 | 42.2 |
| 15 | 40 | −15 | 26.2 | 43.8 |
| 15 | 40 | −20 | 29.8 | 45.7 |
| 15 | 30 | 10 | 6.3 | 29.1 |
| 15 | 30 | −10 | 23.6 | 31.8 |
| 5 | 40 | 10 | −3.7 | 29.9 |
| 5 | 40 | −10 | 13.6 | 30.8 |
3 Discussion
At our institution, it is routine to begin the fluoroscopic assessment of the patient with an AP image centered on the pelvis. This is followed immediately by an adjustment of c-arm sagittal tilt to match the patient’s preoperative standing pelvic radiograph. This step ensures that the apparent position of the cup on intraoperative fluoroscopy will match the eventual position of the cup when the patient stands, rather than assuming that the c-arm being vertical while the patient is supine represents a true AP.
Our model demonstrates that this step of matching the tilt of intraoperative fluoroscopy to preop x-ray is one that merits careful attention. Assuming an acetabular component is ideally positioned in 40° inclination and 15° anteversion, having the c-arm differ from preoperative radiographs by only 10° in a caudal/cephalad direction will result in a change in anteversion of approximately 7.5° (true anteversion of 7.3 or 22.5°, respectively) (Table 1). The inclination only changes by ≈2° with this amount of c-arm tilt. This effect becomes more pronounced in more horizontally placed components; those with 30° of inclination will have their anteversion changed by almost 9° for every 10° of c-arm tilt. The same holds true for less anteverted cups.
An understanding of the geometric relationship between c-arm position and the location of the pelvis in space can also be used to the surgeon’s advantage when dealing with patients with abnormal pelvic tilt, which can affect functional component position.4 Patients with abnormal sagittal balance in the lumbar spine, or those status post spinal fusion surgeries, may have obligate pelvic malpositioning when on the operating room table. The understanding of how abnormal spinopelvic parameters affects cup positioning is rapidly expanding.5 The simple, fast, and cost-effective step of adjusting c-arm tilt has the potential to mitigate some of the effects of abnormal pelvic tilt or spinal fusion on cup positioning.
A heightened awareness of the importance of c-arm tilt, and an understanding of why it has such a significant effect on final cup positioning, may serve as an important tool in the armamentarium of a surgeon performing direct anterior hip arthroplasty to ensure more consistent outcomes and reduced complication rates. This rule of thumb of “10° c-arm tilt equals 7.5° anteversion change” should be a useful guide in accounting for differences between preoperative and intraoperative imaging.
Conflicts of interest
Dr. Cooper reports and paid consultant for Corin, KCI, and Zimmer-Biomet; however, the submitted work describes a technique/presents a mathematical model, does not mention or endorse any specific products.
Dr. Geller reports and paid consultant for Smith & Nephew and Biosensor; however, the submitted work describes a technique/presents a mathematical model, does not mention or endorse any specific products.
Dr. Jang has nothing to disclose.
Dr. Lin has nothing to disclose.
Dr. Shah reports and paid consultant for Link Bio; however, the submitted work describes a technique/presents a mathematical model, does not mention or endorse any specific products.
References
- Anterior vs. posterior approach for total hip arthroplasty, a systematic review and meta-analysis. J Arthroplasty. 2015;30(3):419-434.
- [Google Scholar]
- Does fluoroscopy with anterior hip arthoplasty decrease acetabular cup variability compared with a nonguided posterior approach? Clin Orthop Relat Res. 2014;472(6):1877-1885.
- [Google Scholar]
- Pelvic tilt makes acetabular cup navigation inaccurate. Acta Orthop. 2005;76(4):517-523.
- [Google Scholar]
- Pelvic tilt in patients undergoing total hip arthroplasty: when does it matter? J Arthroplasty. 2015;30(3):387-391.
- [Google Scholar]
- Acetabular anteversion changes due to spinal deformity correction: bridging the gap between hip and spine surgeons. J Bone Jt Surg Am. 2015;97(23):1913-1920.
- [Google Scholar]
