Short-term results of metaphyseal sleeves in total knee arthroplasty for severe bone defects of the tibia
Article History
-
Received: ,
Accepted: ,
Abstract
Abstract
Severe bone defects pose a major challenge in total knee arthroplasty (TKA); however, metaphyseal sleeves may be a viable option to treat Anderson Orthopedic Research Institute (AORI) classification Type II bone defects. This study reports the short-term results of metaphyseal sleeves for severe bone defects in the proximal tibia.
From 2013 to 2019, a total of 10 cases were treated at our hospital with metaphyseal sleeve implants (ATTUNE revision system; DePuy-Synthes MBT, Warsaw, IN, USA), of which 2 cases underwent primary TKA for severe varus deformity, 5 underwent revision, and 3 underwent re-revision. The mean age was 71.2 (57–81) years, and the mean observation period was 5.3 years.
There were 5 cases of AORI Type IIa and 5 cases of Type IIb. The mean implant survival time was approximately 4.7 (3.0–6.2) years. There was one case that underwent revision surgery. One patient underwent arthrodesis after implant removal due to infection. The Knee Society score (KSS) was 75 (60–95) points preoperatively and 120 (90–145) points 12-months postoperatively. Excluding 2 cases that underwent revision surgery, no progression of the radiolucent zone, implant migration, or fracture was observed up to the final follow-up of at least 24-months postoperatively.
Metaphyseal sleeves can potentially replace more invasive procedures if the severe defect has little effect on the medial collateral ligament or patellar tendon attachment site with no observable mediolateral instability.
Keywords
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
1 Introduction
A careful assessment and management of bone defects in the proximal tibia are important during primary total knee arthroplasty (TKA) of severe valgus or varus knee deformity and subsequent revision or re-revision surgery. The Anderson Orthopedic Research Institute (AORI) classification is often used to evaluate bone defect.1 Implant selection and joint reconstruction remain difficult in AORI Type II and III cases with extensive and severe bone defects. For severe bone defects in the tibia, previous reports have described the use of impaction bone graft, allograft, and metal augmentation; however, bone grafting carries the risk of nonunion, bone resorption, and collapse of the grafted bone,2–4 and tibial tray metal augmentation is limited to a defect of up to approximately 20 mm in depth.5 Although rotating hinge and segmental (distal femoral or tibial replacement prostheses) implant systems are also potential treatment options, there are concerns about the possibility of long-term loosening, surgical invasiveness, and large amount of bone resection.6
Porous-coated metaphyseal sleeves are designed to fill bone defects and facilitate osseointegration. Although metaphyseal fixation with porous metal cones for severe bone loss management has demonstrated promising early results,7,8 there are few reports that focus on their use in severe bone defects and the degree to which the technique can be safely used.9–11 The purpose of this study is to evaluate short-term results and clinical efficacy of porous-coated metaphyseal sleeves for severe bone defects. We hypothesized that porous-coated metaphyseal sleeves may be a viable option for managing severe bone defects.
2 Patients and methods
Patients underwent tibial metaphyseal sleeve implantation during primary surgery of high-grade varus deformity and revision TKA between January 2013 to December 2019 at our hospital were included in this study. Written informed consent was obtained from all patients, and the study protocol was conducted in accordance with the principles of the Declaration of Helsinki and approved by the institutional review board of [blinded for review]. A cementless metaphyseal sleeve (ATTUNE revision system; DePuy-Synthes, Warsaw, IN, USA) was used. There were 2 cases that underwent primary TKA for severe varus deformity, 5 that underwent revision, and 3 that underwent re-revision. The mean age was 71.2 (57–81) years, of which 4 were men and 6 were women. The mean follow-up period was 4.7 (3.0–6.2) years. Preoperative classifications of the bone defects are listed in Table 1. In terms of the surgical technique, a metaphyseal sleeve was used to a depth and size that allowed the reamer and broach to be press-fit into the bone. If cortical bone remained around the sleeve with presence of a defect in the cancellous bone, the defect was filled with grafting of resected bone, and if the cortical bone was defected, it was left as-is with no cement used around the sleeve. A nonhinged constrained mobile bearing design (Sigma® TC3 system) was used in all cases, of which 6 patients used a cemented stem and 4 patients used a sleeve alone without a stem. One patient also required a medial tibial augment in conjunction with the sleeve.
