View/Download PDF
18 (); 41-49
doi:
10.1016/j.jor.2019.09.019

Translate:

Experimental and finite element investigation of total ankle replacement: A review of literature and recommendations

Author affiliations
School of Engineering, Indian Institute of Technology Mandi, Kamand, Mandi, 175005, Himachal Pradesh, India
Article History
Corresponding author(s)

∗Corresponding author: Rajesh Ghosh. rajesh@iitmandi.ac.in

Article tools · PDF & citations

Abstract

Abstract

This paper briefly reviews the different methodology, technology, challenges, and outcomes of various studies related to TAR prosthesis based on numerical and experimental techniques. Very less in-vitro experimental studies on TAR have been found than finite element (FE) studies. Due to the invasive nature of the experimental approach, inadequacy and less clinical information, computational modelling has been widely used by the researchers. This paper critically examines the part related to FE modelling and experimental analysis. Some recommendation related to modelling of bones, cartilages, ligaments, muscles, and implant-bone interface condition were discussed for better understanding the results and better clinical significance.

Keywords

  • Ankle
  • Total ankle replacement
  • Implant
  • Finite element method
  • Experimental analysis
Disclaimer:
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
1

1 Introduction

The ankle joint is one of the most complicated joints in the human body. In the case of ankle replacement, post-traumatic is the dominant one, accounting for 54%, whereas primary osteoarthritis had 19%, and rheumatoid had 14.6%.1 Ankle fusion was the earlier method of treatment, but the unsatisfactory result of this treatment made it infamous, and it was not suitable for long-term outcomes. The total ankle replacement (TAR) was first time introduced in the 1970s. Initially, it was not so popular because of its poor results. Later on, the new designs which are very close to ankle anatomy give the good short and intermediate-term results and lead to an interest in TAR.2 However, long term performance has been the leading causes of concerns. Hence, there is a need to improve the survival rate and the long term performance of the existing TAR.

The annual report of National Joint Registry for England and Wales, (2018)3 reported a total number of 4,687 primary ankle operation were done from 1 April 2010 to 31st December 2017. Out of these, 4,602 (98.2%) were uncemented, and 82 (1.8%) were used cemented fixation for the implant. More operations were performed in men (59%) than women. The most common age group was 61–74 years, whereas the overall range of 17–93 years as reported by the National Joint Registry for England and Wales, (2018).3 The aseptic loosening (both tibial and talar component) was the primary reason for the revision of 211 (4.5%) number of patients after primary TAR.3 A total number of 2235 ankle arthroplasties was done from July 2006 to December 2016 as reported by the Australian Orthopaedic Association, (2017).4 Aseptic loosening/bone lysis (29.2% cases) was the common indication for revision as per the Australian Orthopaedic Association (2017)4 supplementary report. The annual report of NewZealand Joint register, (2017)5 reported a total number of 1380 registered ankle arthroplasty from January 2000 to December 2016. An 18% increase in ankle replacement was observed as compared with the last year annual report.5 The median age was found to be 65.91 years, along with the overall range of 32.32–95.91 years. Ankle arthroplasty was performed in 842 (61.01%) male and 538 (38.99%) female patients. A revision procedure was done in 152 (11%) cases which were caused owing to the pain (65), loosening of talar component (46), loosening of tibial component (32), deep infection (17), and fracture talus (3) as reported by the NewZealand Joint register, (2017).5 The Swedish Ankle Registry, (2017)6 reported that 296 primary ankle replacements were performed between 2013 and 2017. A total number of 277 ankle replacements have been revised, since 1993. Including all designs, the prosthetic survival at five years was estimated to 0.81 (95% CI: 0.79–0.83) and 0.69 (95% CI: 0.67–0.71) at ten years as reported by the Swedish Ankle Registry, (2017).6 The Dutch Arthroplasty Register, (2017)7 reported a total number of 405 registered ankle replacement in the years from 2014 to 2016. Out of these, 334 were primary ankle arthroplasty, and 71 were revision arthroplasty. Inlay wear and loosening of talus component were the main indications of revision after total ankle replacement in the Netherlands.7 Finnish Arthroplasty Register, (2017)8 reported a total number of 573 ankle arthroplasties were done between the year 1982 and 2006. In their annual report, survivorship was 83% for the past five years. A total number of 1371 ankle replacements, out of these 1047 primary and 324 revision were registered between 1994 and 2015 in the Norwegian Arthroplasty Register, (2017) report.9 Posttraumatic arthritis is the primary indication for ankle replacement surgery.9 The main purpose of TAR should try to close the actual anatomy of the joint as much as possible, with a suitable range of motion, should follow the proper gait pattern. The design of TAR expected to have a reproducible surgical technique, minimal bone resection, rapid and adequate bone growth, and minimum constraint, replication of physiological ankle motion and pain relief.

Finite element method (FEM) is a widely used technique nowadays to solve biomechanical problems; it has been efficiently used to control the different parameters related to the design of ankle prostheses. While in vitro experimental studies also have been done for analyzing the performance of various parameters which influence the prosthetic design of TAR. But due to the invasive nature of the experimental approach, inadequacy and less clinical information, computational modelling such as FE method have been used by the researchers, due to its capability of modelling structures with irregular geometry and complex material properties, and efficiently simulate the complicated boundary and loading conditions in both the analyses of static and dynamic.10

Vickerstaff et al. (2007)2 reviewed the current status of TAR and suggested that the existing designs have still various essential issues which must be addressed for the better clinical results in long-term. Heuvel et al. (2010)11 reviewed the evolution in the TAR design from the first generation to the newer design. Gougoulias and Maffulli, (2013)12 also studied the history of TAR. They were suggested that the three-component mobile-bearing design showed better agreement than the two-component design of TAR.12 Yu and Sheskier, (2014)13 focused on the design advancement and forthcoming applications of TAR. A failure rate of 18% for the patients who received the TAR surgery and 28% who was undergoing the ankle arthrodesis was observed in long-term outcomes as reported by Yu and Sheskier, (2014).13 They suggested that the TAR has become a sustainable selection for surgical treatment of ankle arthritis and fracture.13 In the existing literature it has been observed that most of the reviews were focused on to investigate the clinical outcomes of TAR after a certain period of follow up study.14–21 It has been found that multiple parameters such as infection, aseptic loosening, lysis, misalignment, implant fracture, wear or polyethylene component, and meniscal insert dislocation, component migration/dissociation, pain, stiffness, soft tissue impingement, etc. have been a significant influence on the survival rate of TAR as per the clinical feedback. Different work has been found in the literature which deals with the various factors of TAR prosthesis such as, manufacturing aspects, internal stress distribution on the bones before and after the insertion of prostheses, micromotion, the effect of bone prosthesis interface with contact, bone remodelling, implant failure, improved design of ankle prostheses etc. by FEM and experimental investigation. Any review based on FE and experimental analysis has not been done so far. In this paper, we have critically reviewed the past work done related to FEM and experimental analysis. The motivation of this study came from the studies had been done by the various researchers on the different aspects of ankle prostheses, and the aspects did not address in their reports which lead to the interest on a review of the ankle joint prostheses by FEM and in vitro experimental studies. Therefore, the objective of this paper was to critically examine the current research related to FEM and experimental analysis and present an overview of the total ankle joint replacement prostheses and some recommendation. This paper also discusses the steps involved to create the finite element model of the intact and prosthetic ankle, and discuss the development of the new design of ankle joint prostheses and their survival rates, as well as the feasibility of new design within the controllable parameters.

