View/Download PDF
67 (); 223-229
doi:
10.1016/j.jor.2025.05.035

Translate:

A novel three-dimensional proximal femoral intramedullary nail (TD-PFIN) system with better biomechanical advantages: a finite element analysis study

Author affiliations
Department of SICU, Beijing Chaoyang Hospital, Capital Medical University, Beijing, 100020, China
Department of Orthopedics, Beijing Chaoyang Hospital, Capital Medical University, Beijing, 100020, China
Article History
Corresponding author(s)

⁎Corresponding author: Dong Wang. dongwang@mail.ccmu.edu.cn

Article tools · PDF & citations

Abstract

Abstract

The biomechanical strength of a new three-dimensional proximal femur intramedullary nail (TD-PFIN) was compared with that of the commonly used internal fixation by finite element analysis.

The femur CT imaging data of a healthy volunteer were collected. Three-dimensional full-length femur was conducted by software and osteotomy was performed at the intertrochanteric crest to make AO/OTA 31-A1.1 fracture model. DHS, PFNA, InterTAN and TD-PFIN fixation systems were simultaneously made.TD-PFIN had three head nails, which were vertically and horizontally dispersed in the proximal femur. Normal physiological weight bearing was simulated by finite element analysis, and peak von Mises stress and the maximum displacement in the femur and internal fixation were recorded.

An axial load of 700N was applied to the femoral head to simulate normal adult weight. Among the fixation systems, the TD-PFIN had the lowest peak von-Mises stress, which was 111.79 MPa. the DHS model had the highest stress concentration, which was 231.71 MPa. In the Von-Mises stress distribution of the femur, the maximum stress was 14.696 MPa in DHS fixed model and the least was 10.162 MPa in TD-PFIN fixed model. In the displacement distribution of the internal fixations, The DHS model was 3.8991 mm, which was the most. The TD-PFIN fixed model had a minimum displacement of 3.1571m. In the displacement distribution in femur, the DHS internal fixation model was the largest (4.017 mm). The TD-PFIN fixed mod was the smallest (3.1886 mm).

TD-PFIN designed by us had the characteristics of multi-plane fixed, with better anti-rotation ability. Its mechanical stability was better than DHS, PFNA, and InterTAN.

Keywords

  • Three-dimensional intramedullary nail
  • Multi-plane fixed
  • Intertrochanteric fracture of the femur
  • Finite element analysis
Disclaimer:
This article was originally published by Reed Elsevier India Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
1

1 Introduction

Intertrochanteric fracture of the femur (IFF) is among the most common types of fractures in clinical practice. As the population aging increased, the number of IFF patients is reaching approximately 3 million at 2050.1,2 For patients without surgical contraindications, early surgical intervention should be performed.3,4 After surgery, patients can move the affected limb freely in a non-weight-bearing state, which prevent the complications of prolonged bed rest, such as pressure ulcers, deep vein thrombosis, and hypostatic pneumonia.5–7

Surgical techniques include intramedullary fixation and extramedullary fixation. Extramedullary fixation utilizes plate and screw, which require extensive soft tissue dissection and make more intraoperative blood loss.8,9 Moreover, extramedullary fixation is an eccentric fixation that fails to provide sufficient medial support, making it unsuitable for reverse intertrochanteric fractures.10

Intramedullary fixation is using intramedullary nail for fracture fixation, which has minimal surgical trauma, shorter operation time, and less intraoperative blood loss.8–10 It is now widely used in clinic. Although intramedullary nailing is central fixation, currently used such as proximal femoral nail anti-rotation (PFNA) and InterTan nail, all are single-plane fixation, which provide limited resistance to rotational forces.11,12

Based on the design scheme of proximal humeral intramedullary nails, we have developed a three-dimensional proximal femoral intramedullary nail (TD-PFIN). The head screws of TD-PFIN were three, which was pointing anteriorly, centrally, and posteriorly. The three head screws formed a multi-planar fixation system with the main nail and distal locking screws. We thought that TD-PFIN could effectively prevent rotation of fracture. In this study, we used finite element analysis to employe the fixation strength of TD-PFIN with that of the PFNA, InterTan, and DHS, in order to investigate the biomechanical advantages. The flow chart of this study was shown in Fig. 1 (By Figdraw.).