| AORI type | N |
| Type ⅡA | 5 |
| Type ⅡB | 5 |
The following items were considered for evaluation of clinical outcomes: 1) a breakdown of the AORI classification of each case, 2) intraoperative findings and the degree of bone loss on postoperative radiographs, 3) mean implant survival time, 4) appearance and progression of a radiolucent zone of 2 mm or more on postoperative radiographs, 5) presence of implant migration of 2 mm or more, and pre- and post-operative Knee Society Scores (KSS). The severity of the bone defect was evaluated based on intraoperative findings, and the bone coverage ratio (%) was defined as the ratio of the maximum diameter of the bone defect in each direction to the mediolateral (M/L) and anteroposterior (A/P) diameter and depth of the proximal sleeve (Fig. 1). Measurements were performed visually during surgery, and the area around the sleeve that was not covered by cortical or cancellous bone was measured with a tape measure. The diameter was calculated based on the diameter of the sleeve size that was used. In addition, the gap (radiolucent zone) or contact area between the sleeve and cancellous/cortical bone were assessed according to a measurement technique described by Borus et al.9 and Gerad et al.10 The measurements were obtained using postoperative anteroposterior and mediolateral radiographs of the four surfaces (anterior, posterior, medial, and lateral) where the porous sleeve was positioned. If there was no gap (radiolucent zone), the sleeve was considered to have complete contact. The total contact surface area was scored on a scale from 0 to 4 points. If resected bone was grafted to the cancellous bone defect during surgery, the grafted site was also considered to have contact. Complete contact over the entire surface was scored as 4 points, with the number of complete contact surfaces representing the corresponding score. A score of 0 was assigned if there was no complete contact on any surface.9,10 Because the bone coverage ratio was evaluated intraoperatively, a bone coverage ratio of 100 % was still determined to have no contact if there was no clear zone or coverage by cortical bone on the contact surface of the sleeve on postoperative radiographs. Complete full-surface contact was scored as 4 points, and the number of surfaces in full contact was scored as a corresponding point. If there was no complete contact on any surface, it was scored as 0 points (Fig. 2). The radiolucent zone was evaluated using KSS12 using anteroposterior and lateral radiographs obtained at 3, 6, 12, and 24 months after surgery and at final follow-up (Fig. 3). Clinical outcomes were evaluated using the new Knee Society knee scoring system13 (KSS: 180 points) before surgery and 24 months after surgery. To evaluate lateral instability, preoperative plain anteroposterior radiographs were taken at 0° and 30° valgus stress. In all cases, we confirmed that there was no lateral instability by verifying that there was no difference in opening angles between the healthy side and the 0° side nor difference in opening angles of 2° or more at 30°. In addition, the lateral stability was checked during surgery to confirm that there were no problems.



3 Results
The clinical outcomes, degree of bone defects, and bone coverage ratios of the sleeve, are summarized in Tables 2–4. The M/L bone coverage ratio of the sleeve was 85.0 % (75–100) for the anterior side and 94.0 % (90–100) for the posterior side, and the A/P bone coverage ratio of the sleeve was 80.5 % (75–100) for the medial side, 97.0 % (95–100) for the lateral side, and 91.5 % (85–100) for the depth (Table 5). The score value of absolute radiographic contact between the metaphyseal sleeve and cortices was a mean of 3.2 (2–4), with a score of 0 or 1 in 0 cases, 2 in 3 cases (both exhibiting posterior and lateral contact), 3 in 2 cases (both exhibiting posterior and mediolateral contact), and 4 in 5 cases (Table 5). The mean implant survival was approximately 4.7 years (3.0–6.2). Revision surgery was performed in one case due to infection, and no loosening was observed after the second revision surgery. KSS was 75 points (60–95) preoperatively and 120 points (90–145) 12-months postoperatively. Excluding 1 case that underwent revision surgery, radiolucent zones that showed 2 mm or more on radiographs obtained 3 months after surgery were as follows: A/P view, zone 1 and 2 in 3 cases, zone 3 and 4 in 1 case; M/L view, zone 1 in 3 cases. A large number of cases with radiolucent zones were found in the anteromedial area just below the tibial tray. No subsequent progression of the radiolucent zone, implant migration, or fracture was observed until the final follow-up, which was at least 24 months after the surgery.