2

2 Finite element modelling and analysis

The FE analysis evolves one of the most efficient techniques for solving complex biomechanical problems. As the three dimensional (3D) FE model of the prosthetic foot-ankle complex consists of different components such as bone, soft tissue, cartilage, ligament, prosthetic components, and plantar fascia, etc. each is having different material characteristics. The material properties of the various anatomical components of the FE model used in the different study are presented in Table 1.22–37 Whereas the properties of the ligaments and different TAR material are shown in Table 222,23,25,38–44 and Table 322–27,29,45–49 respectively.

Table 1 Material properties of the different anatomical components of finite element model.
Components Material Properties References
Young's Modulus E(MPa) Poisson's Ratio(??)
Cortical Bone 19000 0.3 22 23
17580 0.3 24
Cancellous Bone E=4778ρ1.99 0.3 25
3.60ρ−0.14 0<ρ≤0.118.49ρ1.93 1<ρ≤0.378.87ρ- 0.57 0.37<ρ≤1.54.83ρ2.39ρ>.1.5 0.3 26
E=3790ρ3 0.3 27
500 0.3 22
280 0.3 24
Bone (Cortical and Cancellous) 7300 0.3 28–33
Soft tissues Hyperelastic – 29 , 31
0.15 0.45 30
Hyperelastic 0.452 34
Hyperelastic 0.5 32
Cartilage 10 0.4 29 , 32
1 0.4 30 , 31 , 33 , 35
260 Incompressible 31 , 33 , 36
260 – 30
Plantar Fascia 350 0.4 29
350 – 30 , 32 , 33
350 Incompressible 31 , 37
Table 2 Stiffness and Material Properties of Ligaments used in the previous literature.22,23,25,38–44.
Ligaments Represented in Previous Models Stiffness (N/mm) Young's modulus (E), MPa Poisson's ratio(ν)
Interosseous I– IV (Inte I– IV) 400 260 0.4
Anterior tibiofibular (ATiFL) 90 160 0.49
Posterior tibiofibular (PTiFL) 90 160 0.49
Calcaneofibular (CaFL) 70 512 0.49
Tibiocalcaneal (TiCa) 122 512 0.49
Anterior talofibular (ATaFL) 90 255.5 0.49
Posterior talofibular (PTaFL) 70 216.5 0.49
Anterior tibiotalar (ATiTL) 70 184.5 0.49
Posterior tibiotalar (PTiTL) 80 99.5 0.49
Interosseous talocalcaneal (ITaCL) 70 260 0.4
Lateral talocalcaneal (LTaCL) 70 260 0.4
Medial talocalcaneal (MTaCL) 70 260 0.4
Posterior talocalcaneal (PTaCL) 70 260 0.4
Dorsal talonavicular (DTaNa) 70 260 0.4
Calcaneonavicular (CaNa) 70 260 0.4
Cuboidenavicular (CuNa) 70 260 0.4
Table 3 Properties of TAR implant materials.
Material Material Properties References
Young's Modulus E(MPa) Poisson's Ratio(??)
Cobalt chromium alloy (Co–Cr) 250,000 0.29 45 , 46
Cobalt chromium-Molybdenum (Co–Cr–Mo) 210,000 0.3 22 , 23 , 26 , 27 , 29 , 47
241,000 0.3 48
Ceramic 350,000 0.26 25
Carbon-fiber reinforced polyetheretherketone (CFR-PEEK) 130,00 0.4 25
180,00 0.4 49
Titanium 100,000 0.3 47
110,000 0.33 24
Stainless steel 200,000 0.3 45 46
189,600 0.3 24
Ti-6Al-4V 110,000 0.342 45 , 46
114,000 0.342 49
115,000 0.36 48
Ultra High Molecular Weight polyethylene (UHMWPE) 556 0.461 22 , , 26 , 45 , 46
556.31 0.461 49
600 0.46 29
1174 0.4 23,25
2.1

2.1 Modelling of different elements

To accurately create a 3D FE model, the 3D image resulting from magnetic resonance imaging (MRI) or computed tomography (CT) scan is used. However, in the past, mostly CT scan data was used for the development of FE models of bones. Ligaments were modelled as a 3D array of five fibers by Reggiani et al. (2006).50 Wei et al. (2011)43 and Ramlee et al. (2013)38 used the linear spring elements to model the ligaments. Ozen et al. (2013)29 and Rodrigues, (2013)22 used the tension only truss elements for the ligaments. Martinelli et al. (2017)48 were modelled the ligaments as a one-dimensional non-linear spring element. Recently, Mondal and Ghosh, (2018, 2019)23,25 modelled the ligaments by using linear spring elements in their FE models. Cartilages were modelled as a solid part which were generated from the adjacent surfaces of all joint bone.29,38,44 Prosthetic components were developed independently by using design software.

2.2

2.2 Geometry discretization

Discretization of the solid model of bony structure and prosthetic components is the next vital step. Most of the previous studies were used tetrahedral and hexahedral elements for the meshing. Miller et al. (2004)24 was used a different type of elements for the meshing of bones, and prosthetic components. Ten node tetrahedral element was used for the meshing of the metallic components and bone whereas the layer of cortical bone in the distal tibia and fibula was discretized with four-node quadrilateral shell elements (thickness of 1.0 mm) and the eight-node hexahedral elements was used for polyethylene mesh.24 Eight-noded-hexahedral elements were used by Reggiani et al. (2006)50 in their study for the meshing of prosthetic components only. The implants were modelled as three-dimensional solid hex elements by Espinosa et al. (2010).47 Reggiani et al. (2006)50 and Espinosa et al. (2010)47 did not consider bony structures. Terrier et al. (2014)51 was used quadratic ten nodes tetrahedral elements for bones and the rigid quadrilateral elements for the implants. However, most of the earlier studies preferred the tetrahedral elements for the meshing of bones and implant components,22,23,29,45,52,53 since, the tetrahedral configuration is more suitable for meshing of non-prismatic geometry. Tadepalli et al. (2011)54 performed a study to compare the hexahedral and tetrahedral elements in FE analysis of foot biomechanics. They reported that the tetrahedral element formulation seems to be very promising for its expedited model development and eased mesh generation in foot biomechanics than the hexahedral element simulations.