Schematic diagram of the research process of this study Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
Fig. 1 Schematic diagram of the research process of this study Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
2

2 Material and methods

2.1

2.1 Construction of AO/OTA 31-A1.1 IFF model

The study was conducted in accordance with the ethical standards set out in the Declaration of Helsinki and its subsequent amendments. The design of the study was approved by the Ethics Committee of Beijing Chaoyang Hospital, Capital Medical University.

A healthy volunteer femoral CT imaging data were collected. The volunteer was 40 years old, 170 cm height, and 73 kg weight. The CT scan was performed with a slice thickness of 0.625 mm, a voltage setting between 70 and 140 kV, and a current setting between 30 and 800 mA. The scan data were saved in DICOM format and imported into Mimics 21.0 software (Materialize's Interactive Medical Image Control System/Materialize NV, Belgium) for three-dimensional reconstruction. Cortex and cancellous bones were identified by Hounsfield Unit (HU) values, with a boundary value set at 700.13,14 Based on previous literature and the AO/OTA classification, an osteotomy was performed at the intertrochanteric crest to construct an AO/OTA 31-A1.1 IFF model (Fig. 2).

Fracture model Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
Fig. 2 Fracture model Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
2.2

2.2 Construction of internal fixations

The internal fixation devices were modeled using SolidWorks 2021 (Dassault Systèmes, USA). Based on published literature and manufacturer specifications, models of the DHS, PFNA, and InterTAN systems were constructed.13–15 The parameters of the TD-PFIN were set to match those of the InterTAN in terms of the lengths of the main nail, cephalic screws, and distal locking screw. The TD-PFIN features three cephalic screws. The central axis of the middle cephalic screw is aligned with the longitudinal axis of the main nail, while the other two cephalic screws are oriented at an angle of 12.15° relative to the middle screw. All cephalic screws are threaded, with the threaded portion accounting for half the total length of each screw (Fig. 3).

Internal fixator model Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
Fig. 3 Internal fixator model Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
2.3

2.3 Material characteristics

Geomagic Wrap 18.0 (Raindrop, USA) was used to generate mesh models of the femur and internal fixation devices. Material properties were assigned according to values reported in previous studies (Table 1).13–15 All materials were assumed to be homogeneous, isotropic, and linearly elastic. A tetrahedral mesh was applied for model discretization.

Table 1 Material parameters of bone and internal fixations.
Part Component Young's modulus (MPa) Poisson's ratio
Bone Cortical bone 17000 0.33
Cancellous bone 1000 0.30
Implant Plate & Screw & Nail 110000 0.35

For the contact conditions, frictional contact was applied between bone surfaces with a friction coefficient of 0.46, and between the bone and the internal fixation with a friction coefficient of 0.3.13–15 The contacts between the screw and the main screw/plate as well as between screws were defined as bonded contacts to simulate their fixed state (Table 2).

Table 2 Number of nodes and elements divided by Meshing.
Mesh Bone DHS PFNA InterTAN TD-PFIN
Number of nodes 2033153 150919 184174 212068 243103
Number of elements 1473729 94797 135125 165894 186215
2.4

2.4 Boundary conditions and loading settings

ANSYS Workbench 2021 (Ansys Inc., USA) software was used to make finite element analysis. The numbers of finite element mesh units and nodes were shown in Table 1.

The distal end of the femur was fully constrained in all degrees of freedom to simulate physiological fixation. A vertical load of 700 N, corresponding to approximately the body weight of an average adult during single-leg stance, was applied to the femoral head to simulate hip joint loading.13–15 The loading direction was aligned with the mechanical axis of the femur. To ensure numerical stability and avoid stress concentration, the load was distributed over a small circular area at the femoral head. All contacts and constraints were set based on anatomical and biomechanical relevance (Fig. 4).

