Finite element analysis of three reconstruction strategies for AORI IIA tibial bone defects in TKA
Article History
-
Received: ,
Accepted: ,
Abstract
Abstract
AORI IIA-type bone defects are frequently encountered during total knee arthroplasty (TKA), posing a challenge for achieving durable and stable fixation. Although various reconstruction strategies have been proposed, the optimal method balancing mechanical stability and biological integration remains unclear.
This study aimed to evaluate and compare the biomechanical performance of three reconstruction techniques—locking plate fixation (LP Group), screw-assisted bone cement filling (SC Group), and metal block augmentation (MB Group)—for managing AORI IIA tibial defects using finite element analysis.
A finite element model of a medial AORI IIA tibial defect was constructed. Three reconstruction strategies were simulated, and the deformation and Von Mises stress distribution were analyzed under two loading conditions: (1) a vertical load of 800 N, and (2) a combined load of 600 N vertical force and 6000N・mm internal torsional moment.
Under vertical loading, the MB group showed the smallest maximum deformation of the graft material (0.299 mm) and the lowest maximum fixation stress (1.688 MPa), while the SC and LP groups exhibited similar maximum deformation (∼3.4 mm) with higher maximum internal stresses. Under combined loading, the MB group again had the lowest maximum deformation (0.227 mm) and fixation stress (1.153 MPa); the LP group showed the highest maximum fixation stress (62.841 MPa). The graft material in the MB and SC groups experienced higher maximum stress, suggesting improved load transfer, whereas the LP group showed stress shielding effects. Across both conditions, the MB group demonstrated the lowest residual stress in the cancellous bone.
The metal block technique offers superior biomechanical performance by combining structural stability with effective load sharing. The screw-assisted bone cement method provides moderate support and potential for bone integration, while the locking plate technique offers rigid initial fixation and preserves bone stock, it limits stress transmission. Selection of reconstruction strategy should be tailored to patient-specific bone quality, defect morphology, and functional demands.
Keywords
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
1 Introduction
Bone defect is a major challenge in total knee arthroplasty (TKA), most commonly involving the tibia, femur, or patella—particularly the posteromedial aspect of the tibial plateau.1 In patients undergoing TKA with tibial plateau defects, uncorrected lower limb malalignment and improper joint line reconstruction may lead to stress concentration around the prosthesis,2 compromising long-term implant stability and knee function.
In 1998, Engh and Parks proposed the Anderson Orthopaedic Research Institute (AORI) classification,3 which remains one of the most widely adopted systems for evaluating bone defects based on metaphyseal damage. AORI type IIA defects involve a single femoral condyle or one side of the tibial plateau, typically presenting with central or peripheral metaphyseal bone loss, joint line alteration, and intact collateral ligament attachments. Proper reconstruction of AORI IIA tibial defects is essential for ensuring prosthesis stability and reducing revision risk.
In this study, a three-dimensional finite element model simulating an AORI IIA tibial defect was developed. Three reconstruction techniques—locking plate fixation (LP group), screw-assisted bone cement filling (SC group), and metal block augmentation (MB group)—were evaluated under different loading conditions to compare their biomechanical performance. By analyzing stability, stress distribution, and deformation, this study aims to provide a theoretical basis for selecting appropriate reconstruction strategies for tibial bone defects.
2 Method
2.1 Model establishment
A male volunteer with normal knee joint development and no history of trauma was selected after providing informed consent. A 64-slice spiral CT scan was performed to acquire high-resolution images of the tibia and fibula. Using Mimics 17.0 software, both cancellous and cortical regions of the tibia were segmented and extracted. Surface smoothing and refinement were performed using Geomagic Studio 2013, and the final assembly was completed in Creo 7.0.
2.2 Finite element model of the tibial plateau defect
In Creo 7.0, a 10 mm wedge-shaped osteotomy was created on the medial tibial plateau to simulate an AORI type IIA defect. A standard TKA osteotomy (10 mm depth, 5° posterior slope) was also performed. A defect filler matching the osteotomy geometry was inserted to simulate intraoperative reconstruction.