| View | Ratio (%) | |
| M/L | Anterior | 85.0 (75–100) |
| Posterior | 94.0 (90–100) | |
| A/P | Medial | 80.5 (75–100) |
| Lateral | 97.0 (95–100) | |
| Depth | 91.5 (85–100) |
| Cases | Age | AORI type | Bone coverage ratio of sleeve (%) | Score value of absolute radiographic contact | ||||
| M/L anterior | M/L posterior | A/P medial | A/P lateral | Depth | ||||
| 1 | 65 | ⅡA | 90 | 100 | 90 | 100 | 100 | 4 |
| 2 | 66 | ⅡA | 90 | 95 | 85 | 95 | 90 | 4 |
| 3 | 69 | ⅡA | 95 | 100 | 80 | 95 | 95 | 4 |
| 4 | 57 | ⅡA | 80 | 90 | 75 | 95 | 95 | 4 |
| 5 | 73 | ⅡA | 100 | 100 | 100 | 100 | 100 | 4 |
| 6 | 76 | ⅡB | 88 | 100 | 75 | 100 | 90 | 2 |
| 7 | 77 | ⅡB | 75 | 80 | 75 | 95 | 80 | 2 |
| 8 | 81 | ⅡB | 75 | 90 | 75 | 100 | 85 | 2 |
| 9 | 75 | ⅡB | 75 | 90 | 75 | 95 | 90 | 3 |
| 10 | 73 | ⅡB | 80 | 90 | 75 | 95 | 90 | 3 |
| Cases | Radiolucent zone | KSS (Pre-operation) | KSS (Post-operation) | ImplantSurvival time (year) | |
| A/P | M/L | ||||
| 1 | – | – | 60 | 90 | 3.3 |
| 2 | 1.2 | – | 75 | 95 | 4.2 |
| 3 | – | 1 | 70 | 120 | 5.5 |
| 4 | 3.4 | – | 70 | 135 | 3.8 |
| 5 | – | – | 80 | 135 | 6.2 |
| 6 | 1.2 | – | 95 | 145 | 5.5 |
| 7 | 1.2 | – | 65 | 135 | 5 |
| 8 | – | – | 75 | 120 | 3 |
| 9 | – | – | 85 | 110 | 3.9 |
| 10 | – | – | 75 | 120 | 4.3 |
| Score | Cases (N) |
| 0, 1 | 0 |
| 2 | 3 |
| 3 | 2 |
| 4 | 5 |
3.1 Case presentation
A 73-year-old man underwent primary TKA at our hospital 2 years prior to visiting our hospital for follow-up. A radiolucent zone that surrounded the entire tibial component was observed 2 years after primary TKA. His joint fluid culture, blood tests for C-reactive protein (CRP), and erythrocyte sedimentation rate (ESR) were negative. Bone scintigraphy showed no abnormal accumulation on the femoral side, leading to a diagnosis of aseptic loosening on the tibia side (Fig. 4-A to D). A non-stem M.B.T. revision tibial tray (DePuy-Synthes, Warsaw, IN, USA) with 15 mm thickness and size 29 porous sleeve was used. The contact between the tibia tray and the osteotomized surface was good, and fixation in Zone 1 was also determined to be good. In Zone 2, the use of a sleeve alone was determined to be sufficient for fixation, and no stem was used. Intraoperative culture was negative, and no findings suggestive of infection were found in the histopathological tissue. During revision surgery, the intraoperative bone coverage of sleeve was 100 % for M/L, A/P, and depth, and the score value of absolute radiographic contact between the metaphyseal sleeve and cortices was 4 (Fig. 5A–B).


Although no bacteria were detected in the synovial fluid that could cause loosening after revision surgery, coagulase negative staphylococcus (CNS) was detected in two sets of arterial blood cultures that suggested the possibility of a mild infection and loosening caused by infection. Three years after revision surgery, implant loosening was observed in both the femur and tibia. As at the time of revision, joint fluid culture and blood tests for CRP and ESR were negative. However, since CNS was detected in the arterial blood culture, it was determined that the loosening was due to a mild infection (Fig. 6-A-B). A single-stage re-revision surgery was performed. The cortical defect of the bone on the anteromedial side of the tibia partially reached the diaphysis and was diagnosed as AORI Type Ⅱ B. Preoperative plain anteroposterior radiographs revealed no lateral instability, and no lateral instability was observed intraoperatively. However, a bone defect was observed at the medial collateral ligament attachment site; therefore, a semi-constrained non-hinged design (Sigma® TC3 system; DePuy-Synthes, Warsaw, IN, USA) was used as a precautionary measure. The femur was fitted with a 30 mm long cement stem, and the tibia was size 37 porous sleeve and 30 mm long cement stem. A rotating platform polyethylene component was used as an insert. The intraoperative M/L bone coverage of the sleeve was 88 % for the anterior side and 100 % for the posterior side, and the A/P bone coverage of the sleeve was 75 % for the medial side, 100 % for the lateral side, and 90 % for depth. The score value of absolute radiographic contact between the metaphyseal sleeve and cortices was 2 on the lateral and posterior sides (Fig. 6C–D). Five years after re-revision, no loosening was observed. Osteointegration was observed around the sleeve, exhibiting radiographic osseous in-growth with well-fixed metaphyseal sleeves (Fig. 7A–B).