The forces loaded on the model and the fixed plane Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
Fig. 4 The forces loaded on the model and the fixed plane Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
2.5

2.5 Model validation strategy

To validate the accuracy and reliability of the finite element model, the simulated results were compared with data reported in previous studies under similar boundary and loading conditions.13–15 Key parameters such as stress distribution patterns and displacement trends were analyzed to ensure consistency. Mesh sensitivity analysis was also conducted to confirm that the results were independent of element size.

2.6

2.6 Main evaluation parameters

To evaluate the biomechanical performance of different internal fixation systems, several key parameters were assessed. These included the peak von Mises stress and the maximum displacement in the femur and internal fixation. Peak stress values were used to assess the risk of mechanical failure, while displacement served as indicators of overall construct stability. Lower stress concentrations and reduced micromotion were considered indicative of better biomechanical performance and favorable conditions for fracture healing.

3

3 Results

3.1

3.1 The Von-Mises stress distribution analysis of the internal fixations

The von-Mises stress distribution of the four fixations was analyzed under the same loading and boundary conditions. Stress was primarily at regions subjected to high mechanical load transfer. In DHS, peak stress was concentrated around the junction of cephalic screw and the lateral plate.

The PFNA, InterTAN and TD-PFIN showed stress concentration primarily around the junction of the cephalic screws and the main nail. Among the fixation systems, the TD-PFIN had the lowest peak von-Mises stress, which was 111.79 MPa, indicating a more favorable stress distribution (Fig. 5).

The Von-Mises stress distribution of the internal fixations Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
Fig. 5 The Von-Mises stress distribution of the internal fixations Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.

In contrast, the DHS model showed the highest stress concentration, which was 231.71 MPa, suggesting a higher risk of implant fatigue and failure. The PFNA and InterTAN had intermediate stress levels compared to DHS and TD-PFIN (Fig. 6).

The Von-Mises stress peak of the internal fixations and femurs Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
Fig. 6 The Von-Mises stress peak of the internal fixations and femurs Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
3.2

3.2 The Von-Mises stress distribution of the femur

In the DHS fixation model, the femoral stress was concentrated at the femoral lateral cortex. In the intramedullary nail-fixation model, the PFNA, InterTAN and TD-PFIN models, the femoral stress was concentrated at the femoral internal cortex (Fig. 7).

The Von-Mises stress distribution of the femurs Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
Fig. 7 The Von-Mises stress distribution of the femurs Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.

The maximum stress of femur in DHS fixed model was 14.696 MPa. The stress of TD-PFIN fixed model was the least, which was 10.162 MPa. PFNA and InterTAN stress were between DHS and TD-PFIN, 12.801 MPa and 10.868 MPa, respectively (Fig. 6).

3.3

3.3 The displacement distribution analysis of the internal fixations

The displacement distribution of the four internal fixation model was evaluated to assess their structural stability under physiological loading. In all models, the largest displacements were at the femoral head, gradually decreasing along the implant. The DHS model had the most displacement, which was 3.8991 mm (Fig. 8).

The displacement distribution of the internal fixations Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
Fig. 8 The displacement distribution of the internal fixations Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.

The displacement of PFNA and InterTAN internal fixation models was between DHS and TD-PFIN fixation models, 3.465 mm and 3.2816 mm, respectively. The TD-PFIN fixed model had a minimum displacement of 3.1571 mm, which meant that it could provide a stronger fixation (Fig. 9).

The displacement peak of the internal fixations and femurs Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
Fig. 9 The displacement peak of the internal fixations and femurs Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
3.4

3.4 Analysis of displacement distribution in femur

The displacement of the femur was used to assess the fixation strength of the internal fixation in the physiological weight bearing. In the four internal fixation models, the greatest area of femur displacement was the femoral head, and then gradually decreased along the longitudinal axis of the femur (Fig. 10).

The displacement distribution of the femurs Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.
Fig. 10 The displacement distribution of the femurs Notes: Abbreviations: TD-PFIN, three-dimensional proximal femur intramedullary nail; IFF, Intertrochanteric fracture of the femur; PFNA, proximal femoral nail anti-rotation.

