Translate this page into:

56 (); 1-5
doi:
10.1016/j.jor.2024.02.025

Gait analysis of a kinematic retaining implant for Total knee replacements during walking and running

University Hospital Dorset, Castle Ln E, Bournemouth, BH7 7DW, United Kingdom
Department of Design and Engineering, Faculty of Science and Technology, Bournemouth University, United Kingdom

⁎Corresponding author: Suraj Kohli. suraj.kohli@nhs.net

Disclaimer:
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.

Abstract

Abstract

The analysis of gait is an important tool for evaluating postoperative outcomes of a Total Knee Replacement (TKR). There are few studies which have evaluated the gait parameters of a Kinematic retaining (Kr) prosthesis. This study therefore aims to investigate the kinetic and kinematic differences of running and walking, in the ankle, knee and hip joints in patients who underwent a Kr TKR.

This study assessed the gait of 12 patients with physica lima Kr TKRs at 1 year follow up and 8 healthy controls using 3D video analysis. Data was collected on the kinetics and kinematics of walking and running at the ankle, knee and hip. Comparison was made between the operated and non-operated limbs of the patients, and between the operated and control limbs.

Gait analysis showed no statistically significant difference in the hip, ankle and knee angles or moments between the non-operated and operated legs during walking and running. However, there was a statistically significant difference between the knee angles of initial contact, maximum flexion during stance and swing in the TKR knees vs controls in walking and running. Similarly, there was also a statistically significantly higher max knee flexion moment between operated knees and controls in both walking and running.

This study has shown that a quadriceps avoidance gait persists in patients after TKR, and that there was symmetry and reciprocated gait parameters in non-operated limbs. These findings suggest that Kr TKRs could be capable of replicating normal knee kinematics when running and walking.

1

1 Introduction

Osteoarthritis (OA) is a degenerative condition causing varying degrees of pain and disability. The incidence of knee OA in UK adults is approximately one in ten, with a recent increase in cases among 35-44-year olds.1,2 A Total Knee replacement (TKR) is the definitive surgical management of end stage arthritis in the knee.3

TKRs relieve arthritic pain and can improve patients function in activities of daily living.4 The traditional mechanically aligned TKRs whether Cruciate Retaining (CR) or Posterior Sacrificing (PS) have similar outcomes in terms of range of motion, gait parameters, electromyographic waveforms, and isokinetic muscle testing parameters.5 Mechanically aligned PS and CR knees also generates similar values of maximum knee flexion angles and flexion moments during loading.6 However, literature on the assessment of gait parameters in a Kinematic retaining (Kr) prosthesis are sparse.7

Previous studies on gait analysis post TKR have focused solely on assessing walking.8–12

Some studies, although infrequent have assessed more challenging activities such as lunging, squatting and sit-to stand.13–16 There are also some studies which have assessed walking at an ‘elevate speed’,12,17,18 but to our knowledge there have been no studies performed looking at gait analysis of running post TKR.

This study therefore analysed the kinetic and kinematic differences of running and walking, in the ankle, knee and hip joints in patients who underwent a kinematic retaining TKR.

2

2 Materials & methods

This study involved assessing patients who had undergone TKRs using the Physica KR (LimaCorporate, Italy) knee implant. Regional ethical committee approval has been sought for this study (IRAS ID 172937, REC reference 15/SC/0725).

2.1

2.1 Participants

This study was prospective and recruited sixteen patients who had undergone unilateral TKRs using the Physica KR prosthesis and were at minimum 1-year post-op. Exclusion criteria were as follows: a) musculoskeletal or neurological disorders that may affect gait, b) previous surgery on the contra-lateral leg, c) back pain, and d) patients who were unable to perform the gait testing sessions unassisted. Eight healthy control patients who did not suffer from musculoskeletal disorders were also recruited.

2.2

2.2 Data acquisition

Three-dimensional video analysis was used to perform and assess gait, with ten cameras (Oxford Metrics, VICON) co-ordinated with 2 Kistler force platforms (Hampshire, Kistler Instrument Ltd) to assess the kinetic and kinematic parameters of gait.