2.3

2.3 Material modelling

After the mesh generation, each element in the mesh has to be assigned material properties. To overcome the problem related to the partial volume effect, each bone was separated by a cortical and cancellous bone. The cortical bone was assumed as elastic, linear and isotropic having Young's modulus and Poisson's ratio of 19 GPa and 0.3 respectively (Table 1).22,23

Cancellous bone material properties were considered to be heterogeneous, elastic, and isotropic.23,25–27 Young's modulus of cancellous bone was determined using Young's modulus-density relationships.55–57 The different Young's modulus-density relationships were used for the assignment of material properties of cancellous bone, which are presented in Table 1. Rodrigues, (2013)22 considered Young's modulus and Poisson's ratio of 500 MPa and 0.3, respectively, for cancellous bone. In some studies Young's modulus and Poisson's ratio of bony structures were selected as 7300 MPa and 0.3, respectively.10,28–30,32

In most of the studies soft tissues (skin, plantar fascia) were assumed as Hyperelastic materials,10,29,32,34 except Hsu et al. (2006)30 where Young's modulus and Poisson's ratio were selected as 0.15 MPa and 0.45 respectively. Young's modulus and Poisson's ratio of cartilage were chosen as 10 MPa and 0.4 29,32 and 1 MPa and 0.4 respectively.10,30,35 Similarly for the ligaments and plantar fascia, Young's modulus was selected as 260 MPa29,30,31,36 and 350 MPa respectively.29–32,37 Corazza et al. (2003)41 and Reggiani et al. (2006)50 considered eight major ligaments namely, tibiocalcaneal (TiCa), calcaneofibular (CaFi), anteriortalofibular (ATaFi), posterior talofibular (PTaFi), tibionavicular (TiNa), deep anterior tibiotalar (DATiTa), deep posterior tibiotalar (DPTiTa) and superior tibiotalar (STiTa)in their study. Rodrigues, (2013)22 considered the same eight number of ligaments as similar to Reggiani et al. (2006).50 Ozen et al. (2013)29 considered the 76 and 71 number of ligaments corresponding to the intact and prosthetic ankle-foot complex. Martinelli et al. (2017)48 used the five major ligaments, namely, ATaFi, TiCa, CaFi, TiNa, and STiTa. Wei et al. (2011)43 considered the total 20 number of ligaments for their model. A total number of 103 ligaments were considered by Wang et al. (2018)33 in their FE model. Ramlee et al. (2013)38 represents the 37 number of ligaments in their study. Mondal and Ghosh, (2017)44 used the total sixteen number of ligaments in their models. The ligaments around the ankle joint used by the previous studies corresponding their stiffness and material properties are presented in Table 2. Material properties of prosthetic components are shown in Table 3.

2.4

2.4 Boundary and loading conditions

The boundary and loading conditions are direct factors contributing to model accuracy. Loads acting on ankle joint during various sub-phases of the gait cycle, initial contact, heel strike, midstance, forefoot contact, push-off, and toe-off were 675N, 1350N, 2100N, 2550N, 3000N, 300N respectively.58 The effect of axial impulsive loading applied at the plantar surface was studied by Bandak et al. (2001).59 Miller et al. (2004)24 applied a total load of 666 N, which was five times bodyweight carried by the tibia and fibula applied to the joint during normal gait. They applied the reaction forces at the bottom surface of the polyethylene component.24 Anderson et al. (2007)60 applied a 600N compressive load from the proximal end of tibia and fibula across the joint and the flexion/extension rotations of the foot were fixed during the application of force. For the stance phase of gait, the tibial component was permitted to translate along and rotate about the vertical axis while the talar component was translated anteroposteriorly and to rotate about the mediolateral axis.50 The talar component was rigidly fixed by transverse translation and rotation relative to the sagittal axis. The ground reaction force was considered as a concentrated force underneath the ground support.29 For the normal foot-ankle and prosthetic foot-ankle models during balanced standing; only the Achilles tendon loading was considered while other muscle forces were neglected.29 The other muscle forces were not considered because their magnitude is very less, and their effects on load transfer through the ankle joint is negligible.29 However, Hill's muscle model would be beneficial for the better understanding of the effect of all the muscle forces around the ankle joint on load transfer through the ankle joint.

All tibial component degree of freedom was constrained except for a single translational degree of freedom along the tibial axis.47 Espinosa et al. (2010)47 applied the load along the tibial axis according to the stance phase of gait and assuming a bodyweight of 80 kg (a heel-strike of 800 N, midstance of 2000 N, heel-off of 2800 N, and toe-off of 800 N). A static force of 5200 N was applied distally to the tibia and the talus bone by Bouguecha et al. (2011).27 Elliot et al. (2014)45 applied a uniform pressure corresponding to the normal component of the ankle gait to the talar component in each finite element model in their study. The proximal part of the tibia was fully constrained, whereas an axial compressive force of 2 kN was applied at the center of the universal joint.51

Wear on the polyethylene liner also would take place due to the shear and torsion loading conditions.46 Kerschhofer et al. (2016)49 only considered the prosthetic components in which they applied a load of 876.4 N at the top surface of the tibial component. Sopher et al. (2017)26 applied the loading conditions as point forces evenly distributed between the nodes forming the implant articular surfaces. The physiological peak axial contact load was acting on the implant (tibial and talar components) during the six segments of the stance phase of gait (2%, 12%, 31%, 45%, 50%, and 55%). The proximal end of the tibia and talus were fixed.26 Wang et al. (2018)33 and Smyth et al. (2017)84 were applied the loading and boundary conditions as per data obtained from Reggiani et al. (2006)50 during the stance phase of gait.

Recently, Mondal and Ghosh, (2018, 2019)23,25 were applied the loading condition corresponding to the three positions of the ankle during gait such as dorsiflexion, neutral, and plantar flexion and the magnitude of applied reaction forces were based on the data provided by Reggiani et al. (2006).50 They used the compressive and anterior-posterior tangential forces from the force predictions reported by Seireg and Arvikar, (1975)61 and the internal-external torque and plantar-dorsiflexion rotation from the data predicted by Procter and Paul, (1982).62 Mondal and Ghosh, (2018)23 applied the muscle force (Achilles tendon) at the extreme posterior position of the calcaneus bone which was 75% of the total body weight as similar to Ozen et al. (2013).29

3

3 Finite element investigations

3.1

3.1 Intact ankle joint

A 3D model of the human ankle joint was developed using CT scan data set which includes the major ligaments, layer of soft tissues of the foot and the bone geometry.59 This is the advanced FE models of lower extremities by Bandak et al. (2001)59 than the previous FE models by Beaugonin et al. (1995)63 and Schauer et al. (1995).64 The axial impulsive load was considered as applied loading condition to investigate the injury mechanisms at the foot geometry.59 Their results suggested that maximum stress obtained at the calcaneus than the tibia and talus bone geometry.59 Gefen et al. (2000)58 developed a model of the foot for performing the analysis during the gait which included the ligaments and soft tissues same as the Bandak et al. (2001).59 They have done a statistical study to characterize the specific gait on the various parameters such as age, gender, human activities, etc. Their results revealed the von Mises stress distribution at the foot skeleton and plantar pressure during the six sub-phase of gait, namely, initial contact, heel strike, midstance, forefeet contact, push off and toe-off.58