Three reconstruction models were established based on different techniques:
Locking plate fixation: The defect was reconstructed using autologous cancellous bone grafts. A locking plate and six locking screws were applied laterally for fixation.
Screw-assisted bone cement filling: The defect was filled with bone cement reinforced by two vertically inserted cancellous screws, anchored into the underlying bone. The screw heads were aligned flush with the resected surface.
Metal block augmentation: A central metal block (8 mm diameter, 20 mm height) was used. Its upper surface aligned with the resected tibial plateau, and the lower portion was embedded into the cancellous bone. Remaining defect areas were filled with bone cement.
In all models, a cemented tibial stem extender (14 mm diameter, 100 mm length) was incorporated to enhance structural stability, as shown in Fig. 1.

2.3 Mesh generation and material properties
All models were meshed using ANSYS software. The number of nodes and elements for each model is summarized in Table 1. Mesh convergence testing confirmed result accuracy within a 5 % margin. All materials were modeled as homogeneous, isotropic, and linearly elastic. Elastic modulus and Poisson's ratio values were sourced from published literature,4–7 as shown in Table 2. A frictional contact (coefficient = 1.0) was applied between the tibial cut surface and the bone cement. All other contacts were set as fully bonded.8
| Model | Nodes | Elements |
| LP Group | 1771947 | 1139611 |
| SC Group | 2597266 | 1724648 |
| MB Group | 1005180 | 626443 |
| Material | Elastic Modulus (MPa) | Poisson's Ratio |
| Cortical Bone | 17000 | 0.30 |
| Cancellous Bone and autografted bone | 700 | 0.30 |
| Polyethylene Liner | 2300 | 0.25 |
| Tibial Prosthesis and Tibial Stem Extender | 110000 | 0.30 |
| Locking Screw-Plate | 190000 | 0.27 |
| cancellous screw | 190000 | 0.27 |
| Bone Cement Layer | 2270 | 0.46 |
| Metal Block | 117000 | 0.3 |
2.4 Boundary and loading conditions
Two loading conditions were applied to all three reconstruction models. In both scenarios, the cross-section of the tibial prosthesis was fixed.(1)Vertical loading condition: Under vertical loading, a downward force of 800 N was applied to the surface of the tibial component to simulate tibial loading during standing.(2)Combined torsional and vertical loading condition: Under combined torsional and vertical loading, a torque of 6000N・mm was applied along the central axis of the tibia, along with a vertical load of 600 N, to simulate the stress distribution during internal rotation and axial compression of the lower limb.
3 Result
3.1 Engineering condition 1: simulated vertical load (800 N)
Under a vertical load of 800 N, the maximum residual stress in the cancellous bone reached 0.888 MPa in the SC group and 0.884 MPa in the LP group, while the MB group showed the lowest value of 0.404 MPa. For the internal fixation components, the maximum stress was highest in the LP group (50.704 MPa), followed by the SC group (2.656 MPa), and lowest in the MB group (1.688 MPa). Regarding the graft material, the maximum stress was 1.744 MPa in the MB group, 1.063 MPa in the SC group, and 0.42 MPa in the LP group. As for maximum deformation of the graft region, the SC group exhibited the highest value (3.409 mm), followed closely by the LP group (3.382 mm), while the MB group showed the least deformation (0.299 mm), as shown in Fig. 2.

3.2 Engineering condition 2: simulated combined torsional and vertical loading condition (6000N・mm torque + 600 N axial load)
Under the combined load, the maximum residual stress in the cancellous bone was again lowest in the MB group (0.304 MPa), and higher in the LP and SC group (0.638 MPa and 0.618 MPa). The maximum internal fixation stress reached 62.841 MPa in the LP group, followed by 3.82 MPa in the SC group, and 1.153 MPa in the MB group. For the graft material, the SC group exhibited the highest maximum stress (1.217 MPa), followed by the MB group (1.158 MPa), and the LP group (0.469 MPa). In terms of maximum deformation, the SC group showed the highest value (2.591 mm), the LP group was nearly identical (2.586 mm), while the MB group again had the lowest deformation (0.227 mm), as shown in Fig. 3.

