Robotic S2Alar iliac screw placement: Comparing conventional and robot assisted minimally invasive techniques
Article History
-
Received: ,
Accepted: ,
Abstract
Abstract -
To evaluate the clinical, imaging, and peri-operative results of robotic-guided conventional versus robot assisted minimally invasive techniques for S2 Alar-Iliac (S2AI) screw placement during spinopelvic fixation.
A prospective comparative study comparing the clinicoradiological parameters of patients undergoing conventional robotic assisted S2AI screw insertion and those who underwent Robotic minimally invasive percutaneous screw insertion between November 2023 and June 2025 was done. Data collected included demographics, operative time, intraoperative blood loss, postoperative pain, hospital stay, radiographic angles, radiation dose, screw accuracy, and complications.
A total of 102 patients underwent spinopelvic fixation during the study period - 56 patients (116 screws) underwent conventional robotic assisted S2AI screw insertion and 46 patients (92 screws) underwent robot assisted percutaneous S2AI screw insertion. Both groups showed accurate screw placement without any intraoperative adverse events. The robot assisted percutaneous group had markedly reduced postoperative VAS scores (3.2 ± 1.1 vs. 5.8 ± 1.5, p < 0.0001) and reduced length of stay (3.1 ± 0.9 vs. 5.0 ± 1.3 days, p < 0.0001). Intraoperative blood loss,wound complications and operative time were also lesser in the robot assisted group but was not statistically significant. All other radiographic parameters and radiation dose were comparable across groups.
Robot assisted minimally invasive S2AI screw placement is a safe and efficient alternative to the conventional open technique. It offers comparable radiographic accuracy and biomechanical stability, with the added benefits of reduced postoperative pain and shorter hospital stay.
Keywords
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
1 Introduction
Posterior spinopelvic fixation techniques are designed to address spinal instability resulting from a variety of lumbosacral and sacral disorders. Their primary purpose is to provide segmental stabilization and enhanced biomechanical support across the lumbosacral junction, thereby reducing the risk of failure in long-segment constructs. Traditionally, spinopelvic fixation involves the placement of iliac screws, either in the conventional form or as S2 alar-iliac (S2AI) screws.1
To mitigate the drawbacks associated with conventional iliac screws—particularly implant prominence and wound-related issues such as distal dehiscence—the S2AI technique was introduced.2 Despite these advantages, inserting S2AI screws requires subperiosteal stripping of the erector spinae and multifidus muscle attachments from both the medial and lateral aspects of the posterior sacrum. This maneuver can compromise the dorsal sacroiliac, iliolumbar, and sacrotuberous ligaments, which play a vital role in lumbopelvic stability. Disruption of these soft tissues may also lead to devascularization, heightening the risk of poor wound healing, deep hematoma formation, increased intraoperative bleeding, and delayed recovery. As a result, nearly one-quarter of patients may require early reoperations due to wound complications.3,4
Advances in surgical technology, including robotic navigation systems, now offer an alternative approach for inserting S2AI screws with minimal soft tissue disruption, while providing real-time imaging guidance to enhance accuracy and safety.5 Most existing literature on robotic-assisted S2AI screw placement focuses on open exposure followed by robotic insertion. Although robotic guidance improves precision, it still necessitates disruption of the surrounding stabilizing structures. In this context, we conducted a prospective study to compare the clinical and radiological outcomes of two robotic-assisted S2AI screw insertion techniques: the conventional open approach and a percutaneous method.
2 Materials and methods
2.1 Patient population
This prospective study was carried out at a single tertiary spine care center after receiving approval from the Institutional Review Board (ECR/34/Inst/KA/2013/RR-26). All consecutive patients who underwent spinopelvic fixations between November 2023 to June 2025 were enrolled in the study. We excluded patients less than 18 years of age and patients who underwent revision spinopelvic fixations (broken iliac or S2AI screws).
All surgical procedures were performed by the senior author (VS) who is a fellowship trained spine surgeon with over 20 years of experience. All screws were inserted using the MazorX Stealth Edition (MXSE) robotic platform (Medtronic Ltd., Dublin, Ireland).