Cheung et al. (2005)65 constructed an FE model of foot-ankle complex using MRI to understand the biomechanics of the foot and ankle. They reported that the increase in plantar pressure about 35% and 33% at the forefoot and rear foot respectively owing to the increment of the stiffness of soft tissues whereas the bone stress was found to be highest at the third metatarsal bone.65 An FE model of the ankle was developed, and physical validation was done by Anderson et al. (2007).60 Two cadaveric specimens were used for their study. A good agreement was observed between experimental and FE results of the ankle in terms of stress and contact area. The discrepancy in contact stress and contact area results for FE and experimental analysis were found to be 1.5% and 1.7%, respectively.60 Ozen et al. (2013)29 were developed the model for normal foot ankle and prosthetic foot ankle, and compared the results in terms of plantar pressure, von Misses stress distribution during the balance standing positions. MIMICS software was used to generate 3D images of soft tissues, and bones of the normal ankle-foot complex. The results revealed that maximum plantar pressure occurs in case of normal ankle-foot complex, about 0.159 MPa at the rear foot regions of the first, third, fifth metatarsals head. In case of the normal ankle-foot complex, the peak von Mises stress was obtained about 6.766 MPa, 5.50 MPa, 3.464 MPa, and 3.371 MPa at 3rd metatarsals, tibia, calcaneus, and 2nd metatarsals respectively. The constraint is the resistance of movement of the foot in any particular direction; external rotation causes the ankle injury.43 Wei et al. (2011)43 developed a model for predicting the foot constraint effect on the ankle injury, which is due to the external rotation. They have studied the kinematics of the ankle joint due to the various external foot rotation and predict the strains in various ligaments.43 Ramlee et al. (2013)38 developed a 3D FE model of foot ankle complex and validated the results of the displacement with the experimental work. They considered the tibia, talus, fibula, calcaneus, and navicular, cuboid, and metatarsals bones and applied seven different forces at the tibia.38 An FE model of the foot-ankle complex was developed by Chen et al. (2014)66 to predict the internal plantar pressure, displacements, and stress. This model was validated with the experimental measurements in terms of plantar pressure, displacements, and stress.66 They reported that the stress concentration due to the soft tissue contacts with geometrically irregular bony structures caused the high von Mises stress at the plantar fascia.66

Mondal and Ghosh, (2017)44 studied the natural ankle joint to investigate the load transfer through the tibia and talus bone. The effect of ligaments, muscle force, and material distribution on stress distribution at tibia and talus bone has been analyzed. They reported that muscle force, material distribution, and ligaments have significant effects on stress distribution at the tibia and talus bone. They suggested that for the development of realistic 3D FE model of intact ankle joint; ligaments, muscle force, material distribution of bone and proper musculoskeletal boundary and loading condition should be considered. Most recently, Wang et al. (2018)33 investigated the biomechanical difference in the ankle and foot before and after the joint replacements. They have developed 3D FE models of an intact foot and an implanted foot to simulate the stance phase of gait. Computational prediction of plantar pressure and joint contact pressure was validated and compared with experimental results.33 Some of the studies focused on the development of intact human foot-ankle complex for investigation in the various applications such as footwear design, insole design, ankle-foot orthosis, etc.10,54,67–76

3.2

3.2 Implanted ankle joint

In the year of 1970s, the first ankle joint was replaced by Lord and Marotte. Initially, this method of treatment was not so effective due to its poor clinical outcomes due to this arthrodesis was a better choice at that time. Later on, the new designs of ankle prostheses, which are very much close to the actual anatomy give a satisfactory result lead to the interest of TAR. The different researchers have developed numerous FE models; among these, some model and their analysis have been critically reviewed in this paper.

Polyethylene liner is generally used between the tibia and talar component for the ankle joint in which the stress is developed, Miller et al. (2004)24 compute the contact pressure and stress distribution in polyethylene lines for a semi-constrained ankle prosthesis. Miller et al. (2004)24 used two different talar component design. They have studied the parameters which effecting the polyethylene stress such as variation in the shape of talar components, design variations in the ankle replacement, and validated it with different physiologic conditions. They reported that wider talar component would reduce the stresses at the polyethylene.24 Kinematics of the joint and the pressure generated at the prosthetic surface, these are the two important factors in which the success of TAR depends. Reggiani et al. (2006)50 developed an explicit FE model of a three-component TAR during the stance phase of gait. They have studied the different influenced parameters like kinematics, contact forces, ligament forces during the standing and gait conditions. Their results revealed that during the simulated stance condition, the meniscal bearing shifted from its initial position to a distance of about 5.6 mm, and the talar component was about 8.3 mm due to the anterior-posterior displacement. The contact pressure distribution at the articulating surface was observed at a maximum value of 16.8 MPa.50 The ligament forces were found to be within reasonable values during the simulated stance condition.50 Misalignment is one of the major issues for failure mode of the TAR; a study has been done by Espinosa et al. (2010)47 on the positioning of the components of the prosthetic model which results into high joint contact pressure distribution and wear in the liners. Two FE prosthetic models were used to predict the maximum and average pressure on the polyethylene insert surfaces. As the alignment of the components deviated from its initial position, the magnitude of pressure distribution was increased. The result showed that the value of yield stress of polyethylene (18–20 MPa) was less than the average contact pressure.47 This paper also compared the three-component prosthetic design with two-component design and found that three-component design developed less pressure at joint contact than the two-component design.47

Apart from this failure mode of TAR, there are various factors which affect the survivals rate of TAR. Elliot et al. (2014)45 have also been studied about the failure mode of the TAR. They indicated that high contact stress and wear in the polyethylene liner were the main causes of the failure of the material over time. Reggiani et al. (2006)50 already described the performance of TAR during the stance phase of gait. They developed an FE model of a wright state university (WSU) patented TAR prosthesis. The optimization technique has been used to predict the best method to wear characterization and reduction. The value of maximum wear rate (25.598 MPa, 3.74 mm^3/year) and maximum von Mises stress (11.52 MPa)45 was obtained from the optimized model. Putra et al. (2014)52 has done a very much similar study about the wear on total ankle replacement. However, they have used the stance phase of the gait cycle as a loading condition which was not studied before. They developed a 3D model of the right implanted ankle by using the BOX@R TAR model. They predicted that the linear wear depth was 0.1614 mm/million cycle, and cumulative volumetric wear was 30.5 mm^3, respectively.52 Wear on the polyethylene liner also would take place due to the shear and torsion loading conditions.52

An FE model was developed with viscoelastic properties of the polyethylene liner to predict the wear rate by the help of wear rate equation.46 They have also predicted the von Mises and contact stress at the polyethylene liner. Due to the torsion load, the maximum average stress was obtained 14.46 MPa and wear rate was obtained 0.67 mm^3/year for the mobile bearing model while it was 10.55 MPa and 0.33 mm^3/year when the analytical method was used. Wear on the polyethylene liner would be more in axial loading conditions than the torsion and shear loading conditions; this prediction was made by Gundapaneni et al. (2015).46 Ozen et al. (2013)29 were developed the model for the intact and prosthetic ankle. MIMICS software was used to generate 3D images of soft tissues, bones, and the components of the normal and prosthetic ankle-foot complex. After the reconstruction of the solid model, the 3D FE model of the intact and prosthetic ankle was developed.29 The von Misses stress increased from 0.01 MPa to 0.03 MPa at the doral soft tissues after the implantation. In case of normal ankle-foot complex the peak von Mises stress was obtained about 6.766 MPa, 5.50 MPa, 3.464 MPa, and 3.371 MPa at 3rd metatarsals, tibia, calcaneus, and 2nd metatarsals respectively while for the prosthetic one it was obtained 7.560 MPa, 5.58 MPa, 4.794 MPa at the tibia, 4th metatarsals, and 5th metatarsals respectively.29

It is necessary that the exist FE model of ankle prostheses must be validated by clinical or experimental data's. Some literature exists on the experimental validation of the FE model of foot-ankle complex, such as Terrier et al. (2014)51 developed an FE model of total ankle replacement and validated the results obtained from the FEM analysis with the experimental data. Eight cadaveric tibias were used for the experimental investigation. A non-cemented mobile-bearing prosthetic design was inserted into the eight cadaveric tibias, 2 KN axial forces were applied to predict the strain at the bone surface, and digital image correlation was used to measure the strain. The FE model was developed from the CT images, the first step of CT was used to develop the geometry and evaluate the mechanical properties of the tibias, the second one was used for the position of the implant, and the final step of CT was used to analyze the bone-implant interface.51 The maximum strain value was predicted around the implant keel as predicted by Terrier et al. (2014).51 Bone remodelling is an important parameter which must be taken into consideration for the optimization of implant geometry. Bouguecha et al. (2011)27 developed the FE models of the intact and implanted ankle joint. Their results showed that 2% and 13% bone mass loss were found at the tibia and talus respectively due to the bone remodelling. However, Bouguecha et al. (2011)27 reported that the more density decline was observed at the talus due to the large implant dimensions (talar component) than the size of the talus.

