Low risk of periprosthetic fracture and subsidence with automated impaction device in arthroplasty following femoral neck fracture: A retrospective study
Article History
-
Received: ,
Accepted: ,
Abstract
Abstract
Manual broaching in hip arthroplasty for femoral neck fractures (FNFx) may introduce variability in force application, increasing the risk of periprosthetic fractures. Automated impaction devices (AID) deliver consistent, uniform force vectors during femoral preparation, potentially reducing implant-bone mismatch, stem subsidence, and fracture risk. This study evaluates early complication rates following AID use in hip arthroplasty for FNFx.
A retrospective cohort study was conducted on consecutive patients undergoing total hip arthroplasty (THA) or hemiarthroplasty (HA) for FNFx by two surgeons between January 1, 2019, and June 1, 2023. All patients received a cementless femoral stem implanted with an AID and had a minimum 30-day follow-up. Outcomes assessed included 30-day revision-free stem survivorship, intraoperative and postoperative periprosthetic fractures, femoral component subsidence, and 30-day reoperation and readmission rates.
The cohort included 118 patients [72.0 % women; mean age 77.4 (range, 52–95) years], with 82 (69.5 %) undergoing THA and 36 (30.5 %) HA. One intraoperative fracture (0.8 %, Vancouver AG) and one postoperative fracture (0.8 %, Vancouver B1) occurred. The 30-day reoperation rate was 3.4 % (n = 4), and the readmission rate was 16.3 % (n = 15). Mild femoral component subsidence occurred in 1.7 % (n = 2) without evidence of loosening. No femoral stem revisions were reported within 30 days, yielding a 100 % short-term survivorship rate.
Early outcomes suggest that AID use in hip arthroplasty for FNFx is associated with low rates of periprosthetic fracture, subsidence, and early complications. Further prospective studies are needed to assess long-term outcomes.
Keywords
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
1 Introduction
Hip arthroplasty is a well-established surgical intervention that significantly improves the quality of life in patients with debilitating hip conditions including osteoarthritis, avascular necrosis, and femoral neck fractures.1,2 Hip fragility fractures, particularly among the elderly, are largely secondary to decreased bone mineralization and remain a major public health concern.3–5 As the aging population continues to grow, the incidence of hip fractures is expected to rise, underscoring the critical need for effective surgical management strategies, including arthroplasty, to optimize patient outcomes and reduce the burden of hip fracture-related morbidity and mortality.3–7
The accurate placement and stability of the femoral stem are critical to the success of hip arthroplasty, ensuring proper load distribution, minimizing complications, and enhancing implant longevity.8–10 Precise femoral component sizing and coronal alignment is essential for reducing the risk of aseptic loosening, periprosthetic fractures, and malalignment, all of which can negatively impact patient outcomes and implant survivorship.7,8,11–17 Traditionally, femoral canal preparation in hip arthroplasty has been performed using manual broaching and impaction techniques, in which a handheld mallet drives a broach into the femoral canal to create an envelope for implant insertion.10,18,19
Automated impaction devices (AID) have recently been introduced to address the limitations of manual femoral canal preparation. These devices deliver consistent, unidirectional forces during broaching, reducing variability and potentially improving the accuracy and efficiency of femoral stem implantation.20 By standardizing the impaction process, AIDs aim to enhance reproducibility, reduce the physical strain on surgeons, and minimize the risk of periprosthetic fracture brought on by variability in forces applied to the femur.21,22 Early studies have demonstrated promising outcomes in elective total hip arthroplasty, suggesting that automated impaction improves femoral stem alignment, decreases operative times, and optimizes canal fill ratios without increasing complication rates.10,23 However, early outcome studies in patients with femoral neck fractures undergoing hip arthroplasty with the use of an AID are lacking.
We, therefore, asked the following questions regarding the use of an AID when performing THA or HA following FNFx: (1) What was the incidence of intraoperative femoral fracture, early postoperative femoral fracture, femoral subsidence, and loosening? (2) What was the incidence of early stem revision, reoperation, readmission, and complications?