In DHS internal fixation model, the femur displacement was the largest (4.017 mm). In TD-PFIN fixed model, the femur displacement was the smallest (3.1886 mm). The femur displacement of PFNA and InterTAN fixed models was between the two, 3.3635 mm and 3.2805 mm, respectively (Fig. 9).

4

4 Discussion

This study was based on a new three-dimensional proximal femur intramedullary nail (TD-PFIN) system designed and developed by ourselves. The design concept of TD-PFIN was derived from the proximal humerus intramedullary nail.16 Proximal humerus intramedullary nails, such as the multilock fixation system, were multi-plane fixation to prevent rotation and inversion of the humerus head.16,17 Currently, the proximal femoral intramedullary nail was fixed in a single plane, such as Gamma III intramedullary nail, PFNA intramedullary nail, intertan intramedullary nail, etc.11,12,18 Although single-plane fixation made the operation of these intramedullary nails simple and easy to learn and master, the incidence of postoperative internal fixation failure is still high.10,12,18 Clinicians often paid more attention to the anti-varus function of intramedullary nails, ignoring the anti-rotation ability. In fact, the femoral neck had a forward angle of about 15°, so when walking normally, the proximal femur also carried a certain torsional stress. Therefore, there was a great clinical need to design a proximal femoral intramedullary nail with anti-rotation function.

In this study, TD-PFIN was compared with internal fixations commonly used in clinical practice, such as DHS extramedullary fixation system, PFNA, intertan intramedullary nail, to analyze whether it had better fixation strength. We used the most common intertrochanteric fracture model in the clinic. Normal physiological load was simulated on the femoral head. We found that the TD-PFIN fixation model had the smallest internal fixation Von-Mises stress distribution, the smallest femur Von-Mises stress distribution, the smallest internal fixation displacement, and the smallest femur displacement compared to the other three internal fixation systems. These meant that the TD-PFIN intramedullary nail had better fixation strength. It could allow patients to carry weight earlier, walk down, and recover hip function faster. TD-PFIN had three head nails to fix the proximal fracture fragment. The femoral head, femoral neck and intramedullary main nail were fixed together with three head nails, and the fracture end could also be pressurized to a certain extent. Therefore, TD-PFIN was also suitable for basal femoral neck fractures than the other intramedullary nail system.

In this study, finite element method was used to analyze the mechanical fixed strength of internal fixators, which had some limitations. First of all, the finite element model did not add soft tissue, such as the iliopsoas muscle, gluteus media and other structures, resulting in certain differences between the model and the normal body. Second, the femur model used in the study was an Asian normal adult male, and the population might have limitations. Thirdly, the simulated loads used in this study were static loads, not dynamic loads. In the future, we plan to conduct cadaver and clinical studies.

5

5 Conclusion

TD-PFIN was a novel proximal femoral intramedullary nail fixation system designed by us, which had the characteristics of multi-plane fixation. Compared with the internal fixation system commonly used in clinical practice, it had better biomechanical advantages.

CRediT authorship contribution statement

Xi Zheng: Writing – original draft. Hanzhou Wang: Made the finite element analysis and prepared Figures. 2-5,7,8,10. Junlin Zhou: Provide technical guidance and review the original draft. Dong Wang: Prepared Figures. 1,6,9, All the authors reviewed the manuscripts.

Ethics statement

The Ethics Committee of Beijing Chaoyang hospital approved the study and the use of computed tomography data for this study (Approval number 2025-ke-21). Informed consent was obtained from the healthy volunteer whose computed tomography data were used in this study. All methods were carried out in accordance with relevant guidelines and regulations.

Research involving human data was conducted in accordance with the ethical standards set out in the Declaration of Helsinki and its subsequent amendments.

Funding

This work was supported by the National Natural Science Foundation of China (grant number: 82272469) and the Beijing Key Clinical Specialty Project.