Kerschhofer et al. (2016)49 studied the effect of poly-ether-ether-ketone (PEEK) and its composites combination on the wear rate of TAR devices. They have identified that PEEK and carbon fibre reinforced polyether-ether-ketone (CFR-PEEK) as an alternative to metal for orthopaedic implants. Second generation Wright State University (WSU) TAR devices were used by Kerschhofer et al. (2016)49 to determine the wear characterization. The maximum von Mises stress was obtained 25.06 MPa for the model having implant material combination of CFR-PEEK and UHMWPE which was lower than the maximum von Mises stress of 26.85 MPa for the model having a material combination of Ti–6Al–4V and UHMWPE. An FE study on contact stresses, contact pressure, and contact area for a fixed bearing TAR were done by Martinelli et al. (2017).48 CT scan data sets were used to generate the 3D FE model of the implanted ankle using Zimmer Trabecular Metal prosthesis to investigate the contact pressure, von Mises stresses and contact area of the polyethylene component during the stance phase of the gait. The maximum von Mises stress of 14.1 MPa was obtained at the anterior section of the polyethylene component during the 40% of the gait cycle whereas the peak value of von Mises stresses were 10.8 MPa and 5.4 MPa at the central (37% of gait cycle) and posterior section of the polyethylene component (end-stage of gait) respectively. Their proposed fixed bearing design of TAR showed an excellent agreement with the mobile bearing design of TAR in terms of contact stress and wore rate.48 The maximum Round resection of bone gives better bony support than the flat resection, as stated by Bischoff et al. (2017).77 CT scan data sets of 116 subjects were considered for their study. They observed a major reduction in bony support for tibia and talus bone for flat resection of bone as compared to round resection.77

Sopher et al. (2017)26 studied the effect of implant positioning on implant-bone micromotion and bone strains. They have modelled three prosthetic devices, namely, BOX@R, MOBILITY@R, and SALTO@R. The largest micromotion of all tibial and talar components was observed for MOBILITY@R and SALTO@R prosthesis, respectively. Due to malpositioning of the implant components, increased in micromotion and bone strains were observed, which result in increasing the risk for TAR failure. Recently, Mondal and Ghosh, (2018)23 studied the effect of implant orientation and implant-bone interfacial conditions on strain energy density distribution at tibia due to TAR. They have modelled FE models of the optimal implanted ankle, and varus and valgus angles of 5° and 10°, respectively to understand the effect of implant orientation on strain energy density distribution at tibia bone. They were assumed debonded and bonded implant-bone interfacial condition for their study. They found that micromotion was lower than 50 μm, which suggests the possibility of osseointegration onto the coated surface of the implant and supports their assumptions of the bonded implant-bone interface condition. They concluded that proper osseointegration between the implant-bone interfaces and implant orientation would be beneficial for the long-term success of the implant. Most recently, the effect of implant orientation and implant material on the cancellous bone strain, maximum contact pressure at the meniscal bearing surface, wear depth at the polyethylene component, and micromotion between the tibia bone and tibial component was investigated by Mondal and Ghosh, (2019).25 They were considered three different implant material combination, namely, (cobalt-chromium molybdenum) Co–Cr–Mo with UHMWPE, Ceramic with UHMWPE, and CFR-PEEK with Ceramic. Implant orientation influences the wear depth, contact pressure, and micromotion. Contact pressure and wear depth at the meniscal bearing surface, and micromotion between the tibia and tibial component was found to be high due to the implant orientation.25 Implant material does not have a significant effect on contact pressure and micromotion.25 They concluded that alumina ceramic as an alternative to metal and CFR-PEEK in place of UHMWPE could be used for the better performance of TAR. Their results are an excellent agreement with the Kerschhofer et al. (2016).49

Wang et al. (2018)33 developed 3D FE models of an implanted foot to simulate the stance phase of gait. The plantar pressure and joint contact pressure for the implanted foot were validated and compared with experimental results.33 They observed that the medial cuneonavicular joint bears the highest contact pressure among all the joints in the foot after total ankle arthroplasty.

4

4 Experimental investigations

Apart from the FE analysis, some studies have been done on experimental work. Valderrabano et al. (2003)78–80 have done the in-vitro study to investigate the changes in the range of motion (ROM), movement transfer, and talar movement after the ankle fusion and TAR. They were considered six cadaveric fresh frozen specimens for their study. To predict the ROM in the sagittal and frontal plane and internal and external tibial rotation, a moment of 100Nm was applied to the footplate and tibia respectively. They performed this experiment for normal ankle, a fused ankle, Agility@TM, HINTERGA@TM, and S.T.A.R@TM prostheses and found that the HINTERGA@TM and S.T.A.R@TM prosthesis does not alter the ROM significantly for the inversion/eversion than the two-component prostheses and fused ankle.78 For the movement transfer an axial load of 200 N was applied at the top surface (proximal end) of the tibia specimen and predicts that for the dorsiflexion/plantarflexion and inversion/eversion ankle joint fusion increased the movement transfer to tibial rotation by a factor of 2.4 and 18.5 respectively while for the two and three-component prostheses these changes were not observed by Valderrabano et al. (2003).79 Similarly, for the talar movement, the same load was applied at the top surface (proximal end) of the tibia specimen, results predict that three-component designs seem to allow the talar range of motion whereas the two-component design restricts the talar movement within the ankle joint complex.80

Miller et al. (2004)24 investigate the stress distribution in polyethylene lines for a semi-constrained ankle prosthesis by using the digitization of cadaveric specimen. They have observed that wider talar component which satisfied the primary kinematic criteria will reduce the stresses.24 Saltzman et al. (2004)81 have studied the effect of misalignment on the peri-ankle ligament length due to the TAR. Agility@TM prosthesis was implanted in seven fresh frozen cadaveric specimens.81 Saltzman et al. (2004)81 considered proper physiological motion and loading condition. They observed that the anterior talofibular (ATaFi) ligament and tibiocalcaneal (TiCa) ligament were subtle to the transverse plane and coronal plane displacements respectively due to the misalignment of the tibial component. Richter et al. (2007)82 used the two different prosthetic design HINTERGA@TM and GERMEN ANKLE SYSTEM, which was implanted in seven paired cadaveric specimen. The specimens with and without prostheses were mounted on a robot and compared the results in terms of forces, torques, and motion of the bone. They were found that specimen without prosthetics generate the fewer forces and torques than the specimen with prostheses with one exception (one sample test p ≤ 0.01, p = 0.34 for GERMEN ANKLE SYSTEM when measured the lateral force). GERMEN ANKLE SYSTEM obtained the lower value of force, torque, and bone motion than the HINTERGA@TM (t-test, each p ≤ 0.05) prostheses.82 They reported that the GERMEN ANKLE SYSTEM prosthesis might have shown better clinical results than the HINTERGA@TM prosthesis.