2 Materials and methods
This retrospective local Institutional Review Board (IRB)-approved cohort study was performed using clinical data from a single institutional database. A consecutive series of 118 arthroplasty procedures performed for FNFx between January 1, 2019, and June 1, 2023, by two fellowship-trained surgeons was reviewed. The mean age was 77.4 years (±9.7, range, 52 to 95). There were 85 (72.0 %) women and 33 (28.0 %) men, and 113 patients (95.8 %) had an injury mechanism of ground-level fall (Table 1).
| n (%) unless otherwise specified | |
| Age in years (mean [SD, range]) | 77.4 (±9.7, range, 52 to 95) |
| Sex | |
| Women | 85 (72) |
| Men | 33 (28) |
| Mechanism of Injury | |
| Ground level fall | 113 (95.8) |
| Trauma | 3 (2.5) |
| Other | 2 (1.7) |
| ASA Classification | |
| Class ll | 20 (17.0) |
| Class lll | 80 (67.8) |
| Class lV | 18 (15.3) |
| Dorr Classification | |
| A | 35 (29.7) |
| B | 65 (55.1) |
| C | 10 (8.5) |
2.1 Data collection
Data collection included reviewing the electronic health records for each patient, including operative reports and clinic notes. In addition to demographic information (age, sex, American Society of Anesthesiologists (ASA) score), we recorded mechanism of injury, laterality, and time from the emergency department (ED) arrival to the operating room (OR). Surgical details including procedure type (THA or HA), estimated blood loss (EBL), operative duration, stem sizes, the use of wires or cables, and length of stay (LOS) were also recorded. Time to latest follow-up including clinical visits, radiographs, and/or phone-call responses was recorded (Table 2). Any complications within 30 days, including intraoperative and postoperative periprosthetic fractures, stem subsidence, radiographic loosening, dislocation, periprosthetic joint infection, readmission, and any subsequent reoperation on the same hip, were recorded.
| n (%) unless otherwise specified | |
| Time from ED to OR in days (mean [SD, range]) | 2.4 (±5.3, range, 0 to 41) |
| Mean Operative Time in minutes (mean [SD, range]) | 60.4 (±23.0, range, 27 to 182) |
| Laterality | |
| Right | 58 (49.2) |
| Left | 60 (50.8) |
| Procedure | |
| THA | 82 (69.5) |
| HA | 36 (30.5) |
| EBL in mL (mean [SD, range]) | 276 (±183, range, 25 to 1500) |
| Stem Size | |
| 2 | 1 (0.9) |
| 3 | 3 (2.5) |
| 4 | 9 (7.6) |
| 5 | 20 (17.0) |
| 6 | 22 (18.6) |
| 7 | 27 (22.9) |
| 8 | 17 (14.4) |
| 9 | 8 (6.8) |
| 10 | 8 (6.8) |
| 11 | 1 (0.9) |
| 12 | 1 (0.9) |
| Offset Stem Type | |
| Standard | 81 (68.6) |
| High | 37 (31.4) |
All procedures were performed in the supine position using a fluoroscopically-assisted direct anterior (DA) approach on a specialized radiolucent table with independently movable leg spars (Hana Mizuho OSI). Intraoperative fluoroscopy was used to assess femoral component size, leg length, femoral offset, and the presence of any intraoperative fractures. Femoral fractures were classified utilizing the Vancouver classification of postoperative periprosthetic femur fractures.24 All femoral broaching and stem placement were performed with an AID (KINCISE™, DePuySynthes, Warsaw, IN).21 All patients were implanted with a modern collared cementless stem (ACTIS™, DePuySynthes, Warsaw, IN). Patients who underwent HA received a bipolar femoral head component, while patients who underwent THA received a titanium acetabular shell with a highly crosslinked polyethylene liner (DePuySynthes, Warsaw, IN). No patients during the study period received dual mobility or cemented components.
2.2 Radiographic evaluation
Orthogonal radiographs, including AP pelvis and AP/cross-table lateral hip, obtained on presentation for initial injury were used to classify proximal femoral morphology using the Dorr classification system. Intraoperative fluoroscopy was used to verify appropriate femoral component size, leg length, femoral offset, and the presence of any intraoperative fractures. When possible, follow-up radiographs were conducted at two weeks, six weeks, three months, and 1-year to assess stem subsidence, loosening, lucency, and periprosthetic fracture. Stem subsidence was defined as a 3 mm or greater discrepancy between the intraoperative fluoroscopic image and the follow-up x-ray. Stem loosening was identified through changes in the stem's position within the proximal femur when comparing intraoperative images to the most recent follow-up radiograph.
2.3 Data analysis
Continuous variables were presented as means with standard deviations, while categorical variables were reported as counts and percentages. Univariate logistic regression analysis was conducted to examine the potential influence of variables such as surgical procedure, cement use, position of the femoral stem collar, Dorr classification, age, and sex on the incidence of fracture, 30-day stem revision, and 30-day reoperation. All statistical comparisons were performed using RStudio (Vienna, Austria).