Overall, under both loading conditions, the MB group demonstrated the smallest maximum deformation of the graft material, lowest internal fixation stress, and most balanced stress distribution, indicating superior overall biomechanical performance. The SC group offered moderate load-sharing capacity but lower structural stability. In contrast, the LP group experienced pronounced internal fixation stress and stress shielding, while providing strong initial rigidity, as shown in Fig. 4.

4 Discussion
For AORI type IIA bone defects, commonly employed treatment strategies include adequate osteotomy with the use of thick polyethylene liners, bone cement filling technique, and metal block augmentation.9–11 Although there is currently no consensus in the academic literature regarding the clinical outcomes, implant longevity, or revision rates associated with thick versus thin liners during TKA,12–14 it is widely recognized that the use of thicker liners often necessitates more extensive bone resection. This may result in increased bone loss, soft tissue imbalance, and postoperative patellofemoral complications. The tibial stem extender is designed to transmit and distribute stress. In cases with insufficient tibial bone quality, it enhances implant stability by dissipating stress across the tibial plateau and reducing shear forces at the bone–implant interface.15–17 This mechanism plays a key role in maintaining long-term biological fixation in patients with severe bone loss. Previous finite element analyses have demonstrated that the use of a tibial stem extender significantly reduces average micromotion at the bone–cement interface compared to reconstructions without a stem extender, thereby improving overall implant stability.18,19 To ensure consistent stress transfer conditions, a tibial stem extender was used across all three reconstruction models in this study.
Pure bone cement filling techniques are primarily suited for smaller, contained bone defects. In contrast, larger or uncontained defects require reinforcement with screws to improve mechanical anchorage and biological stability. However, the inherent material properties of bone cement—such as its brittleness—still pose risks of fracture, loosening, and implant subsidence.20,21 In this study, the screw-assisted bone cement technique demonstrated moderate stability. The maximum stress in the defect-filling region was relatively high, potentially promoting bone integration. However, this method exhibited greater maximum deformation of the graft material under both conditions, indicating limited resistance to both axial compression and rotational forces. Additionally, the internal fixation stress was higher than that in the metal block group, suggesting increased mechanical demand on the screws. The screw-assisted bone cement technique is cost-effective but has limited indications and is not suitable for patients with large bone defects.
Compared to the screw-assisted bone cement technique, the metal block technique exhibited superior load-bearing capacity and provided more reliable support for implant fixation and joint line restoration.22,23 However, it requires intraoperative adjustment of the resection surface to accommodate the block, which may result in passive bone loss. Reduced bone retention could complicate future revision procedures. The metal block group showed the most favorable biomechanical profile, characterized by the least maximum deformation of the graft material, the lowest internal fixation stress, and the most uniform stress distribution within the residual cancellous bone. These findings suggest that metal block augmentation offers an optimal balance between structural support and load transfer, fulfilling both mechanical and biological reconstruction goals. Metal blocks are an ideal material for bone defect reconstruction and should be the first choice for patients requiring early weight-bearing after surgery.
Locking plate fixation provides strong initial stability, especially in patients with poor bone quality. Its fixed-angle configuration helps resist micromotion and maintain alignment under complex loading conditions. This technique also facilitates bone preservation, offering favorable conditions for future revisions.24 Autologous cancellous bone grafting was harvested during the required TKA osteotomy, making it readily available, cost-effective, and associated with lower infection risk. Compared to metal block reconstruction, autologous bone grafting helps maximize the preservation of native bone and prevents additional bone loss. Although autologous bone grafting is considered the “gold standard” for its biological potential, it does not substantially improve mechanical stability when used alone, and may carry a risk of failure in large defects.25,26 Biomechanically, the locking plate group exhibited the highest internal fixation stress and the lowest stress in the graft region, reflecting a pronounced stress shielding effect that may impair long-term bone remodeling. Nonetheless, this method remains a viable option for patients requiring rapid postoperative stabilization or those with compromised bone quality.