Thirty-nine retro reflective spherical markers of 14 mm size were placed on landmarks following the VICON plug-in-gait guidelines. Force platform and marker data was performed at 2000 Hz and 250 Hz respectively.

The Helen- Hayes plug-in-gait model was followed for application of the markers to the patient. A calibration trial for both dynamic and static movements was performed and evaluated to ensure there were no blind spots.

2.3

2.3 Running and walking gait

Participants were asked to warm up initially for the walking gait analysis.

Subjects were then requested to ambulate in straight line, with three successful trials recorded. Subsequently, for the running analysis, participants were asked to trial and simulate laboratory environment running, before three successful trials of running were obtained.

2.4

2.4 Data processing and analysis

Three successful trials for walking and running were performed and recorded. Data was processed and assessed using Nexus 2.0. (VICON, Oxford Metrics, England).

The gait cycle was plotted, and time normalised to consecutive heel strikes of the same leg. Data was then evaluated and computed using MATLAB®, for the ankle, knee, and hip joints.

The comparisons performed were the following:1)TKR operated leg kinematics and kinetics vs non-operated leg2)TKR operated leg kinematics and kinetic vs control legs

3

3 Results

3.1

3.1 Patient demographics

Twelve patients with unilateral KrTKRs were evaluated at a minimum of one year follow up. The mean age of the patients was 66 ± 5 years, with an age range of 57–75. There were six females and six males, and the mean weight was 84 ± 9 kg and 1.72 ± 0.1 m.

3.2

3.2 Walking

There was no statistically significant difference between the non-operated and operated knees for the knee, hip or ankle angles (see Table 1). Additionally, there was no statistically significant difference between the moments of the ankle, knee or hip between the operated and non-operated knees (see Table 2).

Table 1 Table comparing operated knee, non-operated knee and control angles for the knee, hip and ankle joint during walking. * = statistically significant difference.
Operated knee (Mean ± SD) Non-operated knee (Mean ± SD) P value operated vs non-operated TKR Control Right/Left (Mean ± SD) P value Control vs operated
Knee Angles (Degrees)
Angle on initial contact 1.79 ± 3.90 2.66 ± 6.73 0.700 6.93 ± 6.99 0.031*
Maximum flexion during stance 10.85 ± 6.15 11.81 ± 9.56 0.771 19.54 ± 8.16 0.005*
Maximum extension during stance 4.02 ± 6.48 2.45 ± 6.99 0.571 5.17 ± 5.96 0.626
Maximum flexion during Swing 42.72 ± 7.43 44.46 ± 12.97 0.691 57.29 ± 7.68 <0.0005*
Maximum extension during swing 0.69 ± 3.99 3.21 ± 7.22 0.298 6.96 ± 1.74 0.156
Hip Angles (Degrees)
Angle on initial contact 28.75 ± 6.59 31.17 ± 4.77 0.335 33.39 ± 12.24 0.264
Maximum extension during stance −13.10 ± 6.57 −12.06 ± 5.89 0.701 −7.65 ± 9.43 0.110
Maximum flexion during swing 27.44 ± 6.93 29.48 ± 5.48 0.453 32.42 ± 10.29 0.175
Ankle Angles (Degrees)
Angle on initial contact 2.58 ± 2.72 2.93 ± 3.66 0.804 **
Maximum dorsiflexion during stance 14.40 ± 3.64 14.08 ± 2.67 0.816 **
Maximum plantarflexion at swing −10.63 ± 6.56 −10.89 ± 7.84 0.936 **
Maximum dorsiflexion during swing 6.31 ± 2.78 5.12 ± 1.95 0.261 **
Table 2 Table comparing operated knee, non-operated knee and control moments for the knee, hip and ankle joint during walking. * = statistically significant difference.
Operated knee (Mean ± SD) Non-operated knee (Mean ± SD) P value operated vs non-operated TKR Control Right/Left (Mean ± SD) P value Control vs operated
Knee Angles (Degrees)
Angle on initial contact 1.79 ± 3.90 2.66 ± 6.73 0.700 6.93 ± 6.99 0.031*
Maximum flexion during stance 10.85 ± 6.15 11.81 ± 9.56 0.771 19.54 ± 8.16 0.005*
Maximum extension during stance 4.02 ± 6.48 2.45 ± 6.99 0.571 5.17 ± 5.96 0.626
Maximum flexion during Swing 42.72 ± 7.43 44.46 ± 12.97 0.691 57.29 ± 7.68 <0.0005*
Maximum extension during swing 0.69 ± 3.99 3.21 ± 7.22 0.298 6.96 ± 1.74 0.156
Hip Angles (Degrees)
Angle on initial contact 28.75 ± 6.59 31.17 ± 4.77 0.335 33.39 ± 12.24 0.264
Maximum extension during stance −13.10 ± 6.57 −12.06 ± 5.89 0.701 −7.65 ± 9.43 0.110
Maximum flexion during swing 27.44 ± 6.93 29.48 ± 5.48 0.453 32.42 ± 10.29 0.175
Ankle Angles (Degrees)
Angle on initial contact 2.58 ± 2.72 2.93 ± 3.66 0.804 **
Maximum dorsiflexion during stance 14.40 ± 3.64 14.08 ± 2.67 0.816 **
Maximum plantarflexion at swing −10.63 ± 6.56 −10.89 ± 7.84 0.936 **
Maximum dorsiflexion during swing 6.31 ± 2.78 5.12 ± 1.95 0.261 **