3 Results
A total of 118 hips in 118 patients were found in the study parameters and included. The mean time from the emergency department to the operating room was 2.4 (±5.3, range, 0 to 41) days. There were 58 procedures (49.2 %) performed on the right hip and 60 procedures (50.8 %) performed on the left hip. THA accounted for 69.5 % (n = 82) of the procedures, while HA accounted for 30.5 % (n = 36). The mean estimated blood loss was 276 (±183, range, 25 to 1500) mL, and the mean operative time was 60.4 (±23.0, range, 27 to 182) minutes (Table 2).
3.1 Fracture
One (0.8 %) femoral fracture (Vancouver AG) occurred intraoperatively secondary to errant retractor placement prior to instrumentation, initially treated conservatively (Fig. 1). Postoperatively, one (0.8 %) patient sustained a Vancouver B1 fracture, treated with cable fixation (Fig. 2). Based on intraoperative fluoroscopy, the femoral stem collar was flush with the calcar in 88 (86.3 %) patients. The collar was proud 1–3 mm in 7 (6.9 %) patients and more than 3 mm proud in 7 (6.9 %) patients, indicating possible slight over-resection of the neck. Stem subsidence of greater than 3 mm was observed in 2 (1.7 %) patients in the first year. No stem loosening was observed in the first year (Table 3).


| n (%) unless otherwise specified | |
| LOS in days (mean [SD, range]) | 6.7 (±10.3 range, 0 to 93) |
| Follow-up Time in days (mean [SD, range]) | 128 (±157, range, 0 to 758) |
| X-ray/CT in days (mean [SD, range]) | 119 (±159, range, 0 to 758) |
| Intraoperative Fractures | |
| Shaft (Vancouver AG) | 1 (0.8) |
| Fracture within 30 days | |
| None | 117 (99.2) |
| Shaft (Vancouver B1) | 1 (0.8) |
| Collar Position (immediate postop) | |
| Flush | 88 (86.3) |
| >3 mm | 7 (6.9) |
| 1–3 mm | 7 (6.9) |
| Subsidence (>3 mm) | 2 (1.7) |
| Loosening | 0 (0) |
3.2 30-day outcomes
Of the 118 patients, 92 (78.0 %) patients had complete data for the first 30 days. Of those 92 patients, 0 (0 %) patients underwent stem revision in the first 30 days. Reoperation occurred in 4 (3.4 %) patients. One (0.8 %) patient underwent reoperation for a periprosthetic fracture, 1 (0.8 %) patient underwent reoperation for dislocation and a fracture of the greater trochanter, 1 (0.8 %) patient was treated for a periprosthetic joint infection, and 1 (0.8 %) patient was treated for an adverse suture reaction. Readmissions were recorded in 15 (16.3 %) patients (Table 4).
| n (%) unless otherwise specified | |
| Stem Survivorship (30 days) | 92 (100) |
| Reoperation (30 days) | 4 (3.4) |
| Readmission (30 days) | 15 (16.3) |
| Dislocation | 1 (1.1) |
| PJI | 1 (1.1) |
| Medical Complication | 20 (21.7) |
| Death | 4 (4.3) |
4 Discussion
AIDs are gaining popularity due in part to their potential to increase the accuracy and efficiency of femoral canal preparation during THA and HA. In this retrospective cohort study, aimed to investigate the safety and efficacy of using an AID in THA and HA following FNFxs, our results suggest a low risk of intraoperative and early postoperative fractures, femoral component subsidence, and all-cause reoperation related to the femoral component.
Our results indicate one intraoperative fracture (0.8 %) and one incidence of early postoperative periprosthetic fractures (0.8 %) within 30 days, demonstrating that preparing the femoral canal with an AID is both safe and effective. These findings are consistent with prior research indicating that AID use does not increase the risk of fractures compared to manual techniques. Low rates of periprosthetic femur fractures and cortical perforations have been reported with powered femoral broaching, suggesting that automated techniques can be safely implemented in hip arthroplasty without elevating fracture risk.23,25 In addition, automated impaction was not found to increase operative time.10,21–23 Thomason et al. found a decreased femoral broaching time with AID with no increase in fracture risk.26 Gordon et al. also recently found no increase in intraoperative calcar fracture risk when using automated impaction in THA, with fractures more likely during impaction of the final stem rather than during broaching, reinforcing the safety and efficiency of this approach.27
We found the incidence of femoral component subsidence to be 1.9 %. Thalody et al. reported that automated impaction optimized canal fill ratios, likely due to the ease of access to the center of the canal, allowing for a greater canal fill ratio.28 Optimized canal fill ratios have been shown to decrease the risk of early subsidence, aseptic loosening, and migration of the prosthesis.29,30 This may be explained by the fact that AIDs deliver consistent, unidirectional forces, enhancing the accuracy of femoral component placement and reducing the likelihood of subsidence. The reduced variability in force application may play a significant role in achieving the initial stem stability demonstrated here.