4 Discussion
The principle of zonal fixation is important for implant fixation in TKA. Strong fixation in at least two zones within Zones 1–3 is required.14 In cases with severe tibial bone defects, fixation in Zone 1 is often difficult; therefore, fixation methods in Zones 2 and 3 are important. Because fixation of Zone 3 can often be achieved by using stems, the fixation of Zone 2 can be difficult. In such cases, if the bone is osteotomized down to the bone defect and augmentation is used, there is a concern that the procedure may affect the medial collateral ligament attachment site. Moreover, performing bone grafting alone would require a large area, and fixation may also be compromised.
The Metaphyseal sleeve uses a simple technique of reaming and broaching,11 and its tapered design provides stable fixation and osseous ingrowth (osseous integration) into the porous area.10 In addition, it has been reported that even in cases of peripheral defect, the sleeve occupies a large area within the medullary canal. The strong fixation achieved with this coverage allows a severe defect to be converted to a mild defect.11 A previous study found that the minimum required cortical coverage by the sleeve is 70–75 %.9,11 Another study reported that no effect on postoperative pain was observed between a group achieving complete contact of the sleeve with the bone cortex on four sides (medial, lateral, anterior, posterior) and a group where there is complete contact with one side alone.10 However, one study found that one case of complete contact on one side alone had caused loosening and that complete contact on at least two sides is desirable.9 In this study, a minimum of two-sided contact was observed, and bone coverage of the sleeve was at least 74 % anterior to the ML, 88 % medial to the AP, and no slack was observed up to 79 % depth in three different cases. There have been reports that firm fixation can be obtained with the metaphyseal sleeve even in AORI type 3 cases.15 However, since type 3 cases often involve bone defects at the collateral ligament attachment site, it was not included in this study. In place of metaphyseal sleeves, hinge-type or constrained-type models were used in these cases.
In terms of the radiolucent zone, previous reports have stated that there are no problems in those that non-progressive with the zone directly below the tibial tray,15 and that even when using a sleeve, a load is generated directly under the tray in the defected area and bone contact area of the sleeve.16 The most common areas of bone defects in this study were also located directly under the anteromedial tray and were non-progressive bone defects, and we believe that these cases are likely non-problematic. Regarding the use of stems, there are reports that it is acceptable practice to use sleeves alone17,18 and reports that it is better to use sleeves in conjunction with stems.19–21 In a study using a 3D computer-aided design model of a metaphyseal cone, it was reported that a stem is not necessary if the bone defect is less than 30 mm deep.22
In this study, in three cases where the tibia tray could be placed almost entirely on the osteotomized tibial surface due to a deep osteotomy, a stem was not used as fixation in Zone 1 and was believed to ensure a strong fixation. Regarding the length of the stem used and whether cement was used, there was no significant difference in stem length in terms of postoperative loosening or pain; however, there are reports that recommend the use of a cementless stem with a diameter that can be press-fit. In addition, it has been reported that there was no significant difference in implant survival rate between using cemented and cementless stems.23 In this study, there were many cases of osteoporosis, rheumatism, and diabetes. Due to the risk of fracture in a press-fit cementless stem and those with strong tibial internal curvature in these cases, we determined that press-fit fixation would be difficult. Therefore, cemented stems were used in all cases. Complications of metaphyseal sleeves include fractures during reaming and broaching, and rigid fixation that make implant removal difficult when revision is required in cases without implant loosening.24 Fortunately, no complications were observed in this study. Limitations of this study include the small number of cases, short follow-up period, lack of comparison with other organ types, and retrospective investigation.