At the beginning of robotic integration at our centre, patients initially underwent the conventional open robotic-guided technique for S2AI screw placement. After standardizing the workflow and gaining adequate experience, the surgical team transitioned to a minimally invasive robot-guided approach. The first 56 patients underwent lumbosacropelvic stabilization using the conventional method (116 S2AI screws), followed by 46 patients who received minimally invasive robot-guided S2AI screw placement (92 screws). All surgeries were performed Demographic data, including age, sex, and preoperative diagnosis, were recorded for all patients.
2.2 Conventional robotic technique -
All patients were operated under general anaesthesia and positioned prone over Allen® table. Mazor X Stealth edition (Medtronics Ltd., Dublin) was used for our cases along with O-arm™ (Medtronics Ltd., Dublin). Surgical field and the robot arm is draped in sterile fashion. “Scan and Plan” workflow was used in all our cases.5,6 The surgical method used followed the steps outlined by Srinivasa et al. in their publication.71.A midline posterior incision was made to expose the spine.2.The distal exposure was extended up to the interval between the S1 and S2 foramina, while the proximal exposure depended on the selected upper instrumented vertebra (UIV).3.After exposure, the MXSE robotic platform was positioned for a scan-and-plan workflow.4.A 40-cm field-of-view O-arm scan (Medtronic Ltd., Dublin, Ireland) was obtained to capture the greater sciatic notch and both femoral heads.5.On axial images, the screw path was planned to remain between the inner and outer cortices of the iliac crest and to obtain reliable purchase within the sacral ala.6.On the coronal view, the intended entry point was positioned between the S1 and S2 foramina, with the screw running just superior to the greater sciatic notch and the tip remaining intraosseous.7.The sagittal images were checked to confirm that the planned trajectory did not breach the greater sciatic notch.8.For screw placement, the robotic arm was aligned to the preset S2AI trajectory.9.A 3-mm high-speed burr was used to create the entry point for the S2AI screw.10.The tract was prepared sequentially using a 4-mm tap, followed by a 6.5-mm tap.11.Finally, an 8.5-mm diameter, 90-mm length S2AI screw was inserted along the planned path using a powered driver (see Fig. 1).
2.3 Percutaneous minimally invasive technique
For the minimally invasive technique, the same 'scan-and-plan' workflow was utilized, but the spine was not exposed. After positioning, painting, and draping, the MXSE was set up for the scan-and-plan workflow. Screw trajectory planning was done similar to the previously explained technique; however, meticulous care was taken to plan the skin incision for the S2AI screws in such a way that both screws could be placed through a single midline stab incision without stripping the midline attachments (Fig. 2). Once planned, both S2AI screws were then placed minimally invasively via a single midline stab incision, without any exposure.


After the procedure, the interconnecting rods are placed. In the minimally invasive technique, only the screw heads of S2AI screws were visualized for placing the rods in them without damaging the soft tissues. Wounds are closed in a subcuticular fashion.
Operative data and clinical results, including operative time, intraoperative blood loss, postoperative pain, and total length of hospital stay were recorded. Information regarding complications was also collected.
All screws were evaluated using postoperative O-arm imaging (Fig. 3). The average radiation dose was recorded, along with the time required for screw placement. The intended screw paths were then superimposed on the postoperative O-arm images to identify any deviations from the preplanned trajectories (Fig. 4).


2.4 Data analysis
Data analysis was carried out using SPSS version 20.0 (IBM, USA). Variables showing a normal distribution were expressed as mean ± standard deviation (SD), and comparisons between the two groups were made using an independent samples t-test. For categorical variables, chi-square, Wilcoxon signed-rank, and Fisher's exact tests were employed as appropriate. A p-value of <0.05 was taken to indicate statistical significance.
3 Results
A total of 102 patients (208 screws) underwent spinopelvic fixation between November 2023 and June 2025. Of these, 56 patients (116 screws) were treated using the conventional approach, while 46 patients (96 screws) underwent the minimally invasive technique. The mean age was 57.9 ± 14.6 years in the conventional group and 61.6 ± 12.0 years in the minimally invasive group (p = 0.342). A higher percentage of women were present in the conventional cohort (71.4 %) compared with the minimally invasive group (43.5 %).