Comparing the knee angles for the operated knee vs controls there was a statistically significant increase in the angle on initial contact 1.79 vs 6.93 (p = 0.031), maximum flexion during stance 10.85 vs 19.54 (p = 0.005), and maximum flexion during swing 42.72 vs 57.29 (p < 0.0005). There was no statistically significant difference between the operated knees and the control hip angles. Due to significant scatter and data capture issues, the control ankles values were not included in this analysis.

When comparing the moments of the operated knees and the controls, there was a statistically significant difference between the knee moment in extension of 902.76 vs 2046.76 (p = 0.023), and the hip extension moment −1412.21 vs −3255.82 (p = 0.011).

3.3

3.3 Running

The was no statistically significant difference between the operated and non-operated knees for the knee, hip or ankle running angles (see Table 3). Additionally, there was no statistically significant difference between the running moments of the hip, knee or ankle between the operated and non-operated knees (see Table 4).

Table 3 Table comparing operated knee, non-operated knee and control angles for the knee, hip and ankle joint during running. * = statistically significant difference.
Operated knee (Mean ± SD) Non-operated knee (Mean ± SD) P value* Op vs Nonop Control Right/Left (Mean ± SD) P value* Control vs Op
Knee Angles (Degrees)
Angle on initial contact 7.89 ± 4.86 9.41 ± 7.45 0.558 14.73 ± 4.85 0.011*
Maximum flexion during stance 22.92 ± 6.95 21.73 ± 10.38 0.756 33.98 ± 11.06 0.005*
Maximum extension during stance 13.08 ± 5.08 12.55 ± 8.06 0.849 15.88 ± 7.99 0.300
Maximum Flexion during Swing 61.88 ± 11.52 64.15 ± 7.92 0.579 75.48 ± 13.66 0.010*
Maximum extension during swing 7.31 ± 5.54 7.35 ± 8.00 0.988 8.07 ± 7.579 0.778
Hip Angles (Degrees)
Angle on initial contact 30.25 ± 5.94 32.94 ± 4.06 0.209 35.74 ± 14.36 0.225
Maximum extension during stance −4.62 ± 6.05 −2.36 ± 4.56 0.313 −0.68 ± 9.63 0.224
Maximum flexion during swing 35.24 ± 6.63 36.98 ± 6.02 0.507 42.86 ± 12.19 0.062
Ankle Angles (Degrees)
Angle on initial contact 6.26 ± 3.70 5.58 ± 6.59 0.758 −0.94 ± 17.01 0.163
Maximum dorsiflexion during stance 22.33 ± 4.47 21.13 ± 3.86 0.489 22.86 ± 6.03 0.803
Maximum plantarflexion during stance −9.68 ± 6.35 −13.27 ± 6.82 0.195 −16.89 ± 12.35 0.077
Maximum dorsiflexion during swing 8.95 ± 4.14 7.02 ± 4.62 0.291 6.77 ± 10.73 0.510
Table 4 Table comparing operated knee, non-operated knee and control running moments for the knee, hip and ankle joint during walking. * = statistically significant difference.
N.mm/kg TKR Op (Mean ± SD) TKR Non-Op (Mean ± SD) Pvalue operated vs nonoperated TKR Control (Mean ± SD) Pvalue Control vs operated
Ankle
Max Flexion 1529.43 ± 1053.88 1484.69 ± 1082.85 0.919 2022.27 ± 1796.53 0.448
Max Extension −352.91 ± 393.49 −198.81 ± 184.46 0.232 −846.05 ± 1178.67 0.192
Hip
Max Flexion 2815.23 ± 1317.34 2946.39 ± 1620.24 0.830 3636.32 ± 1893.70 0.266
Max Extension −1498.51 ± 759.43 −1185.95 ± 423.71 0.226 −2296.52 ± 1269.53 0.094
Knee
Max Flexion 942.78 n±769.35 838.31 ± 511.18 0.699 1884.67 ± 1163.89 0.042
Max Extension −1361.44 ± 703.02 −1595.51 ± 1423.96 0.615 −1902.68 1113.38 0.197
Peak adduction −104.37 ± 212.60 −55.43 ± 48.76 0.445 −264.37 ± 396.66 0.255
Peak abduction 350.67 ± 334.84 363.67 ± 483.76 0.940 331.81 ± 193.71 0.888