Affatato et al. (2009)83 have been developed a three-component artificial joint where the shape of the tibial component is in a convex spherical shape, talar components are anticlastic. A meniscal component made of UHMWPE inserted between the tibial and talar component to predict the wear patterns and the wear rate of the polyethylene liner due to TAR. A maximum axial load of 2.6 was applied at the meniscal bearing component. The wear test was run at a frequency of 1.1 Hz for 2 million cycles. The microscopic and Raman crystallinity-based measurements showed a resemblance between the wear patterns generated experimentally in the wear simulator and those seen in retrievals with similar wear life.83

Recently, Terrier et al. (2014)51 have done the experimental study on total ankle replacement along with the FE analysis. They have used eight cadaveric specimens in which prosthetic design component was inserted in order to predict the strains at the bone surface by using digital image correlation (DIC). They replicated this experimental analysis by FE model analysis and found the very close result of bone strains.51 Most recently, Smyth et al. (2017)84 studied the influence of kinematics on the wear of TAR. They have been tested five TARs to wear assessment. Smyth et al. (2017)84 stated that TAR wears rates to vary depending on the gait inputs and kinematic conditions.

5

5 Discussions

Ankle arthritis such as osteoarthritis, rheumatoid arthritis, and posttraumatic arthritis and ankle fracture caused the loss of movement at the ankle joint. In the case of osteoarthritis, cartilages between the bones become thin and ultimately wear out, causing the direct contact of bones. Rheumatoid arthritis caused due to the body's immune system activity, and it is mostly found in women. Injuries due to accidents, sports activity, etc. come under posttraumatic arthritis. Ankle arthrodesis was the earlier method of treatment to overcome the problem related to ankle arthritis and fracture. In this method, two bones (tibia and talus) are fused together with the help of screws. The unsatisfactory result of this treatment made it infamous, and it was not suitable for long-term outcomes. The TAR is a surgical reconstruction procedure, in which the diseased ankle joint is removed and replaced by the artificial component known as implant or prosthesis. This surgical procedure of treatment is recognized as one of the most popular and successful treatment methods over the last few decades. In the present market, numerous designs are present such as S.T.A.R@TM, AGILITY@TM, BUECHEL-PAPAS@TM, SALTO@TM, HINTERGA@R, MOBILITY@TM, RAMSES@R, OSG@R, AES@R, and BOX@R, TNK@R, etc. Still, there are various clinical issues which need to be addressed for the better design of joint prostheses.

In the current literature survey, the recent FE model of total ankle joint prostheses and in vitro experimental study were investigated. Very few numbers of studies have been found for TAR based on FEA and in vitro experimental work. Reviewed of the studies observed that the knowledge of the stress/strain distribution, plantar pressure, wears in polyethylene liner, etc. are the critical parameters for the development and improvement of the ankle prosthesis. It is necessary to understand the biomechanical behavior of foot and ankle joint before the analysis of implanted ankle joint.

The presented overview of the recent literature based on in vitro experimental work which simplifying the kinematics of the joint after the ankle fused and TAR in terms of the range of motion, talar motion, and movement transfer and has made clear that three-component prostheses are close to the actual joint behavior.78 It has been predicted that the high value of forces, torque, and motion of the bone profoundly influence the survival rate of TAR. Effect of Wear rate of polyethylene liner and strains at the bone surface due to the different loading conditions also reviewed from the recent studies. Still, it is challenging to address various clinical issues by the experimental method because of its invasive in nature, high cost, and inadequacy. FE modelling technique is a powerful tool used by the researchers for biomechanical problems, due to its capability of modelling structures with irregular geometry and complex material properties, and efficiently simulates the complicated boundary and loading conditions. The present review of the literature demonstrated about the challenges in FEM model development of TAR, focusing on the various clinical issues of prosthetic design in various boundary conditions, loading conditions, material characteristics, and the other influencing parameters. For the design optimization the computational evaluation has served as an essential approach, and it expected the further investigation would take place to address the ankle arthritis problems effectively for surgical protocols and the design of joint prostheses with adequate accuracy. Comprehensive 3D FE models replicate the actual anatomy of the joint are necessary for the evaluation of interventions.

A geometrically detailed model consists of bones, soft tissues, cartilage, ligament, and plantar fascia. It is essential to get the proper balance between the accurate details and appropriate simplifications. Material characterization for different components such as bone, soft tissue, cartilage, ligament, and plantar fascia needs to be adequately described, especially for the soft tissues under the large deformations and dynamic conditions. A properly constrained prosthesis is required to replicate the normal ROM and minimize the shear forces at the bone prostheses interface. Therefore this point should be kept in mind before the design of any model of the prosthetic design. Incongruent geometry of the prostheses leads to less stability and low wear rates due to the less contacting surfaces and higher local stress respectively, while congruent or full conforming prostheses show the high stability that does not depend only on the ligaments for the stability, low wear rates due to the large contacting surfaces and lower local stress. The boundary and loading conditions are essential factors for obtaining an accurate model. Presently, boundary and loading conditions are obtained from human motion analysis during the standing and stance phase of gait. As per the loading conditions is a concern, apart from the axial loading shear and torsion loading situations have also been reviewed from the literature and observed that wear takes place not only under the axial loading situations, it also has been found during the shear and torsion loading conditions but significantly less than the axial loading. Still, many aspects are being defined such as manufacturing aspects of components of prosthetic design, minimal bone resection, misalignment, and bone remodelling, etc. need to be addressed for much better design and long term outcomes of the newer designs, for that further attempts should aim towards to eliminate this clinical issues.

6

6 Recommendations

This study focused on to review the existing literature of TAR based on FEM and experimental analysis. Various literatures have been found which deals with the stress/strain related failure, load transfer due to implantation, bone density changes due to bone remodelling, assessed existing and alternative design using 3D FE analysis and experimental methods. Insight of the current clinical attention in TAR, some recommendation has been made for future research on TAR.

Various studies were developed FE models of intact and implanted ankle joint based on CT scan or MRI data. All of the studies were assumed that bone was linear, isotropic material. For the better understanding of the load transfer and bone remodelling etc. it is required to assign the accurate bone material properties to the bone, since, in reality, bone is anisotropic and viscoelastic. The time-dependent simulation would be beneficial as the viscoelastic material is varied with time. Apart from the bones, cartilages should be modelled as a time-dependent variable which was not found in most of the previous studies. All soft tissues should be considered for the modelling of ankle joint in order to better understanding of the load transfer. Only ligaments and cartilages were considered by most of the previous studies. More than one CT or MRI data should be used for the development of intact ankle in order to the qualitative assessment of load transfer and stress/strain distribution. Accurate musculoskeletal loading and boundary condition should be considered for the determination of load transfer through the intact and implanted joint. All of the previous studies related to FE and experimental analysis were based on static analysis. The dynamic analysis would be more beneficial for the development of new implants. Hill's muscle model would be beneficial for the better understanding of the effect of all the muscle forces around the ankle joint.