5 Conclusions
We reported the short-term treatment results of metaphyseal sleeves for extensive bone defects in the proximal tibia. Even if there is a partial cortical defect of the bone in Zone Ⅱand the AORI classification is Type Ⅱ, the use of a metaphyseal sleeve is a viable treatment option if there is no lateral instability, the sleeve can be press-fit, and stable fixation can be achieved.
CRediT authorship contribution statement
Fumiyoshi Kawashima: Conceptualization, Methodology, Investigation, Data curation, Writing – original draft, Writing – review & editing. Hiroshi Takagi: Supervision, Writing – review & editing.
Guardian/patient's consent
Written informed consent was obtained from all patients.
Ethics statement
The study protocol was conducted in accordance with the principles of the Declaration of Helsinki and approved by the institutional review board of Showa University Fujigaoka Hospital (approval number: 2023-317-A; approval date: March 13, 2023).
Funding statement
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
References
- Augments and allografts in revision total knee arthroplasty. J Arthroplast. 2005;20:35-41.
- [Google Scholar]
- Limitations of structural allograft in revision total knee arthroplasty. Clin Orthop Relat Res. 2009;467:818-824.
- [Google Scholar]
- The structural allograft composite in revision total knee arthroplasty. J Arthroplasty. . 2002;17:90-93.
- [Google Scholar]
- Tibial tray with modular metal wedges for tibial bone stock deficiency. Clin Orthop Relat Res. 1989;248:71-79.
- [Google Scholar]
- The kinematic rotation hinge prosthesis for complex knee arthroplasty. Clin Orthop Relat Res. 2001;392:283-291.
- [Google Scholar]
- Porous metal metaphyseal cones for severe bone loss: when only metal will do. Bone Joint Lett J. 2014;96B:118-121.
- [Google Scholar]
- Porous tantalum metaphyseal cones for severe tibial bone loss in revision knee arthroplasty. a five to nine-year followup. J Bone Joint Surg Am. 2015;97(3):216-223.
- [Google Scholar]
- Cementless metaphyseal sleeves used for large tibial defects in revision total knee arthroplasty. J Arthroplast. 2013;28:604-607.
- [Google Scholar]
- Direct, cementless, metaphyseal fixation in knee revision arthroplasty with sleeves-short-term results. J Arthroplast. 2015;30(12):2256-2259.
- [Google Scholar]
- Development of a modern knee society radiographic evaluation system and methodology for total knee arthroplasty. J Arthroplast. 2015;30:2311e4.
- [Google Scholar]
- Zonal fixation in revision total knee arthroplasty. Bone Joint Lett J 2015:147-149.
- [Google Scholar]
- Use of stepped porous titanium metaphyseal sleeves for tibial defects in revision total knee arthroplasty: short term results. J Arthroplast. 2014;29(6):1219-1224.
- [Google Scholar]
- Metaphyseal cones in revision total knee arthroplasty: the role of stems. Bone Joint Res.. 2020;9(4):162-172.
- [Google Scholar]
- The use of metaphyseal sleeves in revision total knee arthroplasty. Knee. 2016;23(3):545-548.
- [Google Scholar]
- Cementless metaphyseal sleeve without stem in revision total knee arthroplasty. Arch Orthop Trauma Surg. 2016;136:1761-1766.
- [Google Scholar]
- Stability of three-dimensional printed custom-made metaphyseal cone for tibial bone defects reconstruction: a finite element analysis and biomechanical study. Orthop Surg. 2023;15(11):2937-2946.
- [Google Scholar]
- Influence of varying stem and metaphyseal sleeve size on the primary stability of cementless revision tibial trays used to reconstruct AORI IIA defects. A simulation study. J Orthop Res. 2018 Jul;36(7):1876-1886.
- [Google Scholar]
- Ast mp. Do metaphyseal cones and stems provide any biomechanical advantage for moderate contained tibial defects in revision TKA? A finite-element analysis based on a cadaver model. Clin Orthop Relat Res. 2021 Nov 1;479(11):2534-2546.
- [Google Scholar]
- Stability of three-dimensional printed custom-made metaphyseal cone for tibial bone defects reconstruction: a finite element analysis and biomechanical study. Orthop Surg. 2023 Nov;15(11):2937-2946.
- [Google Scholar]
- Survivorship of metaphyseal sleeves in revision total knee arthroplasty. J Arthroplast. 2017 May;32(5):1565-1570.
- [Google Scholar]
- Management of metaphyseal bone loss in revision knee arthroplasty. Acta Biomed. 2017 Jun 7;88(2S):98-111.
- [Google Scholar]