Comparing the knee angles for the running operated knee vs controls there was a statistically significant increase in the angle on initial contact 7.89 vs 14.73 (p = 0.011), maximum flexion during stance 22.92 vs 33.98 (p = 0.005), and maximum flexion during swing 61.88 vs 75.48 (p = 0.010). There was no statistically significant difference between the running angles of the operated knee, hip or ankle compared to controls.

When comparing the moments of the operated knees and the controls, there was a statistically significant difference between the knee moments in extension of 942.78 vs 1884.67 (p = 0.042) (see Table 4).

4

4 Discussion

Currently, TKRs are the modality of choice for end stage OA in the knee. However, one in five patients post TKR are unsatisfied with their outcomes, compared to one in twenty for Total Hip Replacements.19,20 TKRs have therefore been assessed radiologically, clinically, and mechanically to identify and address this discrepancy. Implant design and surgical technique have evolved as a result of this. The Physica KR prosthesis is designed with a concave medial articulating surface, and a saddle-like lateral surface to allow greater flexion. This study compared the kinematics and kinetics of a KrTKR limb to the non-operated and control limbs, at the ankle, knee, and hip. Additionally, this study assessed the gait during both walking and running for operated, non-operated, and control limbs.

4.1

4.1 Walking

Gait analysis in this study showed no statistically significant difference in the knee angles between the operated and non-operated legs during walking. However, there was a statistically significant difference between the knee angles of initial contact, max-flexion during stance, and max-flexion during swing in the operated knees vs controls. Similarly, there was also a statistically significantly higher max knee flexion moment between TKR knees and controls. The severity of knee OA has been associated with reduced flexion angles of the knee and show some improvement in patients post TKR.9 However, multiple studies have shown persistence of decreased knee flexion angles and peak flexion moments between TKR and control patients post operatively.21 This has been attributed to the development of a ‘quadriceps avoidance gait’, which is often present in patients with OA of the knee to reduce pain in the affected limb.9,21 A reduction in the force of the quadriceps in the opposition of the knee flexion moment results in reduced compression of the joint, which is advantageous in a painful knee.21 Although the painful stimulus has been removed post TKR, the persistence of these elements highlights the difficulties in changing a patient's gait pattern post procedure. Interestingly, the non-operated limb also had a reduced knee flexion angle and peak knee flexion moment, with no statistically significant differences seen between the operated and non-operated limbs. There was also a high level of symmetry between the operated and non-operated knees. The development of OA in the contralateral limb is well documented,22 with the incidence of bilateral knee OA and TKRs being relatively high.23 Given the symmetrical nature of gait, it is possible that the abnormal gait which develops in the OA knee is reciprocated in the other limb as a compensatory mechanism.21 This could provide an altered mechanical loading environment in contralateral limb joints and lead to the development of OA.