Intraoperative blood loss was greater in the conventional group (539.13 ± 246.32 mL) than in the minimally invasive group (405.26 ± 229.67 mL), although this difference did not reach statistical significance (p = 0.08). Operating times were also slightly longer in the conventional group (244.70 ± 35.69 min) compared with the minimally invasive group (234.58 ± 25.43 min), but this variation was likewise not significant (p = 0.6).
Postoperative pain scores on day one were markedly higher in the conventional group (5.8 ± 1.5) than in the minimally invasive group (3.2 ± 1.1; p < 0.0001). The length of hospital stay also favored the minimally invasive approach, with patients staying an average of 3.1 ± 0.9 days, compared with 5.0 ± 1.3 days in the conventional group (p < 0.0001). A summary of demographic, intraoperative, and postoperative findings is provided in Table 1.
| Parameter | Conventional (n = 56, 116 screws) | Percutaneous Group (n = 46, 96 screws) | p-value |
| Mean Age (years) | 57.9 ± 14.6 | 61.6 ± 12.0 | 0.342 |
| Female Patients (%) | 71.4 % | 43.5 % | – |
| Estimated Blood Loss (mL) | 539.13 ± 246.32 | 405.26 ± 229.67 | 0.08 |
| Operating Time (min) | 244.70 ± 35.69 | 234.58 ± 25.43 | 0.6 |
| Post-op Pain (Day 1, VAS) | 5.8 ± 1.5 | 3.2 ± 1.1 | <0.0001 |
| Hospital Stay (days) | 5.0 ± 1.3 | 3.1 ± 0.9 | <0.0001 |
| Intraoperative Complications | None | None | – |
| Wound-related Problems | 2 | 1 | – |
Radiological assessment showed that medial angulation on axial imaging was similar between groups (37 ± 7° in the conventional group versus 35 ± 6° in the minimally invasive group; p = 0.713). Caudal angulation on sagittal images also demonstrated no significant difference (17 ± 8° vs. 18 ± 9°; p = 0.645). Mean radiation exposure was comparable as well (48.7 ± 13.2 mGy in the conventional group versus 46.8 ± 11.2 mGy in the minimally invasive group). Importantly, no deviation was noted between planned and actual screw trajectories in either group. Radiological and technical outcomes are detailed in Table 2.
| Parameter | Conventional Group | Percutaneous Group | p-value |
| Medial Angulation (°) | 37 ± 7 | 35 ± 6 | 0.713 |
| Caudal Angulation (°) | 17 ± 8 | 18 ± 9 | 0.645 |
| Radiation Dose (mGy) | 48.7 ± 13.2 | 46.8 ± 11.2 | – |
| Trajectory Deviation | None | None | – |
No intraoperative complications were encountered. Minor wound issues—superficial gaping with mild discharge—were observed in two patients in the conventional group and one patient in the minimally invasive group; all resolved with local wound care and an extended antibiotic regimen.
4 Discussion
Achieving rigid spinopelvic fixation has posed a longstanding challenge for deformity surgeons. Allen and Ferguson were among the first to directly address pelvic fixation through the Galveston technique, which incorporated the ilium into the fixation construct and served as a foundation for modern spinopelvic instrumentation methods.8 Although iliac screws evolved from the original Galveston method, they still require extensive sub-fascial exposure, making them prone to complications related to implant prominence.9 To counteract rotational forces while ensuring lumbosacropelvic stability, Kostuik et al. later introduced the use of trans-ilial bars connected to bilateral S1 screws.10 In contrast, the S2AI approach allows lumbopelvic fixation through a midline exposure with minimal subperiosteal dissection, reducing screw prominence and lowering the risk of wound-related issues.11 Additionally, S2AI screws align naturally with cephalad S1 screws, thereby eliminating the need for offset connectors. Although fluoroscopy can assist with precise placement of S2AI screws, it increases radiation exposure for both the surgical team and the patient and requires additional equipment and personnel. Robotic guidance, a more recent development, offers the potential to enhance screw placement accuracy while reducing operative duration and decreasing radiation exposure to the surgeon. Existing literature includes several small case series that highlight the benefits of robotic assistance for S2AI screw placement.12 In this study, we evaluate clinical, operative, and imaging outcomes between robotic-assisted conventional and minimally invasive approaches for lumbosacropelvic stabilization using S2AI screws in a cohort of 102 patients.