Our study observed a 30-day postoperative stem survival rate of 100 %, and no reported cases of femoral component loosening. The overall reoperation rate was 3.4 %, with causes unrelated to the femoral component's fixation, such as dislocation and infection. These findings align with those found by Bhimani et al., who demonstrated that automated impaction maintained femoral component accuracy while reducing operative times and physical workload on surgeons, likely contributing to the low observed complication and reoperation rates.31
The traditional manual impaction technique is physically demanding, requiring the surgeon to strike a 3–5 lb mallet approximately 300 times per day.32–34 This repetitive action can lead to significant surgeon fatigue and a higher risk of Repetitive Strain Injuries (RSI), with arthroplasty surgeons being highly affected.34,35 Importantly, a benefit of AID use is the significant reduction of physical strain on surgeons. Studies have shown that automated devices lead to less muscle fatigue and reduced physical exertion compared to manual techniques.28 For example, Ferrari et al. found that muscle fatigue in the upper limb was reduced with automated femoral broaching.36 Additionally, Vandeputte et al. reported that automated impaction improved ergonomic outcomes, reduced cortisol levels, and decreased mental and physical fatigue for surgeons.17,28 These ergonomic benefits are crucial for orthopaedic surgeons' long-term health and productivity by significantly reducing operative time while maintaining accuracy in femoral component placement. Thalody et al. reported that automated impaction decreased the physical workload on surgeons and improved the consistency of femoral canal preparation.10
5 Limitations
This study is not without limitations. First, the retrospective design inherently limits the ability to establish causality and may introduce bias. However, selection bias was minimized as all patients received the same cementless component using an AID, without the use of cement. Additionally, the relatively short follow-up period of 30 days is unlikely to capture all relevant complications. Unfortunately, obtaining adequate follow-up in this patient population remains a challenge, with significant attrition at later follow-up. To address this limitation, attempts were made to contact the patient and/or family. These procedures were all performed by two fellowship-trained arthroplasty surgeons at a tertiary care community hospital, which may ultimately limit the generalizability of these findings. Future studies should include multicenter participation, including various patient populations, surgical teams, and implants, with longer follow-up to assess the utility of AIDs from both a value perspective and surgeon health.
6 Conclusion
Our cohort of FNFxs treated with arthroplasty using an AID shows a low risk of intraoperative and postoperative fractures as well as overall complications. Thus, the AID offers optimized stem-bone contact and limited subsidence. Further prospective studies are warranted to confirm these benefits and explore additional long-term outcomes, including RSI and procedural efficiency.
CRediT authorship contribution statement
Courtney Levit: contributed to, Writing – original draft, Writing – review & editing, and, Formal analysis. Ryan Decook: contributed to, Investigation, Formal analysis, and, Writing – original draft. Joseph M. Schwab: contributed to, Software, Validation, Formal analysis, Writing – review & editing, Visualization, and, Project administration. Thomas L. Bradbury: contributed to, Writing – review & editing, Supervision, Validation, and, Project administration. Natalie L. Gresham: contributed to, Data curation, Writing – review & editing, and, Formal analysis. Anita “Alex” Bradham: contributed to, Data curation, and, Writing – review & editing. Farideh Najafi: contributed to, Conceptualization, Investigation. Brandon Naylor: contributed to, Conceptualization, Methodology, Validation, Resources, Supervision, and, Project administration.
Informed consent (patient/guardian), mandatory only for case reports/clinical images
Not applicable to our study.
Institutional ethical committee approval
This Research Project was approved by the Northside Hospital Institutional Review Board (IRB). Please feel free to reach out to the Central Research Department (CRD) at Northside Hospital for more information on this (IRB exempt) study.
Call the Northside Hospital Research Department at 404-303-3355 or email ClinicalTrials@northside.com.
Funding sources
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
References
- Hip fracture-related emergency department visits, hospitalizations and deaths by mechanism of injury among adults aged 65 and older, United States 2019. J Aging Health. 2023;35:345-355.
- [Google Scholar]
- Epidemiology and treatment of proximal femoral fractures in the elderly U.S. population. Sci Rep. 2023;13
- [Google Scholar]
- Nationwide analysis of femoral neck fractures in elderly patients: a receding tide. J Bone Jt Surg. 2017;99:1932-1940.
- [Google Scholar]
- Trends in neck of femur fracture incidence in EU15+ Countries from 1990-2017. Injury. 2023;54:645-651.