Apart from these, modelling of implant-bone interfacial condition is also plays a crucial role for evaluation of implants or design of a new implants. Till today for TAR analysis, implant-bone interfacial condition was modelled by assuming debonded (fully osseo-integration) and bonded (non osseo-integration) condition for uncemented implants. Initial fixation is generally achieved by amount of interference fit and co-efficient of friction between implant and bone. When time progress bone is generally grow (bone ingrowth) at the porous coated surface of the implant. So, combined bone ingrowth and bone remodelling simulation might be more important for evaluation or design of ankle prosthesis. One of the other aspects is that, during implantation local bone might fail due to press-fit or bone preparation for implant. So, damage modelling of bone might be more appropriate for better understanding the results and better clinical significance. For FE modelling of cemented fixation, bone-cement and cement-implant interface conditions were considered completely fixed. However, it is well known that cement is weak in tension and strong in compression. For this reason, there might be possibility of crack generation at the interfaces. Inclusion of crack at the interface in the FE model might be more beneficial for realistic analysis of implant performance. In reality, bone-cement and cement-implant rarely form a chemical bond, and strength is achieved by interlock between the materials. So, proper modelling at these interfaces along with experimental validation will be more beneficial for better understanding the results.

Conflicts of interest

None.

References

  1. , , , et al . Epidemiology of ankle arthritis: report of a consecutive series of 639 patients from a tertiary orthopaedic centre. Iowa Orthop J. 2005;25:44-46.
    [Google Scholar]
  2. , , , et al . A brief history of total ankle replacement and review of the current status. Med Eng Phys. 2007;29:1056-1064.
    [Google Scholar]
  3. , , . Finite Element Modeling of the Human Foot and Footwear. 2006:145-159.
    [Google Scholar]
  4. , , , . Total ankle replacement Design evolution and results. Acta Orthop Belg. 2010;76:150-161.
    [Google Scholar]
  5. , , . History of total ankle replacement. Clin Podiatr Med Surg. 2013;30:1-19.
    [Google Scholar]
  6. , , . Total ankle replacement--evolution of the technology and future applications. Bull Hosp Jt Dis. 2014;72(1):120-128.
    [Google Scholar]
  7. , , , . Total ankle replacement early experiences with STAR prosthesis. Ulst Med J. 2005;74:9-13.
    [Google Scholar]
  8. , , , . Revision total ankle replacement: an early look at agility to INBONE. Foot Ankle Spec. 2011;4:235-244.
    [Google Scholar]
  9. , , , . 10-year survival of total ankle arthroplasties: a report on 780 cases from the Swedish Ankle Register. Acta Orthop. 2011;82(6):655-659.
    [Google Scholar]
  10. , , , et al . Systematic review of outcome and failure rate of uncemented Scandinavian total ankle replacement. Int Orthop. 2011;35:1751-1758.
    [Google Scholar]
  11. , , , . STAR TM ankle: long-term results. Foot Ankle Int. 2011;32:473-484.
    [Google Scholar]
  12. , , , . Bilateral differences in gait mechanics following total ankle replacement: a two year longitudinal study. Clin Biomech. 2014;29:418-422.
    [Google Scholar]
  13. , , , . The outcome of total ankle replacement: a systematic review and meta-analysis. Bone Joint Lett J. 2013;95-B
    [Google Scholar]
  14. , , , . Failure modes for total ankle arthroplasty: a statistical analysis of the Norwegian Arthroplasty Register. Arch Orthop Trauma Surg. 2014;134:1361-1368.
    [Google Scholar]
  15. , . Biomechanics of the Total Ankle Arthroplasty: Stress Analysis and Bone Remodeling (Master of Science Thesis) 2013
    [Google Scholar]
  16. , , . The effects of implant orientations and implant–bone interfacial conditions on potential causes of failure of tibial component due to total ankle replacement. J Med Biol Eng. 2018;39(4):541-555.
    [Google Scholar]
  17. , , , et al . Stresses in polyethylene lines in a semi-constrained ankle prosthesis. J Biomech Eng. 2004;126
    [Google Scholar]
  18. , , . Effects of implant orientation and implant material on tibia bone strain, implant–bone micromotion, contact pressure, and wear depth due to total ankle replacement. J Eng Med Proc Inst Mech Eng H. 2019;233(3):318-331.
    [Google Scholar]
  19. , , , et al . Total ankle replacement design and positioning affect implant-bone micromotion and bone strains. Med Eng Phys. 2017;42:80-90.
    [Google Scholar]
  20. , , , et al . Numerical simulation of strain-adaptive bone remodeling in the ankle joint. Biomed Eng Online. 2011;10(58):2-13.
    [Google Scholar]
  21. , , , et al . An analysis of soft tissue loading in the foot- a preliminary report. Bull Prosthet Res. 1981;18:27-34.
    [Google Scholar]
  22. , , , et al . Modeling and stress analyses of a normal foot-ankle and a prosthetic foot-ankle complex. Acta Bioeng Biomech. 2013;15
    [Google Scholar]
  23. , , , et al . Using an optimization approach to design an insole for lowering plantar fascia stress-A finite element study. Ann Biomech Eng. 2008;36:1345-1352.
    [Google Scholar]
  24. , , . Clinical applications of computational simulation of foot and ankle. Sport Traumatol.. 2007;23:264-271.
    [Google Scholar]
  25. , , , et al . Nonlinear 3D foot FEA modelling from CT scan medical images. 2007
    [Google Scholar]
  26. , , , et al . Finite element analysis of biomechanical effects of total ankle arthroplasty on the foot. J Ortho Trans. 2018;12:55-65.
    [Google Scholar]
  27. , , , et al . The effect of insoles in therapeutic foot-wear a finite element approach. J Biomech. 1997;30:615-620.
    [Google Scholar]
  28. , , , et al . Biomechanical topography of human articular cartilage in the first metatarsophalangeal joint. Clin Orthop. 1998;348:269-281.
    [Google Scholar]
  29. , , , et al . The mechanical characteristics of the col-lateral ligaments of the human ankle joint. Foot Ankle. 1988;8:234-242.
    [Google Scholar]
  30. , , . A study of the elastic properties of the plantar fascia. J Bone Jt. Surg. 1964;46:482-492.
    [Google Scholar]
  31. , , , . Three-Dimensional Modelling and Analysis of a human ankle joint. 2013
    [Google Scholar]
  32. , , , et al . A biomechanical evaluation of the tibiofibular and tibiotalar ligaments of the ankle. Foot Ankle Int. 2003;24:426-429.
    [Google Scholar]
  33. , , . Computational modelling to predict mechanical function of joints: application to the lower leg simulation of two cadaver studies. J Biomech Eng. 2007;129:811-817.
    [Google Scholar]
  34. , , , et al . Ligament fibre recruitment and forces for the anterior drawer test at the human ankle joint. J Biomech. 2003;36(3):363-372.
    [Google Scholar]
  35. , , . The distal fascicle of the anterior inferior tibiofibular ligament as a cause of tibiotalar impingement syndrome: a current concepts review. Knee Surg Sport Traumatol Arthrosc. 2007;15(4):465-471.