In our study although the conventional group showed a higher mean blood loss (539.13 ± 246.32 mL vs. 405.26 ± 229.67 mL) and slightly longer operative time (244.70 ± 35.69 min vs. 234.58 ± 25.43 min), the differences were not statistically noteworthy. This is consistent with findings by Kwan et al.13 who reported increased soft tissue dissection and haemostasis in open techniques, contributing to greater blood loss, without noteworthy improvement in surgical efficiency. Similarly, Wang et al.14 showed that while open approaches might facilitate exposure, minimally invasive techniques noteworthyly limit intraoperative bleeding.
The markedly reduced postoperative discomfort scores and shorter hospital stay in the minimally invasive group (3.2 ± 1.1 vs. 5.8 ± 1.5 on Day 1; and 3.1 ± 0.9 vs. 5.0 ± 1.3 days) are in line with the minimally invasive surgery (MIS) philosophy. Park et al.15 both found MIS techniques to be associated with less postoperative pain, earlier mobilization, and faster discharge. This supports the notion that minimally invasive screw placement reduces muscle disruption and inflammatory response, thereby enhancing recovery.
In a review of 35 S2AI screws placed in 18 patients, Hu and Lieberman reported no intraoperative complications and confirmed accurate placement of all screws.16 Vidyadhara et al., in a separate series involving 16 S2AI screws, also documented 100 % accuracy without major adverse events.7 Consistent with these reports, our study likewise demonstrated perfect accuracy for robotically placed S2AI screws in both conventional and minimally invasive approaches.
In their cadaveric investigation, Park et al. observed mean medial and caudal angulations of 32.0 ± 1.8° and 17.3 ± 5.4°, respectively,17 which showed slight variation compared with our measurements. In our cohort, the conventional group demonstrated average medial and caudal angles of 37 ± 7° and 17 ± 8°, while the minimally invasive group showed averages of 35 ± 6° and 18 ± 9°, respectively. Park and colleagues also reported 100 % accuracy in their series of eight screws, although this was based solely on surgeon assessment and plain radiographs. In contrast, our postoperative O-arm evaluation confirmed an overall accuracy rate of 100 % in both the conventional and minimally invasive groups.
Radiation dose was similar between groups (48.7 ± 13.2 mGy vs. 46.8 ± 11.2 mGy), suggesting that advancements in intraoperative imaging and navigation systems have mitigated the historically higher radiation burden associated with minimally invasive fluoroscopic guided techniques as described by O'Brein et al..18
5 Limitations and prospects
Our study was conducted at a single institution, involved a relatively limited patient population, and was performed by a single surgeon.Despite being the largest series of minimally invasively placed robotically instrumented S2AI screws in the literature, a larger sample size collected from multiple institutions and involving various surgeons would noteworthyly enhance the rigor of our assessment and improve external validity. While promising immediate post-operative benefits were observed, including reduced surgical time, lower intraoperative radiation exposure, and improved analgesic effects, longer-term studies are necessary to evaluate the durability of these results and overall efficacy. It is important to note that the absence of breaches or complications in this study does not negate the potential risks associated with the procedure. Future studies with extended follow-up, specifically those assessing long-term functional recovery, will be crucial to definitively confirm the sustained benefits and safety of this technique.
6 Conclusion
Robotic-assisted percutaneous S2 alar-iliac screw placement offers a reliable and minimally invasive option for achieving secure spinopelvic fixation. It provides equivalent radiographic precision and biomechanical stability compared to the open approach, with the added benefits of reduced postoperative pain and shorter hospital stays. This refined technique enhances surgical efficiency while minimizing soft-tissue trauma. Its adoption can further advance minimally invasive spinal deformity correction and fusion strategies.
Patient consent statement
The Institutional Ethics Committee of Manipal Hospital, Bangalore (Reference: ECR/34/Inst/KA/2013/RR-26) has WAIVED the requirement for individual patient informed consent for this retrospective study.