In summary, based on the biomechanical findings and relevant clinical considerations, each reconstruction strategy offers distinct advantages for specific patient profiles. Metal block augmentation provided the most favorable biomechanical performance and is best suited for younger or more active patients with higher demands for prosthesis durability and osseointegration. The screw-assisted bone cement technique delivered moderate mechanical support and stress transfer, making it appropriate for patients with good bone quality and moderate functional needs. Although locking plate fixation resulted in higher internal stresses, its initial structural stability makes it an appropriate choice for elderly patients, those with poor bone quality, or individuals not suitable for biological reconstruction. Ultimately, the choice of reconstruction technique should be individualized based on bone quality, defect morphology, patient age, and postoperative functional expectations.
5 Conclusion
This study established finite element models to evaluate the biomechanical performance of three different reconstruction techniques for AORI IIA tibial defects during total knee arthroplasty. The results demonstrated that the metal block technique provided superior load distribution, minimal graft material deformation, and excellent overall stability. The screw-assisted bone cement technique offered moderate mechanical support with favorable stress transfer, while the locking plate fixation ensured strong initial rigidity but exhibited pronounced stress shielding. Each technique presents specific advantages suited to different clinical scenarios. Therefore, reconstruction strategy should be selected based on patient-specific factors to optimize both mechanical stability and long-term clinical outcomes.
Limitation
The limitations of this study include: (1) The load settings in the finite element analysis only simulated vertical loads and internal rotational loads experienced during human walking, without simulating the dynamic environment of knee flexion and extension; (2) The finite element model considered only the tibial bone, without accounting for the roles of muscles, ligaments, and other human structures. The forces exerted by these omitted soft tissues may result in more complex stress distributions on the tibia and internal fixation. Despite these limitations, this study provides valuable biomechanical evidence supporting the performance of the three reconstruction techniques. Our team plans to further validate these findings through clinical retrospective studies and cadaveric biomechanical experiments to enhance their clinical relevance and applicability.
CRediT authorship contribution statement
Nanshan Ma: and. Jing Zhang: and. Shaoqiang Pei: are co-first authors and were jointly involved in the design and analysis of the finite element model. Qing Xia: and. Qingyu Gao: and. Yanlei Wang: and. Jun Xie: and. Sheng Zhong: performed the data analysis. Pengfei Xin: and. Lianbo Xiao: was responsible for ensuring the accuracy of the descriptions and for confirming that all authors agreed with the final manuscript. All authors read and approved the final manuscript.
Data availability
All data included in this study are available upon request by contact with the corresponding author on reasonable request.
Ethics statement
This study was approved by the Ethics Committee of Shanghai Guanghua Hospital of Integrated Traditional Chinese and Western Medicine and was in accordance with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. Written informed consent to participate was obtained from all of the individual participants included in the study.
Consent for publication
Written informed consent was obtained from volunteer.
Authorship declaration
All authors listed meet the authorship criteria according to the latest guidelines of the International Committee of Medical Journal Editors, and ii) that all authors are in agreement with the manuscript.
Funding
This study was supported by the Shanghai Municipal Health Commission scientific research project (20214Y0168), Changning District Health and Wellness Committee Key Specialty Construction Project (20231003) and Changning District Science and Technology Commission Scientific Research Project (CNKW2022Y22, CNKW2024Y23).
References
- Primary total knee arthroplasty in patients with a significant bone defect in the medial tibial plateau: case series and literature review. Int J Surg Case Rep. 2023 Sep;110
- [Google Scholar]
- Arthroplasty with a composite of an allograft and a prosthesis for knees with severe deficiency of bone. J Bone Joint Surg Am. 1995 Mar;77(3):373-386.
- [Google Scholar]
- Classification and preoperative radiographic evaluation: knee. Orthop Clin N Am. 1998 Apr;29(2):205-217.
- [Google Scholar]
- Initial stability of type-2 tibial defect treatments. Proc Inst Mech Eng H. 2010;224(1):77-85.