- [Google Scholar]
- Risk factors for periprosthetic fractures of the hip: a survivorship analysis. Clin Orthop. 2008;466:1652-1656.
- [Google Scholar]
- Risk factors for periprosthetic femoral fractures around total hip arthroplasty: a systematic review and meta-analysis. ANZ J Surg. 2020;90:441-447.
- [Google Scholar]
- Component alignment in hip resurfacing using computer navigation. Clin Orthop. 2009;467:917-922.
- [Google Scholar]
- Periprosthetic femoral fractures and trying to avoid them: what is the contribution of femoral component design to the increased risk of periprosthetic femoral fracture? Bone Jt J. 2017;99-B:50-59.
- [Google Scholar]
- Does automated impaction improve femoral component sizing and alignment in total hip arthroplasty? J Arthroplast. 2023;38:2154-2158.
- [Google Scholar]
- Diagnosis and detection of subtle aseptic loosening in total hip arthroplasty. J Arthroplast. 2022;37:1494-1500.
- [Google Scholar]
- Do we need intraoperative radiographs for positioning the femoral component in total hip arthroplasty? Arch Orthop Trauma Surg. 2014;134:727-733.
- [Google Scholar]
- Predicting revision risk for aseptic loosening of femoral components in total hip arthroplasty in individual patients—a finite element study. J Orthop Res. 2007;25:779-788.
- [Google Scholar]
- Does femoral morphology and stem alignment influence outcomes of cementless total hip arthroplasty with proximally coated double-tapered titanium stems? HIP Int. 2021;31:354-361.
- [Google Scholar]
- CT of the hip prosthesis: appearance of components, fixation, and complications. Radiographics. 2012;32:1089-1107.
- [Google Scholar]
- The role of the femoral component orientation on dislocations in THA: a systematic review. Arch Orthop Trauma Surg. 2022;142:1253-1264.
- [Google Scholar]
- A technique for accurate reproduction of the femoral anteversion during primary total hip arthroplasty. Arch Orthop Trauma Surg. 2001;121:343-345.
- [Google Scholar]
- The dimensional accuracy of preparation of femoral cavity in cementless total hip arthroplasty. J Zhejiang Univ - Sci. 2004;5:1270-1278.
- [Google Scholar]
- Variability in femoral preparation and implantation between surgeons using manual and powered impaction in total hip arthroplasty. Arthroplasty Today. 2022;14:14-21.
- [Google Scholar]
- Does powered femoral broaching compromise patient safety in total hip arthroplasty? Arthroplasty Today. 2023;23
- [Google Scholar]
- Low risk of fracture using a cementless triple-tapered collared femoral stem with automated impaction in direct anterior approach total hip arthroplasty. J Arthroplast. 2024;39:1001-1006.
- [Google Scholar]
- Use of an automated surgical impactor reduces femoral broaching time in direct anterior approach total hip arthroplasty: results from a randomized, multicenter study. Arthroplasty Today. 2024;28
- [Google Scholar]
- Mitigating calcar fracture risk with automated impaction during total hip arthroplasty. J Orthop. 2025;59:64-67.
- [Google Scholar]
- The effect of automated component impaction on the surgeon's ergonomics, fatigue and stress levels in total hip arthroplasty. J Orthop Exp Innov. 2023;4
- [Google Scholar]
- Influence of undersized cementless hip stems on primary stability and strain distribution. Arch Orthop Trauma Surg. 2017;137:1435-1441.
- [Google Scholar]
- Inadequate metadiaphyseal fill of a modern taper-wedge stem increases subsidence and risk of aseptic loosening: technique and distal canal fill matter. J Arthroplast. 2020;35:1868-1876.
- [Google Scholar]
- Surgical automation reduces operating time while maintaining accuracy for direct anterior total hip arthroplasty. J Orthop. 2020;22:68-72.
- [Google Scholar]
- Adult reconstructive surgery: a high-risk profession for work-related Injuries. J Arthroplast. 2016;31:1194-1198.
- [Google Scholar]
- Are we putting ourselves in danger? Occupational hazards and job safety for orthopaedic surgeons. J Orthop. 2021;24:96-101.
- [Google Scholar]
- Prevalence and spectrum of occupational injury among orthopaedic surgeons: a cross-sectional study. JB JS Open Access. 2023;8
- [Google Scholar]
- Surgical ergonomics and musculoskeletal pain in arthroplasty surgeons. J Arthroplast. 2021;36:3781-3787.
- [Google Scholar]
- The assessment of muscle fatigue in orthopedic surgeons, by comparing manual versus automated broaching in simulated total hip arthroplasty. Proc Inst Mech Eng [H]. 2021;235:1471-1478.
- [Google Scholar]