There was no statistically significant difference between peak adduction moments in TKR limbs vs non-operated limbs, or in TKR limbs vs controls. This suggests that there is similar medial loading of the KrTKR patients compared to controls and non-operated limbs. The knee adduction moment is used to assess loading of the medial compartment in gait.9 A higher knee adduction moment, has been associated with both the severity and progression of OA.24,25 This restoration of knee adduction moments similar to controls highlights the efficiency of the KR implant in maintaining low medial compartment forces.

There was no statistically significant difference in hip or ankle angles and moments in the operated vs non-operated legs. There was however a statistically significant difference between the max hip extension moment between TKR limbs and controls.

4.2

4.2 Running

This study is unique in its assessment of running for patients post TKR. Running is characterised by a decreased stance phase and an increased swing phase.26 There is often an increase in joint excursion, and an increase in eccentric muscle contraction.27 Running is not a common activity performed after a TKR, often due to multiple factors, including lack of motivation, medical comorbidities, and precautions over implant survival.

In this study, the symmetry of gait present with walking between the operated and non-operated limbs was also mirrored in running. There was no statistically significant difference in the knee, hip or ankle angles between the operated and non-operated legs during running. However, there was a statistically significant difference between the knee angles of initial contact, maximum flexion during stance, maximum flexion during swing, as well as the max knee moment in TKR limbs vs controls. This suggests that the quadriceps avoidance gait persists during running in patients who underwent Kr TKR.

Interestingly, even at heightened speed, the TKR group had similar knee adduction moments compared to both controls and non-operated limbs. This suggests that the levels of medial compartment stress during running in TKRs does not rise despite an increase in the forces acting across the knee. Given that high knee adduction moments have been attributed to enhanced poly wear, KrTKRs may be less susceptible to wear.28

Limitations of this study include patient numbers, as well as lack of gait analysis prior to TKR surgery. Additionally, the laboratory environment has inherent limitations with respect to assessing for normal walking and running. It must also be noted that this is an overall assessment, and there is likely to be some scatter at the individual level.

5

5 Conclusion

Overall, this study has shown that a quadriceps avoidance gait may persist in patients post TKR (in both walking and running) and that there is significant symmetry and reciprocated gait parameters in non-operated limbs. Additionally, compared to control subjects, the Kr TKR patients still had lower flexion knee angles in both running and walking. Similarly, the knee adduction moments for Kr TKRs are comparable to control subjects, suggesting that they have a sustained decrease in medial compartment loading, which could suggest that they have some resistance to wear. These findings suggest that Kr TKRs could be capable of replicating normal knee kinematics when running and walking.

Ethical statement

Regional ethical committee approval has been sought for this study (IRAS ID 172937, REC reference 15/SC/0725).

Funding statement

Funding and Conflict of interest: The materials and implant componentry used on all the patients were supplied by Lima corps under the trade name TK-8.

Guardian/parental consent

N/a.

CRediT authorship contribution statement

Adrian Harvey: Conceptualization, Writing – original draft. Suraj Kohli: Writing – original draft, Formal analysis. Samira Al-Nasser: Writing – original draft, Formal analysis. Siamak Noroozi: Conceptualization, Writing – review & editing, Supervision.