- [Google Scholar]
- Finite element analysis of double-plate fixation using reversed locking compression-distal femoral plates for Vancouver B1 periprosthetic femoral fractures. BMC Muscoskelet Disord. 2021;22(1):276.
- [Google Scholar]
- The effect of proximal tibial corrective osteotomy on menisci, tibia and tarsal bones: a finite element model study of tibia vara. Int J Med Robot. 2014 Mar;10(1):93-97.
- [Google Scholar]
- The optimal angle of screw for using cement-screw technique to repair tibial defect in total knee arthroplasty: a finite element analysis. J Orthop Surg Res. 2022 Jul 26;17(1):363.
- [Google Scholar]
- Biomechanical analysis of different fixation methods for Rorabeck II supracondylar femoral fractures after total knee arthroplasty. Knee. 2023;41:204-213.
- [Google Scholar]
- Tibial component fixation in deficient tibial bone stock. Clin Orthop Relat Res. 1984 Apr;184:302-308.
- [Google Scholar]
- Techniques for filling tibiofemoral bone defect during revision total knee arthroplasty. Orthop Traumatol Surg Res. 2021 Feb;107(1S)
- [Google Scholar]
- Tibial tray augmentation with modular metal wedges for tibial bone stock deficiency. Clin Orthop Relat Res. 1989 Nov;248:71-79.
- [Google Scholar]
- "Thicker" polyethylene bearings are not associated with higher failure rates in primary total knee arthroplasty. J Arthroplast. 2018 Sep;33(9):2810-2814.
- [Google Scholar]
- "Thicker" polyethylene bearings are associated with higher failure rates in primary total knee arthroplasty. J Arthroplast. 2010 Sep;25(6 Suppl):17-20.
- [Google Scholar]
- The effect of polyethylene thickness on revision rates of contemporary cruciate retaining knee replacements. A registry analysis. Knee. 2021 Dec;33:243-251.
- [Google Scholar]
- Total knee arthroplasty in the osteoporotic tibia: a biomechanical evaluation of the role of stem extensions and cementing techniques. J Am Acad Orthop Surg. 2019 May 15;27(10):370-374.
- [Google Scholar]
- Primary cementless total knee arthroplasty with or without stem extension: a matched comparative study of ninety eight standard stems versus ninety eight long stems after more than ten years of follow-up. Int Orthop. 2019 Aug;43(8):1849-1857.
- [Google Scholar]
- Metaphyseal cones in revision total knee arthroplasty: the role of stems. Bone Joint Res. 2020 May 16;9(4):162-172.
- [Google Scholar]
- The influence of different tibial stem designs in load sharing and stability at the cement-bone interface in revision TKA. Knee. 2008 Jun;15(3):227-232.
- [Google Scholar]
- The influence of different tibial stem designs in load sharing and stability at the cement-bone interface in revision TKA. Knee. 2008;15(3):227-232.
- [Google Scholar]
- Medial screws and cement: a possible mechanical augmentation in total knee arthroplasty. J Arthroplast. 2004 Aug;19(5):587-589.
- [Google Scholar]
- Finite element assessment of the screw and cement technique in total knee arthroplasty. BioMed Res Int. 2020 Oct 15;2020
- [Google Scholar]
- Finite element assessment of block-augmented total knee arthroplasty. Comput Methods Biomech Biomed Eng. 2015;18(15):1726-1736.
- [Google Scholar]
- Biomechanical comparison between metal block and cement-screw techniques for the treatment of tibial bone defect in total knee arthroplasty based on finite element analysis. Comput Biol Med. 2020 Oct;125
- [Google Scholar]
- The limit of tolerable micromotion for implant osseointegration: a systematic review. Sci Rep. 2021;11(1)
- [Google Scholar]
- Bone regeneration: current concepts and future directions. BMC Med. 2011 May 31;9:66.
- [Google Scholar]
- Treatment of 10-mm-Deep or greater uncontained tibial bone defect in primary total knee reconstruction without metal augmentation: autologous oblique structural peg bone and cancellous chip bone grafting. Clin Orthop Surg. 2021 Jun;13(2):168-174.
- [Google Scholar]