    [Google Scholar]
  36. , , , et al . Development and validation of a computational model to study the effect of foot constraint on ankle injury due to the external rotation. Ann Biomech Eng. 2011;39:756-765.
    [Google Scholar]
  37. , , . A numerical study on stress distribution across the ankle joint: effects of material distribution of bone, muscle force, and ligaments. J Orthop. 2017;14:329-335.
    [Google Scholar]
  38. , , , et al . Finite element analysis of stress and wear characterization in total ankle replacements. J Mech Behav Biomed Mater. 2014;34:I34-I45.
    [Google Scholar]
  39. , , , et al . Wear characteristics of WSU total ankle replacement devices under shear and torsion loads. J Mech Behav Biomed Mater. 2015;44:202-223.
    [Google Scholar]
  40. , , , et al . Misalignment of total ankle components can induce high joint contact pressures. J. Bone Jt. Surg.. 2010;92:1179-1187.
    [Google Scholar]
  41. , , , et al . Contact stresses, pressure and area in a fixed-bearing total ankle replacement: a finite element analysis. BMC Muscoskelet Disord. 2017;18:493.
    [Google Scholar]
  42. , , , et al . Applicability of PEEK and its composites in total ankle replacement devices and wear rate predictions. Biomed Phys Eng Exp. 2016;2
    [Google Scholar]
  43. , , , et al . Finite element analysis of a total ankle replacement during the stance phase of gait. J Biomech. 2006;39:1435-1443.
    [Google Scholar]
  44. , , , et al . Development and experimental validation of a finite element model of total ankle replacement. J Biomech. 2014;47:742-745.
    [Google Scholar]
  45. , , , et al . Study of wear prediction on total ankle replacement. Adv Mater Res. 2014;845:311-315.
    [Google Scholar]
  46. , , , et al . Polyethylene wear in total ankle replacement is influenced by its radial curvature: a computational wear simulation study. 2017
    [Google Scholar]
  47. , , , . Comparison of hexahedral and tetrahedral elements in finite element analysis of the foot and footwear. J Biomech. 2011;44:2337-2343.
    [Google Scholar]
  48. , , , . Compressive mechanical properties of human cancellous bone after gamma radiation. J Bone Joint Surg Am. 1992;74:747-752.
    [Google Scholar]
  49. , , , . Correlations between orthogonal mechanical properties and density of trabecular bone: use of different densitometric measures. J Biomed Mater Res. 1994;28:1329-1336.
    [Google Scholar]
  50. , , , . Trabecular bone modulus-density relationships depend on anatomic site. J Biomech. 2003;36:897-904.
    [Google Scholar]
  51. , , , et al . Biomechanical analysis of three-dimensional foot structure during gait: a basic tool for clinical applications. J Biomech Eng. 2000;122:629-639.
    [Google Scholar]
  52. , , , et al . On the development of an Osseo-ligamentous FE model of the human ankle joint. Int J Solids Struct. 2001;38:1681-1697.
    [Google Scholar]
  53. , , , et al . Physical validation of a patient-specific contact finite element model of the ankle. J Biomech. 2007;40:1662-1669.
    [Google Scholar]
  54. , , . The prediction of muscular load shearing and joint forces in the lower extremities during walking. J Biomech. 1975;8:89-102.
    [Google Scholar]
  55. , , . Ankle joint biomechanics. J Biomech. 1982;15(9):627-634.
    [Google Scholar]
  56. , , , et al . A preliminary numerical model of the human ankle under impact loading. 1995:277-289.
    [Google Scholar]
  57. , , , et al . Lower extremity finite element model development. 1995:263-275.
    [Google Scholar]
  58. , , , et al . Three-dimensional finite element analysis of the foot during standing - a material sensitivity study. J Biomech. 2005;38:1045-1054.
    [Google Scholar]
  59. , , , . The in vivo plantar soft tissue mechanical property under the metatarsal head: implication soft issues' joint-angle dependent response in foot finite element modelling. J Mech Behav Biomed Mater. 2014;40:264-274.
    [Google Scholar]
  60. , , , et al . Effects of internal stress concentrations in plantar soft-tissue—a preliminary three-dimensional finite element analysis. Med Eng Phys. 2010;32:324-331.
    [Google Scholar]
  61. , , , et al . Effects of plantar fascia stiffness on the biomechanical responses of ankle-foot complex. Clin Biomech. 2004;19:839-846.
    [Google Scholar]
  62. , , , et al . Three-dimensional finite element stress analysis of the polypropylene, ankle-foot orthosis: static analysis. Med Eng Phys. 1995;17:372-379.
    [Google Scholar]
  63. , , , et al . Finite element modeling of a 3D coupled foot-boot model. Med Eng Phys. 2011;33:1228-1233.
    [Google Scholar]
  64. , , . Human foot modeling towards footwear design. Com Aid Deg. 2011;43:1841-1848.
    [Google Scholar]
  65. , , . Effects of boundary conditions on foot behaviour in the standing position in 3D finite element foot model. J Foot Ankle Res. 2014;7:A38.
    [Google Scholar]
  66. , , , . Effects of bone Young's modulus on finite element analysis in the lateral ankle biomechanics. Appl Bionics Biomechanics. 2013;10:189-195.
    [Google Scholar]
  67. , , , . 3D finite element model of the diabetic neuropathic foot: a gait analysis driven approach. J Biomech. 2014;47:3064-3071.
    [Google Scholar]
  68. , , , . Numerical and experimental investigation of the structural behaviour of a carbon fiber reinforced ankle-foot orthosis. Med Eng Phys. 2015;37:505-511.
    [Google Scholar]
  69. , , , et al . Ankle–foot simulator development for testing ankle–foot orthosis. Med Eng Phys. 2010;32:623-629.
    [Google Scholar]
  70. , , , et al . Evaluation of total ankle arthroplasty using highly crosslinked ultrahigh-molecular-weight polyethylene subjected to physiological loading. Foot Ankle Int 2019
    [Google Scholar]
  71. , , , et al . Kinematics changes after fusion and total replacement of the ankle. Part 1-Range of motion (ROM) Foot Ankle Int. 2003;24:881-887.
    [Google Scholar]
  72. , , , et al . Kinematics changes after fusion and total replacement of the ankle. Part 2-Movement Transfer. Foot Ankle Int. 2003;24:888-896.
    [Google Scholar]
  73. , , , et al . Kinematics changes after fusion and total replacement of the ankle. Part 3-Talar Movement. Foot Ankle Int. 2003;24:897-900.
    [Google Scholar]
  74. , , , et al . The effect of agility ankle prosthesis misalignment on the peri-ankle ligaments. Clin Orthop Relat Res. 2004;424:137-142.
    [Google Scholar]
  75. , , , et al . Robotic cadaver testing of a new total ankle prosthesis model. Foot Ankle Int. 2007;28:1276-1286.
    [Google Scholar]
  76. , , , et al . Wear behaviour in total ankle replacement: a comparison between an in vitro simulation and retrieved prostheses. Clin Biomech 2009
    [Google Scholar]
  77. , , , et al . Influence of kinematics on the wear of a total ankle replacement. J Biomech. 2017;53:105-110.
    [Google Scholar]
Show Sections