References

  1. , , , et al . Trends in incidence and prevalence of osteoarthritis in the United Kingdom: findings from the clinical practice research datalink (CPRD) Osteoarthritis Cartilage. 2020;28(6):792-801.
    [Google Scholar]
  2. , , , , . Annual consultation incidence of osteoarthritis estimated from population-based health care data in England. Rheumatology. 2015;54(11):2051-2060.
    [Google Scholar]
  3. , , , et al . A randomized, controlled trial of total knee replacement. N Engl J Med. 2015;373(17):1597-1606.
    [Google Scholar]
  4. , , , et al . Outcome following knee arthroplasty beyond 15 years. Knee. 2008;15(2):85-90.
    [Google Scholar]
  5. , , , et al . A comparison of isokinetic strength testing and gait analysis in patients with posterior cruciate-retaining and substituting knee arthroplasties. J Arthroplasty. 1998;13:906-915.
    [Google Scholar]
  6. , , , , , , . No difference in gait between posterior cruciate retention and the posterior stabilized design after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2014;22:3135-3141.
    [Google Scholar]
  7. , , , , , , . Kinematic alignment in total knee arthroplasty better reproduces normal gait than mechanical alignment. Knee Surg Sports Traumatol Arthrosc. 2019;27(5):1410-1417.
    [Google Scholar]
  8. , , . Kinematic analysis of human gait cycle. Al-Nahrain Journal for Engineering Sciences. 2014;16(2):208-222.
    [Google Scholar]
  9. , , , , . The effect of total knee arthroplasty on knee joint kinematics and kinetics during gait. J Arthroplasty. 2011;26(2):309-318.
    [Google Scholar]
  10. , , , . Gait after unilateral total knee arthroplasty: frontal plane analysis. J Orthop Res. 2011;29(5):647-652.
    [Google Scholar]
  11. , , , . Variability of gait in the early postoperative period of total knee arthroplasty with different surgical technique. Int Orthop. 2014;38(3):517-523.
    [Google Scholar]
  12. , , , , . Knee kinematics during walking at different speeds in people who have undergone total knee replacement. Knee. 2011;18(3):151-155.
    [Google Scholar]
  13. , , , . How to quantify knee function after total knee arthroplasty? Knee. 2008;15:390-395.
    [Google Scholar]
  14. , , . Swing phase kinetics and kinematics of knee replacement patients during obstacle avoidance. Gait Posture. 2003;18:95-104.
    [Google Scholar]
  15. , , . Altered loading during walking and sit-to-stand is affected by quadriceps weakness after total knee arthroplasty. J Orthop Res. 2005;23:1083-1090.
    [Google Scholar]
  16. , , , et al . Functional evaluation of the TKA patient using the coordination and variability of rising. J Electromyogr Kinesiol. 2007;17:49-56.
    [Google Scholar]
  17. , , , . Variability of walking and other daily activities in patients with total knee replacement. Gait Posture. 2009;30:288-295.
    [Google Scholar]
  18. , , , et al . The effect of walking speed on gait kinematics and kinetics after endoprosthetic knee replacement following bone tumor resection. Gait Posture. 2014;40:622-627.
    [Google Scholar]
  19. , , , et al . Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res. 2010;468:57-63.
    [Google Scholar]
  20. , , , et al . Analysis of factors influencing patient satisfaction after total hip arthroplasty in a Japanese cohort: the significant effect of postoperative physical activity. J Phys Ther Sci. 2022;34(2):76-84.
    [Google Scholar]
  21. , . Is gait normal after total knee arthroplasty? Systematic review of the literature. J Orthop Sci. 2009;14(1):114-120.
    [Google Scholar]
  22. , , , , . Nonrandom evolution of end-stage osteoarthritis of the lower limbs. Arthritis Rheum. 2002;46:3185-3189.
    [Google Scholar]
  23. , , , , . History of the contralateral knee after primary knee arthroplasty for osteoarthritis. Clin Orthop Relat Res (321):145-150.
    [Google Scholar]
  24. , , , et al . Knee adduction moment, serum hyaluronan level, and disease severity in medial tibiofemoral osteoarthritis. Arthritis Rheum. 1998;41:1233-1240.
    [Google Scholar]
  25. , , , , , , . Dynamic load at baseline can predict radiographic disease progression in medial compartment knee osteoarthritis. Ann Rheum Dis. 2002;61:617-622.
    [Google Scholar]
  26. , , . Biomechanics and analysis of running gait. Phys Med Rehabil Clin. 2005;16(3):603-621.
    [Google Scholar]
  27. , . The biomechanics of walking and running. Clin Sports Med. 1994;13(4):843-863.
    [Google Scholar]
  28. , , , , . Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty. J Bone Joint Surg Am. 2007;89:1306-1314.
    [Google Scholar]
Show Sections