Ethical approval
This prospective, single-center study was approved by the Institutional Ethics Committee of Manipal Hospital, Bangalore (ECR/34/Inst/KA/2013/RR-26) on January 10th, 2024. All patients provided written informed consent before participation.
Author contributions
All authors have read and approved the final manuscript. Contributions were as follows:•Conceptualization: Vidyadhara S•Methodology: Vidyadhara S, Balamurugan T•Formal Analysis & Data Curation: Balamurugan T, Abhishek Soni, Dinesh Iyer•Surgical Procedures & Investigation: Vidyadhara S, Madhava Pai Kanhangad, Abhishek Soni•Writing – Original Draft: Balamurugan T•Writing – Review & Editing: Vidyadhara S, R Dinesh Iyer•Supervision & Project Administration: Vidyadhara S
Ethical statement
This study was approved by the Institutional Ethics Committee of Manipal Hospital, Bangalore.
Ethics Committee Reference: ECR/34/Inst/KA/2013/RR-26.
Study Classification: Retrospective observational study.
Funding declaration
None. This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
References
- Segmental spinal instrumentation for neuromuscular spinal deformity. Clin Orthop Relat Res 1989
- [Google Scholar]
- Complications associated with surgical stabilization of high-grade sacral fracture dislocations with spino-pelvic instability. Spine. 2006;31:S80-S88.
- [Google Scholar]
- Decompression and lumbopelvic fixation for sacral fracture-dislocations with spino-pelvic dissociation. J Orthop Trauma. 2006;20:447-457.
- [Google Scholar]
- Accuracy of S2 Alar-Iliac screw placement under robotic guidance. Spine Deform. 2018 Mar-Apr;6(2):130-136.
- [Google Scholar]
- Pelvic fixation in adult and pediatric spine surgery: historical perspective, indications, and techniques. J Bone Jt Surgery-Amer. 2015;97(18):1521-1528.
- [Google Scholar]
- Robotic-assisted spinopelvic fusions using S2 alar iliac screws. Ind Spine J. 2025 May-Aug;8(2):152-156.
- [Google Scholar]
- The galveston technique of pelvic fixation with L-rod instrumentation of the spine. Spine. 1984;9(4):388-394.
- [Google Scholar]
- Minimum 2-year analysis of sacropelvic fixation and L5--S1 fusion using S1 and iliac screws. Spine. 2001;26(18):1976-1983.
- [Google Scholar]
- Pelvic fixation in spine surgery. Historical overview, indications, biomechanical relevance, and current techniques. J Bone Joint Surg Am. 2005;87(suppl 2):89-106.
- [Google Scholar]
- Use of S2-alar-iliac screws associated with less complications than iliac screws in adult lumbosacropelvic fixation. Spine. 2017;42:E142-E149.
- [Google Scholar]
- Robotic guidance for S2-alar-iliac screws in spinal deformity correction. Clin Spine Surg. 2017;30(1):E49-E53.
- [Google Scholar]
- Comparison of percutaneous versus open techniques of S2AI screw placement: a systematic review and meta-analysis. Eur Spine J. 2019;28(8):1810-1820.
- [Google Scholar]
- Development of a minimally invasive technique for S2-alar-iliac screw placement with navigation: feasibility and accuracy. Neurosurg Focus. 2022;52(1):E6.
- [Google Scholar]
- Comparison of minimally invasive surgery and open surgery for lumbar fusion: a meta-analysis. World Neurosurg. 2017;98:273-284.
- [Google Scholar]
- Bone-mounted miniature robotic guidance for pedicle screw and translaminar facet screw placement: part I—technical development and a test case result. Neurosurgery. 2006;59(3):641-650.
- [Google Scholar]
- Free hand insertion technique of S2 sacral alar-iliac screws for spino-pelvic fixation: technical note, a cadaveric study. J Korean Neurosurg Soc. 2015;58(6):578-581.
- [Google Scholar]
- Comparison of radiation exposure during fluoroscopically guided percutaneous vs open freehand S2-alar-iliac screw placement. J Neurosurg Spine. 2019;30(5):574-580.
- [Google Scholar]