References

  1. , , , et al . Nomogram based on high-density lipoprotein cholesterol for the occurrence of preoperative deep vein thrombosis in patients with intertrochanteric femur fracture: a retrospective study. J Orthop Surg Res. 2024;19(1):22.
    [Google Scholar]
  2. , , , et al . Impact of femoral head size on the localization of proximal femur fractures: retrospective analysis of 400 cases. J Clin Orthop Trauma. 2025;65
    [Google Scholar]
  3. , , , et al . Optimising reduction and implant positioning in intertrochanteric fracture treatment: an evaluation of the effects of a structured educational program. Injury. 2025;56(2)
    [Google Scholar]
  4. , , , et al . Comparative outcomes of proximal femur intramedullary nailing vs. cemented bipolar hemiarthroplasty for treating intertrochanteric fractures in patients aged 75 and older: analysis of risk factors for postoperative all-cause mortality. BMC Surg. 2025;25(1):130.
    [Google Scholar]
  5. , , , et al . The effect of calf muscular vein thrombosis on the prognosis within one year postoperatively of geriatric hip fracture patients: a propensity score-matched analysis. BMC Geriatr. 2024;24(1):1050.
    [Google Scholar]
  6. , , , et al . Treatment of irreducible intertrochanteric femoral fracture with a minimally invasive clamp reduction technique via the anterior approach. J Orthop Surg Res. 2023;18(1):167.
    [Google Scholar]
  7. , , , et al . Nutritional interventions for pressure ulcer prevention in hip fracture patients: a systematic review and meta-analysis of controlled trials. Nutrients. 2025;17(4):644.
    [Google Scholar]
  8. , , , et al . Risk of internal fixation treatment in intertrochanteric fracture based on different lateral femoral wall thickness: finite element analysis. BMC Musculoskelet Disord. 2024;25(1):462.
    [Google Scholar]
  9. , , , et al . Mobility and living status at discharge and after three-months for extramedullary versus intramedullary fixation of AO type 31-A1 trochanteric fractures; an analysis of Dutch hip fracture audit data. Eur J Trauma Emerg Surg. 2025;51(1):6.
    [Google Scholar]
  10. , , , et al . Finite element analysis of fixation stability according to reduction position for internal fixation of intertrochanteric fractures. Sci Rep. 2024;14(1)
    [Google Scholar]
  11. , , , et al . Proximal femoral nail antirotation versus InterTan nail for the treatment of intertrochanteric fractures: a systematic review and meta-analysis. PLoS One. 2024;19(7)
    [Google Scholar]
  12. , , , et al . A comparative study of the early postoperative outcome of three intramedullary fixation modalities in the treatment of intertrochanteric fractures of the femur in the elderly. J Musculoskelet Neuronal Interact. 2024;24(3):310-317.
    [Google Scholar]
  13. , , , et al . Finite element analysis of proximal femoral bionic nail (PFBN), proximal femoral nail antirotation and InterTan for treatment of reverse obliquity intertrochanteric fractures. Orthop Surg. 2025;17(3):888-899.
    [Google Scholar]
  14. , , , et al . Biomechanical characteristic differences of two new types of intramedullary nail devices in the treatment of comminuted intertrochanteric fractures of femur: a comparative study based on finite element analysis. J Orthop Surg Res. 2024;19(1):583.
    [Google Scholar]
  15. , , , et al . Biomechanical evaluation of the modified proximal femoral nail for the treatment of reverse obliquity intertrochanteric fractures. Sci Rep. 2025;15(1):3261.
    [Google Scholar]
  16. , , , et al . Angular stable intramedullary nail versus locking plate fixation of osteoporotic surgical neck proximal humerus fractures: a biomechanical comparison. Shoulder Elbow. 2025;13
    [Google Scholar]
  17. , , , et al . Intramedullary nailing of humeral shaft fractures: percutaneous nailing in the lateral position through the Neviaser approach is a reliable technique. J Shoulder Elb Surg. 2024;33(10):2178-2186.
    [Google Scholar]
  18. , , , et al . A novel intramedullary nail design of intertrochanteric fracture fixation improved by proximal femoral nail antirotation. Comput Methods Biomech Biomed Engin. 2025;28(2):292-302.
    [Google Scholar]
Show Sections